Effect of Three Side Hole Infusion Catheter on CT Image Quality Compared to a Standard Infusion Catheter

IV Contrast Administration: Effect of Three Side Hole Infusion Catheter on CT Image Quality, Infusion Pressure, and Infusion Rates, Compared to a Standard Infusion Catheter

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02943226
Enrollment
61
Registered
2016-10-24
Start date
2016-10-31
Completion date
2016-11-30
Last updated
2018-06-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Contrast Enhanced Computed Tomography (CT) Scanning

Brief summary

The purpose of this research study is to compare two types of catheters used for infusing the contrast dye necessary for viewing internal organs during computed tomography (CT) scanning. The standard catheter, which has one hole from which contrast dye enters veins, will be compared to the Becton Dickinson Nexiva Diffusics System catheter, which has three holes for diffusing dye into veins. The purpose of this study is to determine which catheter type gives the best image.

Detailed description

This will be a prospective, randomized, controlled trial. The catheter is already in use at our facilities. After informed consent, patients will be randomly given the standard or new catheter in our practice. Envelopes with the random assignment will be available to the individuals consenting patients. Randomization generally ensures that study groups are similar on average. As age, sex, and weight impact image quality, we will collect data on these factors in case randomization results in an imbalance. This will allow for control of these factors in analysis of the data. Data will be collected from CT abdomen pelvis examinations within a 6 month-1 year period in 2016 - 2017. Our primary outcome is image quality, measured as tissue density and distribution of contrast within the liver, spleen, aorta and IVC. Our secondary outcomes include Peak Contrast Infusion Pressure and Injection Flow Rate. Possible adverse events for use of any type of catheter include contrast infiltration into the subcutaneous tissues which is a minor limited event. All infiltrations will be recorded as part of the permanent medical record, documented in the chart, and the referring physician notified. The patient will be assessed by a physician immediately and referred to the ED if necessary for further evaluation. Current technology utilizes Power injector equipment which alerts the technologist to the pressure being achieved. If the injection pressure exceeds 325 psi the Power injector automatically shuts off.

Interventions

DEVICEBecton Dickinson Nexiva Diffusics System

Becton Dickinson Nexiva Diffusics System will be used to inject IV contrast and will be compared to a single hole catheter to see if it improves image quality in the main vessels in the abdomen and pelvis because of its novel design.

DEVICEStandard intravenous catheter

Standard catheter with one hole at the tip will be compared to a novel 3 hole catheter to see which catheter provides the best image quality.

Sponsors

The University of Texas Health Science Center, Houston
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
DIAGNOSTIC
Masking
TRIPLE (Subject, Investigator, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients undergoing contrast enhanced CT abdomen pelvis examinations at Lyndon B. Johnson General Hospital in Texas * Patients whose access point is the antecubital fossa * Patients receiving 20 or 22 gauge catheters for their CT exam * Patients scanned on two 64 slice CT scanners at the same hospital

Exclusion criteria

* patients whose access point is other than antecubital fossa

Design outcomes

Primary

MeasureTime frameDescription
Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aortaat the time of CT scanningThe ratio of tissue density measured in the main portal vein (MPV) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aortaat the time of CT scanningThe ratio of tissue density measured in the spleen to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aortaat the time of CT scanningThe ratio of tissue density measured in the inferior vena cava (IVC) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.
Image Quality as Assessed by Rating Scaleat the time of CT scanningRadiologists blinded to catheter type will review images and rate them on a scale of 1 to 10. 10 indicates a better outcome.

Secondary

MeasureTime frameDescription
Peak Contrast Infusion Pressureat the time of CT scanningThe peak contrast infusion pressure is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.
Injection Flow Rateat the time of CT scanningThe injection flow rate is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.

