Failed or Difficult Intubation
Conditions
Brief summary
We will conduct this study to show that it is possible to deliver air through a Frova bougie and an Ambu® bag.
Detailed description
Participants will be invited to squeeze an Ambu® bag connected to a Frova bougie, and we will measure the volume of air delivered and peak pressure generated.
Interventions
Squeezing an Ambu bag and delivering air through a Frova bougie. Measuring mean volumes of air delivered and peak pressures generated.
Sponsors
Study design
Eligibility
Inclusion criteria
* staff, fellows and resident anesthesiologists
Exclusion criteria
* injury to hand
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Tidal Volume (ml) | 15 minutes | volume of air delivered via Frova bougie (milliliters) |
Countries
Canada
Participant flow
Pre-assignment details
This study design involves a cross-over of participants (i.e., participants are randomly assigned to one of the four groups after which they crossover to the other 3).
Participants by arm
| Arm | Count |
|---|---|
| Maximum and Normal Effort Using Long and Short Frova Bougies In a random order, participants would squeeze an Ambu bag attached to a long or short Frova bougie using both maximum and normal effort. As such, participants performed a total of 4 sets of squeezes (maximum effort: long, normal effort: long, maximum effort: short, and normal effort: short) in a random order.
Frova bougie: Squeezing an Ambu bag attached to a Frova bougie and measuring mean volumes of air delivered. | 35 |
| Total | 35 |
Baseline characteristics
| Characteristic | Maximum and Normal Effort Using Long and Short Frova Bougies |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 0 Participants |
| Age, Categorical Between 18 and 65 years | 35 Participants |
| Region of Enrollment Canada | 35 Participants |
| Sex: Female, Male Female | 18 Participants |
| Sex: Female, Male Male | 17 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 0 / 35 |
| serious Total, serious adverse events | 0 / 35 |
Outcome results
Tidal Volume (ml)
volume of air delivered via Frova bougie (milliliters)
Time frame: 15 minutes
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Maximum Effort: Long | Tidal Volume (ml) | 954.8 tidal volume (ml) | Standard Deviation 120.7 |
| Normal Effort: Long | Tidal Volume (ml) | 554.4 tidal volume (ml) | Standard Deviation 131.3 |
| Maximum Effort: Short | Tidal Volume (ml) | 893.5 tidal volume (ml) | Standard Deviation 171.6 |
| Normal Effort: Short | Tidal Volume (ml) | 542.9 tidal volume (ml) | Standard Deviation 132.9 |