Skip to content

Clinical Evaluation of Investigational Contact Lenses

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02801396
Enrollment
116
Registered
2016-06-15
Start date
2016-04-24
Completion date
2016-05-22
Last updated
2017-07-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Visual Acuity

Brief summary

Single-visit, unmasked, non-dispensing study where subjects will wear three different lenses bilaterally in one visit, each lens being worn for approximately 30-60 minutes. Subjects will complete a questionnaire for each study lens.

Interventions

DEVICEEtafilcon A Design 1 (Test 1)
DEVICEEtafilcon A Design 2 (Test 2)

Sponsors

Johnson & Johnson Vision Care, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 55 Years
Healthy volunteers
Yes

Inclusion criteria

1. Subjects must be at least 18 years of age and no more than 55 years of age (inclusive). 2. The subject's vertex corrected best spherical distance refraction must be in the range of -1.00 to -6.00 Diopters (D) in each eye. 3. The subject's vertex corrected cylindrical component of distance refraction must be less than or equal to -1.50 Diopters of Cylinder (DC) in each eye. 4. Subjects must have best corrected visual acuity of 20/25 (Snellen or equivalent) or better in each eye. 5. Subjects must self-report as being Asian race. 6. Subjects must self-report as having brown or black eyes. 7. Subjects must be habitual lens wearer (clear or limbal ring cosmetic) . Habitual is defined as at least one (1) month of Contact Lens wear where the lenses were worn for a minimum of six (6) hours per day and a minimum of three (3) days per week. 8. Subjects between 18 to 39 years old must be habitual limbal ring cosmetic lens wearers or must have tried limbal ring cosmetic lens within the past two years. 9. Subjects must be cosmetic contact lens concept acceptors as determined by screening questionnaire. 10. Subjects must possess a functional / usable pair of spectacles and bring them to the visit (only if applicable - to the investigator's discretion). 11. Subjects must read, understand, and sign the Statement of Informed Consent. 12. Subjects must appear able and willing to adhere to the instructions set forth in the clinical protocol. 13. Subjects must be willing to have their eyes video recorded and photographed.

Exclusion criteria

1. Working for any advertising agency, any company involved in public relations or marketing, market research; or any company that makes or sells contact lenses or glasses. 2. Participated in any paid market research within the past 3 months. 3. Pregnant or breast feeding a baby. 4. Subjects with diabetes. 5. Any ocular or systemic allergies or disease which might interfere with contact lens wear. 6. Any systemic disease, or autoimmune disease, which might interfere with contact lens wear, at investigator's discretion. 7. Using medications during and immediately preceding the study period that may interfere with current contact lens wear as determined by the investigators 8. Any infectious diseases (e.g. hepatitis, tuberculosis) or a contagious immunosuppressive disease (e.g. HIV), by self-report. 9. Any grade 3 or greater Biomicroscopy findings (this includes, corneal edema, corneal staining, corneal vascularization, conjunctival injection, tarsal abnormalities, bulbar injection) on the FDA classification scale. 10. Any active ocular abnormalities/conditions that might interfere with contact lens wear (this includes, but not limited to, chalazia, recurrent styles, pterygium, infection etc.). 11. Any corneal distortion, scar, or opacity affecting vision as noted by subjects or investigators during examination. 12. History of any corneal surgery (e.g. radial keratotomy (RK), photorefractive keratectomy (PRK), laser-assisted in situ keratomileusis (LASIK)); confirmed with topography if necessary at the investigator's discretion. 13. Habitual wearers of rigid gas permeable or orthokeratology lens within the past 3 months. 14. Current habitual contact lens modality for extended wear or conventional wear (contact lenses with a replacement schedule of 3 months to \>1 year). 15. Participation in any pharmaceutical, medical device, or contact lens related clinical research trial within 14 days prior to study enrollment. This excludes studies that do not require product treatment/intervention.

Design outcomes

Primary

MeasureTime frameDescription
CLUE Handling30 Minutes Post InsertionCLUE Handling is assessed using the Contact Lens User Experience (CLUE™) Questionnaire. CLUE™ is a validated patient-reported outcomes questionnaire to assess patient experience attributes of soft, disposable contact lenses (comfort, vision, handling, and packaging) in a contact lens-wearing population in the US, ages 18-65. Scores follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable / positive response. 97% of the scores fall within 0 and 120 (mean +/- 3XSD).
Visual Performance (LogMar)30 Minutes Post InsertionDistance time controlled Visual Performance (LogMAR) was assessed for each subject eye under bright-illumination high-contrast lighting conditions at 4m using an ETDRS chart. The average visual performance (LogMAR) was reported for each study lens.
Percentage of Eyes With Acceptable Mechanical Lens Fit30 Minutes Post InsertionMechanical lens fit will be assessed for each subject and eye using a slit lamp. Lens fit is a binary response and 'Yes' = Acceptable Fit and 'No' = Unacceptable Fit. Lens fit is assessed using Lens centration, limbal exposure, primary gaze movement, up-gaze movement, edge lift, lens tightness. Unacceptable fit will be declared if there is any of the following present: limbal exposure, edge lift, excessive movement in primary or up-gaze or insufficient movement in primary gaze and up-gaze. The Percentage of subject eyes with acceptable lens fit will be reported.
Percentage of Eyes With Acceptable Cosmetic Lens Fit30 Minutes Post InsertionCosmetic lens fitting is assessed for each subject eye in the primary gaze position without a slit lamp. Cosmetic lens fit is a binary response and is reported as acceptable or unacceptable. The Percentage of subject eyes with acceptable cosmetic lens fitting is reported.