Countries

United States

Participant flow

Participants by arm

ArmCount
Becton Dickinson Nexiva Diffusics System Fenestrated Catheter
Becton Dickinson (BD) Nexiva™ Diffusics™ System uses three laser-cut tear-drop holes located in the catheter tip to stabilize the catheter tip by reducing flow velocities while maintaining flow rates during intravenous contrast administration into the vein.
31
Standard Intravenous Non-fenestrated Catheter
The standard catheter has one hole at the tip for contrast infusion.
30
Total61

Baseline characteristics

CharacteristicTotalStandard Intravenous Non-fenestrated CatheterBecton Dickinson Nexiva Diffusics System Fenestrated Catheter
Abdominal anterioposterior diameter at lower kidneys26.8 centimeters
STANDARD_DEVIATION 4.7
26.6 centimeters
STANDARD_DEVIATION 4.4
27.0 centimeters
STANDARD_DEVIATION 5.1
Abdominal transverse (left-right) diameter at lower kidneys37.1 centimeters
STANDARD_DEVIATION 6.1
37.0 centimeters
STANDARD_DEVIATION 5.7
38.7 centimeters
STANDARD_DEVIATION 7.9
Age, Continuous51.9 years
STANDARD_DEVIATION 12.5
52.9 years
STANDARD_DEVIATION 13.8
50.8 years
STANDARD_DEVIATION 11.2
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
61 participants30 participants31 participants
Sex: Female, Male
Female
33 Participants17 Participants16 Participants
Sex: Female, Male
Male
28 Participants13 Participants15 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 310 / 30
other
Total, other adverse events
0 / 310 / 30
serious
Total, serious adverse events
0 / 310 / 30

Outcome results

Primary

Image Quality as Assessed by Rating Scale

Radiologists blinded to catheter type will review images and rate them on a scale of 1 to 10. 10 indicates a better outcome.

Time frame: at the time of CT scanning

ArmMeasureValue (MEAN)Dispersion
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterImage Quality as Assessed by Rating Scale8.5 units on a scaleStandard Deviation 0.97
Standard Intravenous Non-fenestrated CatheterImage Quality as Assessed by Rating Scale8.6 units on a scaleStandard Deviation 0.87
p-value: 0.4487two sample t-test
Primary

Image Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta

The ratio of tissue density measured in the inferior vena cava (IVC) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.

Time frame: at the time of CT scanning

ArmMeasureValue (MEAN)Dispersion
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta0.52 RatioStandard Deviation 0.15
Standard Intravenous Non-fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Inferior Vena Cava (IVC) to the Aorta0.52 RatioStandard Deviation 0.18
Primary

Image Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta

The ratio of tissue density measured in the main portal vein (MPV) to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.

Time frame: at the time of CT scanning

ArmMeasureValue (MEAN)Dispersion
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta0.95 RatioStandard Deviation 0.18
Standard Intravenous Non-fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Main Portal Vein (MPV) to the Aorta0.95 RatioStandard Deviation 0.18
Primary

Image Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta

The ratio of tissue density measured in the spleen to tissue density in the aorta on the CT images will be determined and will provide a measure of biodistribution, which is an important indicator of image quality. Dividing each Hounsfield unit measurement by the aorta density and determining these ratios will be done on a per patient basis to assess biodistribution as a measure of image quality.

Time frame: at the time of CT scanning

Population: Data for the Becton Dickinson Nexiva Diffusics System fenestrated catheter arm was not collected for one patient because the patient had a splenectomy.

ArmMeasureValue (MEAN)Dispersion
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta0.72 RatioStandard Deviation 0.12
Standard Intravenous Non-fenestrated CatheterImage Quality as Assessed by Ratio of Tissue Density of Spleen to the Aorta0.69 RatioStandard Deviation 0.12
Secondary

Injection Flow Rate

The injection flow rate is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.

Time frame: at the time of CT scanning

ArmMeasureValue (MEAN)Dispersion
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterInjection Flow Rate2.4 milliliters per second (mL/sec)Standard Deviation 0.4
Standard Intravenous Non-fenestrated CatheterInjection Flow Rate2.3 milliliters per second (mL/sec)Standard Deviation 0.3
p-value: 0.0784two sample t-test
Secondary

Peak Contrast Infusion Pressure

The peak contrast infusion pressure is recorded in the electronic health record (EHR) during every CT examination preformed at the hospitals taking part in this study.

Time frame: at the time of CT scanning

Population: Peak contrast infusion pressure data was not collected by the technologist for one patient in the Standard intravenous non-fenestrated catheter arm.

ArmMeasureValue (MEDIAN)
Becton Dickinson Nexiva Diffusics System Fenestrated CatheterPeak Contrast Infusion Pressure80 Pounds per square inch (PSI)
Standard Intravenous Non-fenestrated CatheterPeak Contrast Infusion Pressure93 Pounds per square inch (PSI)
p-value: 0.0112Wilcoxon (Mann-Whitney)