Clinical Investigation iDesign System With 1.3-PRESBY Treatment and STAR S4 IR™ Excimer Laser System

A Prospective Study to Evaluate the Feasibility of a New Treatment Algorithm to Increase Depth of Focus After Wavefront-guided LASIK Correction of Myopic Refractive Errors With the iDesign System and Star S4 IR™ Excimer Laser System

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02806726
Enrollment
42
Registered
2016-06-21
Start date
2016-07-13
Completion date
2017-11-09
Last updated
2025-02-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Myopia, Presbyopia

Brief summary

A prospective study to evaluate the feasibility of a new treatment algorithm to increase depth of focus after wavefront-guided LASIK correction of myopic refractive errors with the iDesign Advanced Wavescan Studio™ System and Star S4 IR™ Excimer Laser System. The purpose of this study is to determine whether wavefront-guided LASIK correction of myopic refractive errors with CustomVue combined with iDesign 1.3-PRESBY (presbyT-LASIK) treatments mitigate the effects of presbyopia by increasing the depth of focus compared to iDesign CustomVue treatments.

Interventions

DEVICEiDesign 1.3-PRESBY treatment

Software used to calculate the LASIK treatment profile including a presbyopic correction component for the iDesign Advanced Wavescan Studio™ System within the Star S4 IR™ Excimer Laser System.

DEVICEiDesign 1.3 treatment

Software used to calculate the LASIK treatment profile for the iDesign Advanced Wavescan Studio™ System within the Star S4 IR™ Excimer Laser System.

Sponsors

Abbott Medical Optics
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
45 Years to No maximum
Healthy volunteers
No

Inclusion criteria

All criteria apply to each eye 1. Signed informed consent and Personal Health Information Protection Act authorization. 2. At least 45 years of age at enrollment (date informed consent signed). 3. The refractive error, based on the iDesign displayed refraction selected for treatment (4.0 Rx calc at 12.5 mm), must be myopia with or without astigmatism with sphere up to -6.00 D, and cylinder between 0.00 D and -5.00 D with a maximum spherical equivalent (SE) of -8.00 D. 4. Require an add power of +1.00 D or more during near testing at 40 cm. 5. Anticipated stromal bed thickness of at least 250 microns based on preoperative central corneal pachymetry minus the maximum ablation depth (as calculated by the iDesign system) plus the intended flap thickness. 6. Distance Best Spectacle Corrected Visual Acuity (BSCVA) of 20/20 or better. 7. Uncorrected Visual Acuity (UCVA) of 20/40 or worse. 8. Less than or equal to 0.75 D difference between cycloplegic and manifest refraction sphere. 9. A stable refractive error (based on a previous exam, medical records, lensometry, or prescription at least 12 months prior to the preoperative manifest refraction), as defined by a change of ≤1.00 D in MRSE. 10. Any eye with a history of contact lens wear within the last 4 weeks must demonstrate refractive stability according to the following: 1. Rigid contact lenses (toric or spherical) must be removed for at least 4 weeks and soft contact lenses (toric or spherical) for at least 2 weeks prior to the first refraction used to establish stability. 2. Two consecutive refractions and keratometric readings must be conducted at least 7 days apart. 3. Refractive stability is defined as a change of not more than 0.50 D in manifest refraction sphere and cylinder as well as mean keratometry between measurements. 4. If the subject/eye meets the refractive stability criteria, contact lens wear is not permitted prior to surgery. 11. Agreement between manifest refraction (adjusted for optical infinity) and iDesign System refraction chosen for treatment, as follows: 1. Spherical Equivalent: Magnitude of the difference is less than 0.625 D. 2. Cylinder: Magnitude of the difference is less than or equal to 0.5 D. 3. Cylinder Axis Tolerance: If either the manifest cylinder entered into the iDesign System or the iDesign cylinder selected for treatment is less than 0.5 D, there is no requirement for axis tolerance. When both cylinders have a magnitude of at least 0.5 D, the axis tolerance as determined by the iDesign system is linearly reduced from 15 (0.5 D) to 7.5 (7.0 D) based on the average magnitude of both cylinders. Note: If the axis tolerance is not in the calculated range, the iDesign system will produce a warning and this exam may not be used for treatment planning. 12. Willing and capable of complying with follow-up examinations for the duration of the study.

