FRailty WAlking Patterns (FRAP) Study

FRailty WAlking Patterns (FRAP) Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02755129
Acronym
FRAP
Enrollment
15
Registered
2016-04-28
Start date
2016-09-20
Completion date
2017-07-20
Last updated
2020-03-25

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Heart Failure

Brief summary

The aim of this study is to evaluate sit-stand phases and gait speed detection using an externally worn Reveal LINQ (TM) compared to an external reference (3D accelerometer, and/or the Computer Assisted Rehabilitation Environment (CAREN) system) in one center in the Netherlands

Detailed description

Frailty is a geriatric syndrome characterized by reduced homeostatic reserves, exposing the organism to extreme vulnerability to endogenous and exogenous stressors. Frailty is prevalent in older people and involves a progressive physiological decline of multiple body systems, typical signs and symptoms include weight loss, fatigue, muscle weakness, slow or unsteady gait declines in activity. Frailty is increasingly recognized as an important prognostic indicator in heart failure (HF) and is more prevalent in HF than the general population. The identification of frailty in its early stage is important because interventions may potentially prevent, or delay the clinical consequences of frailty. Of particular focus in this study will be walking speed as prior research has demonstrated that slow gait speed has the strongest prognostic ability of the traditional components used to assess frailty, and has been reported as one of the strongest to predict adverse outcomes, such as mobility disability, falls, or hospitalization. Also of interest is the detection of posture changes as this may have implications for detecting changes in sleeping habits and could also provide context for other biomarker signals collected by the LINQ device. The literature has been reviewed and the scientific soundness of the proposed analytical techniques evaluated. The rationale for this study design is to evaluate the feasibility of using the Reveal LINQ™ to monitoring walking patterns. HF patients (the target population for the study in discussion) would make it possible to test the sensitivity of the accelerometer embedded in the LINQ device in a small cohort of subjects with reduced mobility and for which frailty is more prevalent than the general population, as well as recognized as an important prognostic indicator. No risk to the subjects is expected with this study.

Interventions

OTHERWalking Exercises-Six Minute Walk (6MW) Test

The Six Minute Walk (6MW) Test measures the distance an individual is able to walk over a total of six minutes on a flat surface. The goal is for the individual to walk as far as possible in six minutes. The Six Minute Walk (6MW) test will be performed within the Computer Assisted Rehabilitation Environment (CAREN) system system environment - single arm

OTHERWalking Exercises-4 Meter Gait Speed (4MGS) Test

The 4 Meter Gait Speed (4MGS) Test measures the gait speed an individual is able to walk over 4 meters on a flat surface. The goal is for the individual to walk as fast as possible over 4 meters. It will be repeated 3 times within the aforementioned 6MW test (after approximately 1, 3 and 5 minutes) to get variability due to subject tiredness. The test will be performed within the CAREN system environment. - single arm

OTHERWalking Exercises-Five Times Sit to Stand (FTSTS) Test

The Five Times Sit to Stand (FTSTS) Test measures the time a subject takes to stand up from an armchair 5 times in a row, without stopping in between. It is a short test which measures dynamic balance and functional mobility - single arm

OTHERWalking Exercises-Expanded Timed Get-Up-and-Go (ETGUG) Test

The Expanded Timed Get-Up-and-Go (ETGUG) Test measures the time it takes a subject to stand up from an armchair, walk a distance of 10 m, turn, walk back to the chair, and sit down. It is a short test of basic mobility skills for frail community-dwelling elderly. - single arm

OTHERReveal LINQ accelerometer

Walking exercises test

OTHER3D accelerometer

Walking exercises test

Sponsors

Medtronic Cardiac Rhythm and Heart Failure
CollaboratorINDUSTRY
Medtronic BRC
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Chronic Heart Failure in New York Heart Association (NHYA) class II and class III * Willing to sign the informed consent form. * At least 18 years of age.

