Hallux Valgus
Conditions
Brief summary
The purpose of the study is to evaluate safety, tolerability and performance of the CyclaPlex implant, a button and suture type implant device and instruments for the correction of the first inter metatarsal angle in subjects suffering from Hallux Valgus deformity.
Interventions
Button and suture type implant that will be implanted between the first and second metatarsal
Sponsors
Study design
Eligibility
Inclusion criteria
* Weight less than 100 kilos * IMA greater than 12 degrees * IMA less than 7 degrees
Exclusion criteria
* Diabetics * Known osteoporosis * Known allergy or hypersensitivity to metals * Gout or any systemic inflammatory arthropathy
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| IMA Reduction | One year | The IMA angle is measured pre-op and post-op on x-ray images using internal software. The IMA at end of study should be less than 10 degrees or a reduction of more than 5 degrees. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| American Orthopedic Foot and Ankle Society (AOFAS) Score | One year | General improvement as assessed by AOFAS Hallux Scale metatarsophalangeal interphalangeal scale score |
Countries
Israel