Obstructive Sleep Apnea (OSA)
Conditions
Brief summary
Subjects will all have a two week baseline period in which they will use a nasal mask using an air view CPAP machine (which records all data). After the two week period all will switch to a full face mask with half using the same CPAP pressure and half with a new a CPAP pressure derived from our formula for the final two weeks.
Detailed description
The goal of this study is to test the efficacy of an algorithm the investigators previously created to adjust Continuous positive airway pressure (CPAP) when a mask interface is switched from a nasal to a full face. Subjects will currently have a moderate to severe Apnea-hypopnea index (AHI - The Apnea-Hypopnea Index or Apnoea-Hypopnoea Index (AHI) is an index used to indicate the severity of sleep apnea. It is represented by the number of apnea and hypopnea events per hour of sleep). (above 15) * Scoring of apneas Score a respiratory event as an apnea when BOTH of the following criteria are met: N1, N2, N3, N4 1. There is a drop in the peak signal excursion by ≥90% of pre-event baseline using an oronasal thermal sensor (diagnostic study), PAP device flow (titration study) or an alternative apnea sensor (diagnostic study) 2. The duration of the ≥90% drop in sensor signal is ≥ 10 seconds. Score an apnea as obstructive if it meets apnea criteria and is associated with absent inspiratory effort throughout the entire period of absent airflow. Score an apnea as central if it meets apnea criteria and is associated with absent inspiratory effort throughout the entire period of absent airflow. Score an apnea as mixed if it meets apnea criteria and is associated with absent inspiratory effort in the initial portion of the event, followed by resumption of inspiratory effort in the second portion of the event. N4 Note 1 - Identification of an apnea does not require a minimum desaturation criterion. Note 2 - If a portion of a respiratory event that would otherwise meet criteria for a hypopnea meets criteria for apnea, the entire event should be scored an apnea. Note 3 - If the apnea or hypopnea event begins or ends during an epoch that is scored as sleep, then the corresponding respiratory event can be scored and included in the computation of the apnea hypopnea index (AHI). This situation usually occurs when an individual; has a high AHI with events occurring so frequently that sleep is severely disrupted and epochs may end up being scored as wake even though \<15 seconds of sleep is present during the epoch containing that portion of the respiratory event. However if the apnea or hypopnea occurs entirely during an epoch scored as wake, it should not be scored or counted towards the apnea hypopnea index because of the difficultly of defining a denominator in this situation. If these occurrences are a prominent feature of the polysomnogram and/or interfere with sleep onset, their presence should be mentioned in the narrative summary if the study. Note 4 - For alternative apnea sensors see tech specifications for Adults. There is not sufficient evidence to support a specific durations of the central and obstructive components of a mixed apnea thus specific durations of these components are not recommended. * AASM 2015 PROTOCOL Scoring of Hypopneas Score a respiratory event as a hypopnea if all of the following criteria are met. N1, N2, N3. 1. The peak signal excursions drop by ≥30% of pre-event baseline using nasal pressure (diagnostic study), PAP device flow (titration study), or an alternative hypopnea sensor (diagnostic study). 2. The duration of the ≥30% drop in signal excursion is ≥10 seconds. 3. There is a ≥3% oxygen desaturation from pre-event baseline or the event is associated with an arousal. Subjects once consented and entered into the study, will all have a two week baseline period in which they will use a nasal mask using an airview cpap machine (which records all data). After the two week period all will switch to a full face mask with half using the same CPAP pressure and half with a new cpap pressure derived from our formula for the final two weeks. Those placed in the randomized into a new pressure will have an experimental procedure of getting a new pressure. The investigators expect the AHI of the controls to increase. Our two new formulas are referred to as Opred (oronasal) (full face mask) and Npred (nasal) (non-full face mask). Npred=(0.017 AHI)-(0.092 LSaO2)+(0.225 NC)+5.534 Opred=(0.03 AHI)-(0.130 LSaO2)+19.732 NC=Neck circumference, AHI= baseline AHI, LSaO2= Lowest blood oxygen saturation from the baseline polysomnogram. These models were developed using multiple linear regressing modeling. The study is self-funded because it has no funding. Subjects will not be compensated
Interventions
Full Face Mask with same CPAP Pressure
New cpap pressure derived from our formula for the final two weeks.
Sponsors
Study design
Intervention model description
Subjects will complete two week baseline with nasal mask then will complete two week treatment with full face either with new algorithm or old algorithm.