Exclusion criteria

* • Not able to walk continuously for a period of 6 minutes and perform the walking exercises as necessary for the study protocol. * Any known allergy to Titanium * Any concomitant conditions which in the opinion of the investigator would not allow accurate measurement of gait and frailty parameters with an externally worn device. * Any concomitant condition which in the opinion of the investigator would not allow a safe participation in the study. * Enrolled in another study that could confound the results of this study, without documented pre-approval from a Medtronic study manager.

Design outcomes

Primary

MeasureTime frameDescription
Correlation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesDuring walking exercises visit, 1 DayThe gait speed will be calculated from the Reveal LINQ accelerometer signals, from the validation accelerometer and / or from the computer assisted rehabilitation system during the walking exercises. We will calculate the average correlation coefficient over all patients and exercises. It is reported as score on a scale, with minimum value= -1 and maximum value = 1. Higher absolute values mean higher correlation.

Countries

Netherlands

Participant flow

Participants by arm

ArmCount
All Study Participants
Patients with Chronic Heart Failure. No control group has been used for this study, but all the patients enrolled will perform the same walking exercises. On arrival to the rehabilitation center, subjects will have the Reveal LINQ, the 3D-accelerometers (one on the chest by medical-grade adhesives and a second at the level of the waist over the top of a medical grade adhesive) and the Holter attached externally. Then, they will be asked to perform the following exercises: * Walking exercises - Six Minute Walk (6MW) Test * Walking exercises - 4 Meter Gait Speed (4MGS) Test * Walking Exercises - Five Times Sit to Stand (FTSTS) Test * Walking Exercises - Expanded Timed Get-Up-and-Go (ETGUG) Test
15
Total15

Baseline characteristics

CharacteristicAll Study Participants
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
10 Participants
Age, Categorical
Between 18 and 65 years
5 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Netherlands
15 participants
Sex: Female, Male
Female
4 Participants
Sex: Female, Male
Male
11 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 0
other
Total, other adverse events
0 / 0
serious
Total, serious adverse events
0 / 0

Outcome results

Primary

Correlation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking Exercises

The gait speed will be calculated from the Reveal LINQ accelerometer signals, from the validation accelerometer and / or from the computer assisted rehabilitation system during the walking exercises. We will calculate the average correlation coefficient over all patients and exercises. It is reported as score on a scale, with minimum value= -1 and maximum value = 1. Higher absolute values mean higher correlation.

Time frame: During walking exercises visit, 1 Day

Population: 15 subjects were enrolled and completed the study. All subjects met the inclusion and exclusion criteria.~One subject did not have readable device data and as such was excluded from the analysis.

ArmMeasureGroupValue (NUMBER)
Correlation Coefficient - Six Minute Walk (6MW) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesCorrelation per each Exercise0.97 Correlation Coefficient
Correlation Coefficient - Six Minute Walk (6MW) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesAverage Correlation Across All Exercises0.91 Correlation Coefficient
Correlation Coefficient - 4 Meter Gait Speed (4MGS) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesAverage Correlation Across All Exercises0.91 Correlation Coefficient
Correlation Coefficient - 4 Meter Gait Speed (4MGS) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesCorrelation per each Exercise0.93 Correlation Coefficient
Correlation Coefficient - Five Times Sit to Stand (FTSTS) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesCorrelation per each Exercise0.86 Correlation Coefficient
Correlation Coefficient - Five Times Sit to Stand (FTSTS) TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesAverage Correlation Across All Exercises0.91 Correlation Coefficient
Correlation Coefficient - ETGUG TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesCorrelation per each Exercise0.91 Correlation Coefficient
Correlation Coefficient - ETGUG TestCorrelation of Gait Speed Between Reveal LINQ Accelerometer, Validation Accelerometer and/or Computer Assisted Rehabilitation System by Walking ExercisesAverage Correlation Across All Exercises0.91 Correlation Coefficient