Contrast Enhanced Computed Tomography (CT) Scanning
Conditions
Brief summary
The purpose of this research study is to compare two types of catheters used for infusing the contrast dye necessary for viewing internal organs during computed tomography (CT) scanning. The standard catheter, which has one hole from which contrast dye enters veins, will be compared to the Becton Dickinson Nexiva Diffusics System catheter, which has three holes for diffusing dye into veins. The purpose of this study is to determine which catheter type gives the best image.
Detailed description
This will be a prospective, randomized, controlled trial. The catheter is already in use at our facilities. After informed consent, patients will be randomly given the standard or new catheter in our practice. Envelopes with the random assignment will be available to the individuals consenting patients. Randomization generally ensures that study groups are similar on average. As age, sex, and weight impact image quality, we will collect data on these factors in case randomization results in an imbalance. This will allow for control of these factors in analysis of the data. Data will be collected from CT abdomen pelvis examinations within a 6 month-1 year period in 2016 - 2017. Our primary outcome is image quality, measured as tissue density and distribution of contrast within the liver, spleen, aorta and IVC. Our secondary outcomes include Peak Contrast Infusion Pressure and Injection Flow Rate. Possible adverse events for use of any type of catheter include contrast infiltration into the subcutaneous tissues which is a minor limited event. All infiltrations will be recorded as part of the permanent medical record, documented in the chart, and the referring physician notified. The patient will be assessed by a physician immediately and referred to the ED if necessary for further evaluation. Current technology utilizes Power injector equipment which alerts the technologist to the pressure being achieved. If the injection pressure exceeds 325 psi the Power injector automatically shuts off.
Interventions
Becton Dickinson Nexiva Diffusics System will be used to inject IV contrast and will be compared to a single hole catheter to see if it improves image quality in the main vessels in the abdomen and pelvis because of its novel design.
Standard catheter with one hole at the tip will be compared to a novel 3 hole catheter to see which catheter provides the best image quality.
Sponsors
Study design
Eligibility
Inclusion criteria
* Patients undergoing contrast enhanced CT abdomen pelvis examinations at Lyndon B. Johnson General Hospital in Texas * Patients whose access point is the antecubital fossa * Patients receiving 20 or 22 gauge catheters for their CT exam * Patients scanned on two 64 slice CT scanners at the same hospital
Exclusion criteria
* patients whose access point is other than antecubital fossa
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta | at the time of CT scanning | The ratio of tissue density measured in the main portal vein (MPV) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality. |
| Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta | at the time of CT scanning | The ratio of tissue density measured in the spleen to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality. |
| Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta | at the time of CT scanning | The ratio of tissue density measured in the inferior vena cava (IVC) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality. |
| Image Quality as Assessed by Rating Scale | at the time of CT scanning | Radiologists blinded to catheter type will review images and rate them on a scale of 1 to 10. 10 indicates a better outcome. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Peak Contrast Infusion Pressure | at the time of CT scanning | The peak contrast infusion pressure is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study. |
| Injection Flow Rate | at the time of CT scanning | The injection flow rate is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter Becton Dickinson (BD) Nexiva™ Diffusics™ System uses three laser-cut tear-drop holes located in the catheter tip to stabilize the catheter tip by reducing flow velocities while maintaining flow rates during intravenous contrast administration into the vein. | 31 |
| Standard Intravenous Non-fenestrated Catheter The standard catheter has one hole at the tip for contrast infusion. | 30 |
| Total | 61 |
Baseline characteristics
| Characteristic | Total | Standard Intravenous Non-fenestrated Catheter | Becton Dickinson Nexiva Diffusics System Fenestrated Catheter |
|---|---|---|---|
| Abdominal anterioposterior diameter at lower kidneys | 26.8 centimeters STANDARD_DEVIATION 4.7 | 26.6 centimeters STANDARD_DEVIATION 4.4 | 27.0 centimeters STANDARD_DEVIATION 5.1 |
| Abdominal transverse (left-right) diameter at lower kidneys | 37.1 centimeters STANDARD_DEVIATION 6.1 | 37.0 centimeters STANDARD_DEVIATION 5.7 | 38.7 centimeters STANDARD_DEVIATION 7.9 |
| Age, Continuous | 51.9 years STANDARD_DEVIATION 12.5 | 52.9 years STANDARD_DEVIATION 13.8 | 50.8 years STANDARD_DEVIATION 11.2 |
| Race and Ethnicity Not Collected | 0 Participants | — | — |
| Region of Enrollment United States | 61 participants | 30 participants | 31 participants |
| Sex: Female, Male Female | 33 Participants | 17 Participants | 16 Participants |
| Sex: Female, Male Male | 28 Participants | 13 Participants | 15 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 0 / 31 | 0 / 30 |
| other Total, other adverse events | 0 / 31 | 0 / 30 |
| serious Total, serious adverse events | 0 / 31 | 0 / 30 |
Outcome results
Image Quality as Assessed by Rating Scale
Radiologists blinded to catheter type will review images and rate them on a scale of 1 to 10. 10 indicates a better outcome.
Time frame: at the time of CT scanning
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Image Quality as Assessed by Rating Scale | 8.5 units on a scale | Standard Deviation 0.97 |
| Standard Intravenous Non-fenestrated Catheter | Image Quality as Assessed by Rating Scale | 8.6 units on a scale | Standard Deviation 0.87 |
Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta
The ratio of tissue density measured in the inferior vena cava (IVC) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Time frame: at the time of CT scanning
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta | 0.52 Ratio | Standard Deviation 0.15 |
| Standard Intravenous Non-fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta | 0.52 Ratio | Standard Deviation 0.18 |
Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta
The ratio of tissue density measured in the main portal vein (MPV) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Time frame: at the time of CT scanning
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta | 0.95 Ratio | Standard Deviation 0.18 |
| Standard Intravenous Non-fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta | 0.95 Ratio | Standard Deviation 0.18 |
Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta
The ratio of tissue density measured in the spleen to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Time frame: at the time of CT scanning
Population: Data for the Becton Dickinson Nexiva Diffusics System fenestrated catheter arm was not collected for one patient because the patient had a splenectomy.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta | 0.72 Ratio | Standard Deviation 0.12 |
| Standard Intravenous Non-fenestrated Catheter | Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta | 0.69 Ratio | Standard Deviation 0.12 |
Injection Flow Rate
The injection flow rate is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.
Time frame: at the time of CT scanning
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Injection Flow Rate | 2.4 milliliters per second (mL/sec) | Standard Deviation 0.4 |
| Standard Intravenous Non-fenestrated Catheter | Injection Flow Rate | 2.3 milliliters per second (mL/sec) | Standard Deviation 0.3 |
Peak Contrast Infusion Pressure
The peak contrast infusion pressure is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.
Time frame: at the time of CT scanning
Population: Peak contrast infusion pressure data was not collected by the technologist for one patient in the Standard intravenous non-fenestrated catheter arm.
| Arm | Measure | Value (MEDIAN) |
|---|---|---|
| Becton Dickinson Nexiva Diffusics System Fenestrated Catheter | Peak Contrast Infusion Pressure | 80 Pounds per square inch (PSI) |
| Standard Intravenous Non-fenestrated Catheter | Peak Contrast Infusion Pressure | 93 Pounds per square inch (PSI) |