Exclusion criteria

1. Women who are pregnant, breast-feeding, intend to become pregnant, or are not using an adequate method of birth control \[examples are any form of barrier contraception (such as condom or diaphragm with contraceptive cream/jelly), birth control pills, hormonal implant, IUD, abstinence or surgical sterilization (tubal ligation, hysterectomy or vasectomy)\]. Note: Women who were pregnant or nursing may not be enrolled until 6 months after either delivery or have stopped nursing and there is documented refractive stability. 2. Concurrent use of systemic (including inhaled) medications that may impair healing, including but not limited to: antimetabolites, isotretinoin (Accutane®) within 6 months of treatment, and amiodarone hydrochloride (Cordarone®) within 12 months of treatment. NOTE: The use of inhaled or systemic corticosteroids, whether chronic or acute, is deemed to adversely affect healing and subjects using such medications are specifically excluded from eligibility. 3. History of any of the following medical conditions, or any other condition that could affect wound healing: collagen vascular disease, autoimmune disease, immunodeficiency diseases, ocular herpes zoster or herpes simplex, endocrine disorders (including, but not limited to unstable thyroid disorders and diabetes), lupus, and rheumatoid arthritis. NOTE: The presence of diabetes (either type 1 or 2), regardless of disease duration, severity, or control, will specifically exclude subjects from eligibility. 4. Subjects with a cardiac pacemaker, implanted defibrillator or other implanted electronic device. 5. History of prior intraocular or corneal surgery (including cataract extraction), active ophthalmic disease or abnormality (including, but not limited to, symptomatic blepharitis, recurrent corneal erosion, dry eye syndrome, neovascularization \> 1 mm from limbus), retinal detachment/repair, clinically significant lens opacity, clinical evidence of trauma, corneal opacity within the central 9 mm and visible on topography, at risk for developing strabismus, or with evidence of glaucoma or propensity for narrow angle glaucoma. NOTE: Subjects with open angle glaucoma, regardless of medication regimen or control, or an IOP greater than 21 mmHg at screening, are specifically excluded from eligibility. 6. Evidence of keratoconus, corneal dystrophy or irregularity, or abnormal topography. 7. Known sensitivity or inappropriate responsiveness to any of the medications used in the postoperative course. 8. Desire to have monovision. 9. Intolerance to multifocal correction based on questionnaire responses to contact lens trial. 10. Participation in any other clinical study.

Design outcomes

Primary

MeasureTime frameDescription
Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 67 cm6 monthsMean distance corrected intermediate visual acuity (DCIVA) for presbyT-LASIK treatments (test) was compared to iDesign CustomVue treatments (control).

Secondary

MeasureTime frameDescription
Monocular Distance Corrected Near Visual Acuity (DCNVA) AT 40 cm6 monthsMean distance corrected near visual acuity (DCNVA) for PresbyT-LASIK treatments (test) was compared to iDesign CustomVue treatments (control).

Participant flow

Pre-assignment details

Per the protocol, the total number of participants enrolled is 42 (the number of participants who agreed to participate in the study and completed the informed consent process). The total number of participants Started is 7 (the number of participants who were assigned to each Arm/Group).

Participants by arm

ArmCount
All Study Participants
Parts of a participant were allocated to different interventions: 1. Left side iDesign 1.3-PRESBY, Right side iDesign 1.3 2. Left Side iDesign 1.3, Right Side iDesign 1.3-PRESBY
7
All Study Participants
Parts of a participant were allocated to different interventions: 1. Left side iDesign 1.3-PRESBY, Right side iDesign 1.3 2. Left Side iDesign 1.3, Right Side iDesign 1.3-PRESBY
14
Total21

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyLost to Follow-up1

Baseline characteristics

CharacteristicAll Study Participants
Age, Continuous49.2 Years
STANDARD_DEVIATION 3.3
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
1 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
6 Participants
Sex: Female, Male
Female
4 Participants
Sex: Female, Male
Male
3 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 7
other
Total, other adverse events
2 / 7
serious
Total, serious adverse events
0 / 7

Outcome results

Primary

Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 67 cm

Mean distance corrected intermediate visual acuity (DCIVA) for presbyT-LASIK treatments (test) was compared to iDesign CustomVue treatments (control).

Time frame: 6 months

Population: Due to the small sample size, descriptive results for all endpoints are provided along with confidence intervals, and no statistical conclusions are made.

ArmMeasureValue (MEAN)
iDesign 1.3-Presby Treatment (Experimental)Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 67 cm0.03 LogMAR
iDesign CustomVue Treatments (Control)Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 67 cm0.1 LogMAR
Secondary

Monocular Distance Corrected Near Visual Acuity (DCNVA) AT 40 cm

Mean distance corrected near visual acuity (DCNVA) for PresbyT-LASIK treatments (test) was compared to iDesign CustomVue treatments (control).

Time frame: 6 months

Population: Due to the small sample size, descriptive results for all endpoints are provided along with confidence intervals, and no statistical conclusions are made.

ArmMeasureValue (MEAN)
iDesign 1.3-Presby Treatment (Experimental)Monocular Distance Corrected Near Visual Acuity (DCNVA) AT 40 cm0.25 LogMar
iDesign CustomVue Treatments (Control)Monocular Distance Corrected Near Visual Acuity (DCNVA) AT 40 cm0.35 LogMar

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026