Servo-Ventilation In-lab PSG Evaluation

Servo-Ventilation In-lab Polysomnography (PSG) Evaluation

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02808338
Enrollment
25
Registered
2016-06-21
Start date
2015-11-09
Completion date
2017-07-06
Last updated
2019-03-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Sleep Apnea Syndromes

Brief summary

Continuous Positive Airway Pressure (CPAP) is the most effective treatment for the Obstructive Sleep Apnea-Hypopnea Syndrome (OSAHS). CPAP stabilizes the airway and prevents instability and collapse. With a stable and patent airway, breathing continues in a normal manner, gas exchange is improved, and there is no disruption of sleep related to disturbed breathing. Auto Servo Ventilation (Auto SV) is a mode of positive airway pressure used to treat obstructive and complex central sleep apnea. The main features of the Auto SV mode include: * Normalization of ventilation by automatically adjusting Inspiratory Positive Airway Pressure (IPAP) pressure to achieve a target ventilation. IPAP is increased or decreased to help stabilize the ventilation. * Provision of timed, back-up breaths during central apneas. The optimal back-up rate is automatically determined by the device based on the patient's breathing. * Automatic control of Expiratory Positive Airway Pressure (EPAP) to treat obstructive events. Several manufacturers produce these types of devices. The algorithms used to determine the IPAP, EPAP and minimum respiratory rate are different. The largest number of these devices currently in use are the BiPAP AutoSV Advanced System One (Philips Respironics, Murrysville PA) and the Variable positive airway pressure (VPAP) Adapt (ResMed Corp., San Diego CA).

Detailed description

This study is to better understand the performance and features of the BiPAP autoSV Advanced System One. Philips Respironics will evaluate the PolySomnoGraphy (PSG) and Encore data from each night. The aim of the study is to characterize the acute outcomes of treatment provided by the device and evaluate any modifications made to the Philips Respironics device's algorithm The participants for this study will be experienced in having used servo ventilation therapy at their home because it is only such patients who are already using servo ventilation as part of routine clinical care that are eligible for participation. Participants will receive four (4) randomized PSG's during which they will receive treatment from the following devices in a randomized manner: * FDA released Philips BiPAP AutoSV Advanced System One * A Modified Philips BiPAP ASV * FDA released ResMed S7 VPAP Adapt * FDA released ResMed S9 VPAP Adapt The participants sought for this effort will be previously prescribed a servo-ventilation device as part of routine clinical care. Baseline with PSG * Informed Consent * Inclusion/Exclusion Criteria Review * Demographics * Anthropometric Measurements * PAP Prescription information (if available; from Device or from Medical records). This is the pressure settings of the device that was prescribed by the patient's clinical sleep medicine provider. * Medical History and physical examination * Sleep History- Including that past 30 day detailed report * Diagnostic PSG history (copies of sleep studies that were performed as part of routine clinical care in the past that qualified the patient for the servo-ventilation device) * Current Medications * Vital Signs (at beginning of PSG night) * Research Trial PSG- Randomized to a device Procedures for Each PSG after Baseline * Current Medications * Vital Signs (at beginning of PSG night) * Research Trial PSG- Randomized to one of the four devices (4 research PSGs will be performed for each subject). 30 day Take Home with Modified Philips BiPAP ASV device After the last PSG participants will be sent home on Modified Philips Auto Servo Ventilation (ASV) study device. A wireless modem with oxygen saturation level (SPO2) connection will be connected to the Modified Philips ASV study device. The wireless modem with SPO2 connection is for device data collection/transmission purposes only that monitors the usage and performance of the device in the home-setting. * The device setting should be set to the following: * P max: 30 * EPAP min: 4 * Expiratory Positive Airway Pressure maximum (EPAPmax):15 * Pressure Support Minimum (PS min): 0 * Pressure Support Maximum (PS max): 15 * BiFlex (Bi-Level Flex): 2 * Rate: Auto After 30 day Take Home Participants Will * Return to the Sleep Lab * Complete the end of study questionnaire * Return all of the study equipment Additional Take Home with the Modified Philips BiPAP ASV device: * Participants may be asked to use the Modified Philips ASV study device for an additional 30 days if more data is needed for analysis. Participants will complete the same end of day questionnaire noted above. After trial completion: * When the Participant is done with the trial they will go back to using their own prescribed device.

