Clinical Evaluation of Bass and Treble Controls for CI Subjects

Clinical Evaluation of Bass and Treble Controls for CI Subjects

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02908737
Acronym
CLASS
Enrollment
34
Registered
2016-09-21
Start date
2016-07-31
Completion date
2016-09-30
Last updated
2018-01-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss

Brief summary

The clinical study is designed to evaluate the safety and effectiveness of recipient controlled Bass and Treble.

Detailed description

Bass and Treble controls provide the CI recipient with access to the adjustment of low and high frequency sound balance, known as Bass and Treble respectively. The clinical study is designed to evaluate the safety and effectiveness of recipient controlled Bass and Treble. Subjects will have access to the Bass and Treble controls during a 4 week take-home period, and will complete questionnaires and diaries to provide descriptive data on the Bass and Treble adjustments made during take home use, and the success of these adjustments. Pre and post intervention speech test results will be used to evaluate the effect of access to Bass and Treble controls on speech perception. The post intervention questionnaire will provide information on the overall ratings of the Bass and Treble controls.

Interventions

DEVICEBass and Treble

Bass and Treble controls are clinician enabled and provide the recipient with access to the adjustment of low and high frequency sound balance, known as Bass and Treble respectively

Sponsors

Cochlear
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
6 Years to 100 Years
Healthy volunteers
Yes

Inclusion criteria

1. Six years of age or older 2. At least 3 months experience with the CP810, CP910 or CP920 sound processor 3. At least 3 months experience with the CI24RE or CI500 series implant 4. Willingness to participate in and to comply with all requirements of the protocol

Exclusion criteria

1. Unrealistic expectations on the part of the subject, regarding the possible benefits, risks and limitations that are inherent to the procedure and prosthetic device 2. Additional disabilities that would prevent participation in evaluations

Design outcomes

Primary

MeasureTime frameDescription
Non-inferior speech perception with preferred Bass and Treble settings compared to the baseline Bass and Treble settings8 weeksCo-primary endpoint #1: This endpoint is to test that speech perception scores in noise with preferred Bass and Treble settings are non-inferior to the baseline settings. Subjects will be tested at visit one with their own settings (Baseline), and then again after four weeks use of the Bass and Treble feature. The speech signal will arrive from zero degrees azimuth at a level of 65 dB SPL, the noise will also arrive from zero degrees azimuth, however the noise will adapt according to subject responses. o-primary endpoint #2: This endpoint is to test that speech perception scores in quiet with preferred Bass and Treble settings are non-inferior to the baseline settings. Subjects will be tested at visit one with their own settings (Baseline), and then again after four weeks use of the Bass and Treble feature. The speech signal will arrive from zero degrees azimuth at a level of 50 dB SPL.

Secondary

MeasureTime frameDescription
Non-inferior sentence recognition in four-talker babble noise with Bass and Treble values set to the minimum (-6 Bass, - 6 Treble) compared to the baseline condition when the speech and noise are from the front8 weeksSecondary endpoint #1: This endpoint is to test that speech perception scores in noise with Bass and Treble settings set to minimum (-6, -6) are non-inferior to the baseline settings. Subjects will be tested at visit one. The speech signal will arrive from zero degrees azimuth at a level of 65 dB SPL, the noise will also arrive from zero degrees azimuth, however the noise will adapt according to subject responses.

Other

MeasureTime frame
Subjective daily diary data on the success of the daily Bass and Treble changes made during take-home use8 weeks
Questionnaire data on the overall subjective ratings of the Bass and Treble controls (importance, confidence, ease and satisfaction)8 weeks
Incidence of adverse events as well as the severity, relationship to treatment, outcome and actions taken will be listed for each subject and summarised.8 weeks