Skip to content

Clinical Comparison of Different Humidification Strategies During Noninvasive Ventilation With Helmet

Clinical Comparison of Different Humidification Strategies During Noninvasive Ventilation With Helmet

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02875379
Enrollment
20
Registered
2016-08-23
Start date
2017-02-01
Completion date
2019-01-23
Last updated
2019-01-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Respiratory Failure

Keywords

Noninvasive Ventilation, Mechanical Ventilation

Brief summary

Background. Non invasive positive pressure ventilation (NIV) is among first line treatments of acute respiratory failure. Several interfaces are available for non-invasive ventilation.Despite full face and oronasal masks are more frequently used, some evidence suggests that helmets may optimize patients' comfort and NIV tolerability. During NIV, humidification strategies (heat and moisture exchangers HME or heated humidifiers HH) may significantly affect patient's comfort and work of breathing. Despite physiological data suggested heated humidification as the best strategy during NIV with full face masks, no differences were found in a randomized controlled study assessing the effects of HME or HH on a pragmatic clinical outcome. However, the higher dead space (i.e. 18 L/min) and rebreathing rate observed during helmet NIV make such results not applicable to this particular setting. The investigators designed a randomized-crossover trial to assess the effect of four humidification strategies during helmet NIV on patients with acute respiratory failure, in terms of comfort, work of breathing and patient-ventilator interaction. Methods. All awake, collaborative, hypoxemic patients requiring mechanical ventilation will be considered for the enrollment. Hypercapnic patients (i.e.PaCO2\>45 mmHg) will be excluded. Each enrolled patient will undergo helmet NIV with all the following humidification strategies in a random order. Each period will last 60 minutes. * Passive humidification, double tube circuit. * Heated humification (MR 730, Fisher & Paykel, Auckland, New Zealand), humidification chamber temperature 33°C. * Heated humification (MR 730, Fisher & Paykel, Auckland, New Zealand), humidification chamber temperature 37°C. * Passive humidification with HME, Y-piece circuit. Ventilatory settings (Draeger Evita xl or Evita infinity ventilators): Pressure support ventilation; pressure support=20 cmH20; FiO2 titrated to obtain SpO2 between 92 and 98%; positive end-expiratory pressure=10 cmH2O; maximum inspiratory time 0.9 seconds; inspiratory flow trigger = 2 l/min; expiratory trigger: 30% of the maximum inspiratory flow; pressurization time=0,00 s. Such settings will be kept unchanged during the whole study period. An oesophageal catheter will be placed and secured to measure oesophageal pressure (Pes) and gastric pressure (Pga) (Nutrivent, Italy): the reliability of the measured pressure will be confirmed with an airway occlusion test during NIV with oronasal mask. Work of breathing will be estimated with the pressure-time product (PTP) of the pleural pressure. A pneumotachograph (KleisTek) will record flow, airway pressure, Pes and Pga on a dedicated laptop. At the end of each cycle, the patient will be asked to rate his/her discomfort on a visual analog scale (VAS) modified for ICU patients. The level of dyspnea will be assessed with the Borg dyspnea scale. The following parameters will be record at the end of each cycle: Arterial pressure, heart rate, respiratory rate, SpO2, pH, PCO2, PaO2, SaO2. Airway and esophageal pressure signals will be reviewed offline to detect patient-ventilator asynchronies (ineffective efforts, double cycling, premature cycling, delayed cycling) and asynchrony index (number of asynchrony events divided by the total respiratory rate computed as the sum of the number of ventilator cycles (triggered or not) and of wasted efforts) will be computed. The trigger delay will be also measured. The pressurization and depressurization velocity will be assessed with the PTP airway index 300 and 500 (inspiratory and expiratory), as suggested by Ferrone and coworkers. The work of breathing (WOB) for each breath will be estimated by PTPes. An hygrometer (Dimar SRL, Italy) will measure and record on a dedicated laptop Helmet temperature, relative and absolute humidity. Primary endpoints: patient's comfort, work of breathing and asynchrony index. Sample Sizing: Given the physiological design of the study, the investigators did not make an a priori sample size and plan to enroll 24 patients.