Eligibility
Inclusion criteria
* AHI above 15. * Willing to switch to full face mask post titration. * Not currently on a weight loss plan and no intention of beginning a weight loss regimen during the duration of the study
Exclusion criteria
* AHI below 15. * Not willing to switch to full face mask post titration. * Currently on a weight loss plan and / or intention of beginning a weight loss regimen during the duration of the study.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Change in Severity of Sleep Apnea as Measured by the AHI | After a 2-week baseline period (for weeks 3 & 4), subjects will be switched from a nasal to full-face CPAP mask interface for two-weeks with either the same CPAP setting (control) or a new formula determined pressure (treatment). | The investigators will compare the Apnoea-Hypopnoea Index (AHI) of subjects with the new formula in place vs. those that don't have the formula. The Apnea-Hypopnea Index is an index used to indicate the severity of sleep apnea. It is represented by the number of apnea and hypopnea events per hour of sleep. A respiratory event is scored as an apnea when both of the following criteria are met: N1, N2, N3, N4. a) There is a drop in the peak signal excursion by greater than or equal to 90% of pre-event baseline using an oronasal thermal sensor (diagnostic study), PAP device flow (titration study) or an alternative apnea sensor (diagnostic study) b) The duration of the greater than or equal to 90% drop in sensor signal is greater than or equal to 10 seconds. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Lowest Oxyhemoglobin Saturation | One month | The investigators will compare the lowest oxyhemoglobin saturation of subjects with the new formula in place vs. those that don't have the formula. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Full Face Mask With Same CPAP Pressure After the two week period all will switch to a full face mask with half using the same CPAP pressure.
Full Face Mask with same CPAP Pressure: Full Face Mask with same CPAP Pressure | 6 |
| Full Face Mask With New Pressure After the two week period all will switch to a full face mask with half using with a new cpap pressure derived from our formula.
Full Face Mask with new Pressure: New cpap pressure derived from our formula for the final two weeks. | 5 |
| Total | 11 |
Baseline characteristics
| Characteristic | Full Face Mask With Same CPAP Pressure | Full Face Mask With New Pressure | Total |
|---|---|---|---|
| Age, Categorical <=18 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical >=65 years | 4 Participants | 4 Participants | 8 Participants |
| Age, Categorical Between 18 and 65 years | 2 Participants | 1 Participants | 3 Participants |
| Ethnicity (NIH/OMB) Hispanic or Latino | 1 Participants | 0 Participants | 1 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 5 Participants | 5 Participants | 10 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 0 Participants | 0 Participants | 0 Participants |
| Region of Enrollment United States | 6 participants | 5 participants | 11 participants |
| Sex: Female, Male Female | 4 Participants | 3 Participants | 7 Participants |
| Sex: Female, Male Male | 2 Participants | 2 Participants | 4 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 0 / 6 | 0 / 5 |
| other Total, other adverse events | 0 / 6 | 0 / 5 |
| serious Total, serious adverse events | 0 / 6 | 0 / 5 |
Outcome results
Change in Severity of Sleep Apnea as Measured by the AHI
The investigators will compare the Apnoea-Hypopnoea Index (AHI) of subjects with the new formula in place vs. those that don't have the formula. The Apnea-Hypopnea Index is an index used to indicate the severity of sleep apnea. It is represented by the number of apnea and hypopnea events per hour of sleep. A respiratory event is scored as an apnea when both of the following criteria are met: N1, N2, N3, N4. a) There is a drop in the peak signal excursion by greater than or equal to 90% of pre-event baseline using an oronasal thermal sensor (diagnostic study), PAP device flow (titration study) or an alternative apnea sensor (diagnostic study) b) The duration of the greater than or equal to 90% drop in sensor signal is greater than or equal to 10 seconds.
Time frame: After a 2-week baseline period (for weeks 3 & 4), subjects will be switched from a nasal to full-face CPAP mask interface for two-weeks with either the same CPAP setting (control) or a new formula determined pressure (treatment).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Full Face Mask With Same CPAP Pressure | Change in Severity of Sleep Apnea as Measured by the AHI | 2.81 AHI | Standard Deviation 1.59 |
| Full Face Mask With New Pressure | Change in Severity of Sleep Apnea as Measured by the AHI | 1.59 AHI | Standard Deviation 1.47 |
Lowest Oxyhemoglobin Saturation
The investigators will compare the lowest oxyhemoglobin saturation of subjects with the new formula in place vs. those that don't have the formula.
Time frame: One month
Population: We were unable to collect oximetry data in either the treatment or control groups.