Countries

Hong Kong

Participant flow

Pre-assignment details

A total of 116 subjects were enrolled into the study. Of the enrolled subjects 15 were not dispensed a study lens. A total of 101 subjects were dispensed a study lens. Of the dispensed subjects 1 was discontinued and 100 subjects completed the study.

Participants by arm

ArmCount
Dispensed Subjects
All subjects that were dispensed at least one study lens.
101
Total101

Withdrawals & dropouts

PeriodReasonFG000FG001FG002FG003FG004FG005
Period 1Withdrawal by Subject000010

Baseline characteristics

CharacteristicDispensed Subjects
Age, Continuous32.43 Years
STANDARD_DEVIATION 9.768
Race/Ethnicity, Customized
Asian
101 Participants
Region of Enrollment
China
101 Participants
Sex: Female, Male
Female
56 Participants
Sex: Female, Male
Male
45 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
— / —— / —— / —
other
Total, other adverse events
0 / 1000 / 1000 / 101
serious
Total, serious adverse events
0 / 1000 / 1000 / 101

Outcome results

Primary

CLUE Handling

CLUE Handling is assessed using the Contact Lens User Experience (CLUE™) Questionnaire. CLUE™ is a validated patient-reported outcomes questionnaire to assess patient experience attributes of soft, disposable contact lenses (comfort, vision, handling, and packaging) in a contact lens-wearing population in the US, ages 18-65. Scores follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable / positive response. 97% of the scores fall within 0 and 120 (mean +/- 3XSD).

Time frame: 30 Minutes Post Insertion

Population: Subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
Test1CLUE Handling55.73 Units on a scaleStandard Deviation 20.146
Test 2CLUE Handling57.58 Units on a scaleStandard Deviation 20.109
ControlCLUE Handling64.39 Units on a scaleStandard Deviation 19.825
Primary

Percentage of Eyes With Acceptable Cosmetic Lens Fit

Cosmetic lens fitting is assessed for each subject eye in the primary gaze position without a slit lamp. Cosmetic lens fit is a binary response and is reported as acceptable or unacceptable. The Percentage of subject eyes with acceptable cosmetic lens fitting is reported.

Time frame: 30 Minutes Post Insertion

Population: Subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (NUMBER)
Test1Percentage of Eyes With Acceptable Cosmetic Lens Fit99.5 Percentage of eyes
Test 2Percentage of Eyes With Acceptable Cosmetic Lens Fit99.5 Percentage of eyes
ControlPercentage of Eyes With Acceptable Cosmetic Lens Fit99.5 Percentage of eyes
Primary

Percentage of Eyes With Acceptable Mechanical Lens Fit

Mechanical lens fit will be assessed for each subject and eye using a slit lamp. Lens fit is a binary response and 'Yes' = Acceptable Fit and 'No' = Unacceptable Fit. Lens fit is assessed using Lens centration, limbal exposure, primary gaze movement, up-gaze movement, edge lift, lens tightness. Unacceptable fit will be declared if there is any of the following present: limbal exposure, edge lift, excessive movement in primary or up-gaze or insufficient movement in primary gaze and up-gaze. The Percentage of subject eyes with acceptable lens fit will be reported.

Time frame: 30 Minutes Post Insertion

Population: Subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (NUMBER)
Test1Percentage of Eyes With Acceptable Mechanical Lens Fit95 Percentage of eyes
Test 2Percentage of Eyes With Acceptable Mechanical Lens Fit97 Percentage of eyes
ControlPercentage of Eyes With Acceptable Mechanical Lens Fit91 Percentage of eyes
Primary

Visual Performance (LogMar)

Distance time controlled Visual Performance (LogMAR) was assessed for each subject eye under bright-illumination high-contrast lighting conditions at 4m using an ETDRS chart. The average visual performance (LogMAR) was reported for each study lens.

Time frame: 30 Minutes Post Insertion

Population: Subjects that completed all study visits without a major protocol deviation.

ArmMeasureValue (MEAN)Dispersion
Test1Visual Performance (LogMar)0.00 LogMARStandard Deviation 0.082
Test 2Visual Performance (LogMar)0.00 LogMARStandard Deviation 0.082
ControlVisual Performance (LogMar)0.01 LogMARStandard Deviation 0.082

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026