Detailed description

Background Non invasive positive pressure ventilation (NIV) is among first line treatments of acute respiratory failure. In patients with new-onset respiratory failure, NIV was showed to reduce the rate of complications and the length of ICU stay, as compared to invasive mechanical ventilation\[1\] Several interfaces are available for non-invasive ventilation: full face masks, oronasal masks, nasal prongs and helmets\[2\]. Despite full face and oronasal masks are more frequently used, some evidence suggests that helmets may optimize patients' comfort and NIV tolerability. The helmet allows patients' interaction, speech, feeding and does not limit cough. In addition, skin necrosis, gastric distension, or eye irritation are seldom observed during helmet NIV, while may be consequences of long-term treatments with face masks. \[3\] On the contrary, helmet NIV hampers tidal volume monitoring, is contraindicated in hypercapnic patients and requires specific ventilator settings\[4\]. Lastly, when compared to face masks, helmets may increase the work of breathing and worsen patient-ventilator interaction\[5\]\[6\]\[7\]. During NIV, humidification strategies (heat and moisture exchangers HME or heated humidifiers HH) may significantly affect patient's comfort and work of breathing \[8\]\[9\]. Despite physiological data suggested heated humidification as the best strategy during NIV with full face masks\[8\]\[9\], no differences were found in a randomized controlled study assessing the effects of HME or HH on a pragmatic clinical outcome\[10\]. However, the higher dead space (i.e. 18 L/min) and rebreathing rate observed during helmet NIV make such results not applicable to this particular setting. One only study assessed the effects of a HH during helmet low-flow continuous positive airway pressure on comfort in healthy volunteers\[11\]. Indeed, patients suffering from acute respiratory failure may behave differently, especially in terms of minute ventilation and maximum inspiratory flow. A recent bench study identified a better patient-ventilator interaction when helmet NIV was provided through a double tube circuit, as compared to the Y-piece system \[12\]. The investigators designed a randomized-crossover trial to assess the effect of four humidification strategies during helmet NIV on patients with acute respiratory failure, in terms of comfort, work of breathing and patient-ventilator interaction. Methods Design: monocentric, randomized, cross-over trial. Each enrolled patient will undergo helmet NIV with all the following humidification strategies in a random order. Each period will last 60 minutes. * Passive humidification, double tube circuit. * Heated humification (MR 730, Fisher & Paykel, Auckland, New Zealand), humidification chamber temperature 33°C. * Heated humification (MR 730, Fisher & Paykel, Auckland, New Zealand), humidification chamber temperature 37°C. * Passive humidification with HME, Y-piece circuit. Ventilatory settings (Draeger Evita xl or Evita infinity ventilators): Pressure support ventilation; pressure support=20 cmH20\[4\]; FiO2 titrated to obtain SpO2 between 92 and 98%; positive end-expiratory pressure=10 cmH2O\[4\]; maximum inspiratory time 0.9 seconds; inspiratory flow trigger = 2 l/min; expiratory trigger: 30% of the maximum inspiratory flow; pressurization time=0,00 s. Such settings will be kept unchanged during the whole study period. An oesophageal catheter will be placed and secured to measure oesophageal pressure (Pes) and gastric pressure (Pga) (Nutrivent, Italy): the reliability of the measured pressure will be confirmed with an airway occlusion test during NIV with oronasal mask\[13\]. Work of breathing will be estimated with the pressure-time product (PTP) of the pleural pressure\[13\]. A pneumotachograph (KleisTek) will record flow, airway pressure, Pes and Pga on a dedicated laptop. At the end of each cycle, the patient will be asked to rate his/her discomfort on a visual analog scale (VAS) modified for ICU patients. The level of dyspnea will be assessed with the Borg dyspnea scale\[14\]. The following parameters will be record at the end of each cycle: Arterial pressure, heart rate, respiratory rate, SpO2, pH, PCO2, PaO2, SaO2. Airway and esophageal pressure signals will be reviewed offline to detect patient-ventilator asynchronies (ineffective efforts, double cycling, premature cycling, delayed cycling) and asynchrony index (number of asynchrony events divided by the total respiratory rate computed as the sum of the number of ventilator cycles (triggered or not) and of wasted efforts) will be computed\[15\]. The trigger delay will be also measured. The pressurization and depressurization velocity will be assessed with the PTP airway index 300 and 500 (inspiratory and expiratory), as suggested by Ferrone and coworkers\[12\]. The work of breathing (WOB) for each breath will be estimated by PTPes. An hygrometer (Dimar SRL, Italy) will measure and record on a dedicated laptop Helmet temperature, relative and absolute humidity. End point: Primary endpoints: patient's comfort, work of breathing and asynchrony index. Sample Sizing: Given the physiological design of the study, the investigators did not make an a priori sample size and planned to enroll 24 patients. Statistical analysis Qualitative data will be expressed as number of events (%) and continuous data as mean ± standard deviation or median \[Interquartile range\]. Comparisons concerning qualitative variables will be performed with the Mc-Namar test. Ordinal qualitative variables or non normal quantitative variables will be compared with the Friedman's Test, the wilcoxon sum of ranks test or the Mann-Whitney test, as appropriate. All analysis will be performed applying a bilateral hypothesis. P ≤ 0.05 will be considered significant. Statistical analysis will be performed with SPSS 20.0.

Interventions

PROCEDUREMeasurements

Measurements of respiratory mechanics and parameters, arterial blood gases and comfort

Sponsors

Catholic University of the Sacred Heart
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Awake and collaborative patients * Age\>18 years * Need for noninvasive mechanical ventilation * Informed consent

Exclusion criteria

* Cardiopulmonary resuscitation * Haemodynamic instability * Coma * Asma * Hypercapnia (paCO2\>45 mmHg) * Recent gastric or abdominal surgery

Design outcomes

Primary

MeasureTime frameDescription
Comfort assessed by visual analogic scale modified for ICU patientsAt the end of each 1-hour ventilation periodPatient's comfort, assessed by visual analogic scale modified for ICU patients
Patient-ventilator asynchrony. Asynchrony indexAt the end of each 1-hour ventilation periodAsynchrony index number of asynchrony events divided by the total respiratory rate computed as the sum of the number of ventilator cycles (triggered or not) and of wasted efforts. Inspiratory trigger delay (time between the onset of patient's effort and ventilatory support). Pressurization and depressurization efficacy.
Work of breathing. Oesophageal pressure time productAt the end of each 1-hour ventilation periodPressure time product of the esophageal pressure (PTPes) and pressure time product of the transdiaphragmatic pressure (PTPdi)

Secondary

MeasureTime frameDescription
Helmet humidityAt the end of each 1-hour ventilation period
PaO2At the end of each 1-hour ventilation period
PaCO2At the end of each 1-hour ventilation period
Helmet temperatureAt the end of each 1-hour ventilation period
respiratory rateAt the end of each 1-hour ventilation period
DyspneaAt the end of each 1-hour ventilation periodBorg dyspnea score

Countries

Italy

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 20, 2026