Apnea
Conditions
Brief summary
Investigators compared tidal volumes for single rescuer ventilation using a modified bag valve mask with a supplemental external handle versus a conventional bag valve mask in a manikin model.
Detailed description
This was a prospective, randomized, crossover study to assess the tidal volume delivered using a standard and modified bag valve mask (BVM) device in a manikin airway model. The modified device comprised a standard bag valve mask with supplemental external bar. Data was collected during May 2016 at the San Antonio Military Medical Center. Emergency medicine providers (physicians, physician assistants, residents, nurses, medics) were randomized to device order. Prior to participation each participant filled out a survey indicating their job status, gender, years of medical experience, and experience level with BVM ventilation. Hand grip strength and size (length, width, span) were measured. Each participant was then asked to provide BVM ventilation using the assigned devices at a rate of 10 breaths per minute for 3 minutes for a total of 30 breaths. Tidal volume of each delivered breath was recorded in milliliters. After a 3 minute rest period, testing was repeated with the second device. After ventilation with each device, participants completed an anonymous questionnaire that used a Likert scale to assess perceived qualities of the modified device including ease of use, ability to provide a superior mask seal, willingness to use in an emergency situation, and overall preference between the two devices. Each participant served as his/her own control.
Interventions
Subjects deliver breaths to manikin (IngMar RespiTrainer manikin model) a modified Ambu Spur II bag valve mask with a supplemental external handle
Subjects deliver breaths to manikin (IngMar RespiTrainer manikin model) a conventional Ambu Spur II bag valve mask
Sponsors
Study design
Eligibility
Inclusion criteria
* Emergency Department healthy volunteers (physician assistants, residents, paramedics, nurses, medics, respiratory therapists).
Exclusion criteria
* Not trained in basic life support
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Received Tidal Volume | 3 minutes | Mean received tidal volume for 30 breaths delivered by the subjects over three minutes as measured via the RespiTrainer Advance manikin model monitor output. |
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Modified Then Conventional Ambu Spur II Bag Valve Mask A health volunteer uses a modified Ambu Spur II bag valve mask with integrated internal handle (experimental device, not yet FDA approved) to deliver 10 breaths per minute for 3 minutes to a manikin (IngMar RespiTrainer manikin model).
Modified Ambu Spur II bag valve mask: Subjects deliver breaths to manikin (IngMar RespiTrainer manikin model) a modified Ambu Spur II bag valve mask with a supplemental external handle.
After completion, subjects repeat all these procedures using a conventional Ambu Spur II bag valve mask. | 35 |
| Conventional Then Modified Ambu Spur II Bag Valve Mask A health volunteer uses a conventional Ambu Spur II bag valve mask to deliver 10 breaths per minute for 3 minutes to a manikin (IngMar RespiTrainer manikin model).
Conventional Ambu Spur II bag valve mask: Subjects deliver breaths to manikin (IngMar RespiTrainer manikin model) a conventional Ambu Spur II bag valve mask.
After completion, subjects repeat all these procedures using the modified Ambu Spur II bag valve mask. | 35 |
| Total | 70 |
Baseline characteristics
| Characteristic | Modified Then Conventional Ambu Spur II Bag Valve Mask | Conventional Then Modified Ambu Spur II Bag Valve Mask | Total |
|---|---|---|---|
| Age, Categorical <=18 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical >=65 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical Between 18 and 65 years | 35 Participants | 35 Participants | 70 Participants |
| Region of Enrollment United States | 35 participants | 35 participants | 70 participants |
| Sex: Female, Male Female | 16 Participants | 16 Participants | 32 Participants |
| Sex: Female, Male Male | 19 Participants | 19 Participants | 38 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | — / — | — / — |
| other Total, other adverse events | 0 / 70 | 0 / 70 |
| serious Total, serious adverse events | 0 / 70 | 0 / 70 |
Outcome results
Received Tidal Volume
Mean received tidal volume for 30 breaths delivered by the subjects over three minutes as measured via the RespiTrainer Advance manikin model monitor output.
Time frame: 3 minutes
| Arm | Measure | Value (MEAN) |
|---|---|---|
| Modified Ambu Spur II Bag Valve Mask | Received Tidal Volume | 662 milliliters |
| Conventional Bag Valve Mask | Received Tidal Volume | 681 milliliters |