Evaluating Heat Transfer With the Esophageal Cooling Device

Prospective, Interventional Study Evaluating Heat Transfer With the Esophageal Cooling Device

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02743884
Enrollment
19
Registered
2016-04-19
Start date
2016-04-30
Completion date
2017-01-31
Last updated
2017-04-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Non-cardiac Surgery

Brief summary

An esophageal heat exchange tube will be inserted in anesthetized surgical patients. Each patient will have 30 minutes of cooling (circulating fluid at 7 degrees C) and 30 minutes of warming (42 degrees C) with 15-minute recovery break in between. The primary outcome will be heat transfer, determined from inflow and outflow temperatures and fluid flow rate for cooling and warming modes.

Detailed description

The ECD is made with medical-grade silicone and has 3 ports; 2 coolant ports and a central port. The coolant ports of the ECD are connected to external heat exchanger while a third, central lumen simultaneously allows gastric decompression and drainage. The ECD can be inserted like a standard gastric tube. The aim of this prospective, interventional study is to quantify heat transfer during warming and cooling using the Esophageal Cooling Device in anesthetized patients having non-cardiac surgery. An esophageal heat exchange tube will be inserted in anesthetized surgical patients. Each will have 30 minutes of cooling (circulating fluid at 7 degrees C) and 30 minutes of warming (42 degrees C) with 15 -minute break in between; the order of the cooling and warming will be randomized. The primary outcome will be heat transfer, determined from inflow and outflow fluid temperatures and fluid flow rate assessed for both cooling and warming modes.

Interventions

30 minutes of warming from circulating water at about 42 degrees C through an esophageal heat exchanger.

30 minutes of cooling from circulating water at about 7 degrees C through an esophageal heat exchanger.

Sponsors

The Cleveland Clinic
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Intervention model description

Cross-over.

Eligibility

Sex/Gender
ALL
Age
18 Years to 80 Years
Healthy volunteers
No

Inclusion criteria

* Elective non-cardiac surgery; * Age 18-80 years; * Surgery expected to last \>2 hours; * General anesthesia with endotracheal Intubation; * Volatile anesthesia; * BMI \< 38 kg/m2; * ASA physical status 1-3.

Exclusion criteria

* Weight \< 45 kg; * Unstable blood pressure; * Known esophageal deformity or evidence of esophageal trauma or esophageal disease that in the opinion of the attending anesthesiologist precludes safe use of ECD.

Design outcomes

Primary

MeasureTime frameDescription
Mean heat transfer at cooling, kcal/h15-60 minutes after intubation during elective non-cardiac surgeryHeat transfer is calculated as difference between inflow and outflow temperatures, multiplied by flow and the specific heat of water
Mean heat transfer at warming, kcal/h15-60 minutes after intubation during elective non-cardiac surgeryHeat transfer is calculated as difference between inflow and outflow temperatures, multiplied by flow and the specific heat of water