Interventions

DEVICEPhilips BiPAP AutoSV Advanced System One

Auto SV Auto Servo Ventilation is a mode of positive airway pressure used to treat obstructive and complex central sleep apnea. This devices is FDA approved and will be set according to a predetermined setting.

DEVICEModified Philips BiPAP ASV

This Modified BiPAP ASV will be set to a predetermined setting. This is the investigational device.

DEVICEResMed S7 VPAP Adapt

This device is FDA approved and will be set to predetermined setting.

DEVICEResMed S9 VPAP Adapt

This device is FDA approved and will be set to predetermined setting.

Sponsors

Philips Respironics
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Ability to provide consent * Age ≥ 21 * Currently prescribed servo ventilation therapy at home * At least two weeks of recent adherence and efficacy data from PAP device demonstrating adequate use of therapy (at least 4 hours of use per night and use on at least 9 of 14 nights) * Current device compliance report demonstrating residual Apnea-Hypopnea Index (AHI) of 5 or more

Exclusion criteria

* Participants who are acutely ill, medically complicated or who are medically unstable * Participants in whom PAP therapy is otherwise medically contraindicated * Participants who are claustrophobic * Symptomatic (Symptomatic defined as hospitalized for heart failure or a change in cardiac medications, within the last two months) chronic heart failure (NYHA 2-4) and reduced LVEF≤45%, AND moderate to severe predominant central sleep apnea * Participants with previously diagnosed respiratory failure or respiratory insufficiency and who are known to have elevated arterial carbon dioxide levels while awake (PaCO2 ≥ 55mmHg). * Participants requiring any kind of oxygen therapy * Participants who have had surgery of the upper airway, nose, sinus, eyes, or middle ear within the previous 90 days * Participants with untreated, non-OSA sleep disorders, including but not limited to; insomnia, periodic limb movement syndrome, or restless legs syndrome (PLMI \> 10).

Design outcomes

Primary

MeasureTime frameDescription
Average Apnea Hypopnea Index Among Different Polysomnography Devices.4 nightsThe Apnea Hypopnea Index is the number of the apneas and hypopneas during one hour of sleep. The average of these events over one night were compared during each overnight device use.

Countries

United States

Participant flow

Recruitment details

A multi-site study that accrued over 2016 and into the middle of 2017.

Participants by arm

ArmCount
Screening
All participants that consented to the study.
20
Total20

Withdrawals & dropouts

PeriodReasonFG000
Overnight In-lab PolysomonographyReason not given1
Overnight In-lab PolysomonographyWithdrawal by Subject1
ScreeningScreen Failure10

Baseline characteristics

CharacteristicScreening
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
10 Participants
Age, Categorical
Between 18 and 65 years
10 Participants
Age, Continuous65.3 years
STANDARD_DEVIATION 12.6
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
20 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
20 Participants
Region of Enrollment
United States
20 participants
Sex: Female, Male
Female
3 Participants
Sex: Female, Male
Male
17 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
EG004
affected / at risk
deaths
Total, all-cause mortality
0 / 120 / 130 / 130 / 120 / 13
other
Total, other adverse events
1 / 120 / 133 / 130 / 121 / 13
serious
Total, serious adverse events
0 / 120 / 130 / 130 / 120 / 13

Outcome results

Primary

Average Apnea Hypopnea Index Among Different Polysomnography Devices.

The Apnea Hypopnea Index is the number of the apneas and hypopneas during one hour of sleep. The average of these events over one night were compared during each overnight device use.

Time frame: 4 nights

Population: 1 participant was excluded because he did not meet the criteria to receive ASV.

ArmMeasureValue (MEAN)Dispersion
Philips BiPAP AutoSV Advanced System OneAverage Apnea Hypopnea Index Among Different Polysomnography Devices.18.4 events per hourStandard Deviation 23
Modified Philips BiPAP ASVAverage Apnea Hypopnea Index Among Different Polysomnography Devices.25.4 events per hourStandard Deviation 25
ResMed S7 VPAP AdaptAverage Apnea Hypopnea Index Among Different Polysomnography Devices.24.1 events per hourStandard Deviation 25.4
ResMed S9 VPAP AdaptAverage Apnea Hypopnea Index Among Different Polysomnography Devices.19.2 events per hourStandard Deviation 23.1