Blood Pressure
Conditions
Keywords
non-invasive blood pressure
Brief summary
The goal of the study is to test the capability of the CareTaker monitor to calibrate its continuous blood pressure readings independently. Currently the device requires another approved blood pressure monitor to provide a starting calibration. The new control module enables the CareTaker to perform a pressure sweep of the internal pressure in the finger cuff. The resulting data is analyzed using a combination of pulse analysis/oscillometry approach.
Detailed description
The specific objective of this project was to demonstrate the CareTaker's ability to perform self-calibration to the same performance standard as its predicate device. That device demonstrated efficacy by showing substantial equivalence to the performance of classical sphygmomanometry using a Riva-Rocci/Korotkoff (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg across 3 paired readings from at least 85 patients, for a total of at least 255 paired readings. The study was performed at the University of Virginia Medical Center in Charlottesville, Virginia. UVA's Institutional Review Board approved (#18686) and supervised the study for subjects who were University of Virginia hospital patients and staff \> 18 years of age and who were able to give verbal consent.
Interventions
Equivalence of CareTaker blood pressure to classical sphygmomanometry
Sponsors
Study design
Eligibility
Inclusion criteria
* \> 18 years of age, who are able and willing to participate and have given verbal assent
Exclusion criteria
* Unable to give verbal assent * \<18 years of age * No or poor finger pulse, as determined through visual inspection for ischemic hands
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Performance comparison | 8 minutes for 3 paired readings | Comparison with performance of classical sphygmomanometry using a Riva-Rocci/Korotkov (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg |