Utilizing Holmium Laser for FURS Renal Stone Lithotripsy Comparing Stone Breaking vs. Stone Dusting Settings

Utilizing Holmium Laser for FURS Renal Stone Lithotripsy Comparing Stone Breaking vs. Stone Dusting Settings

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02683993
Enrollment
20
Registered
2016-02-17
Start date
2016-02-29
Completion date
2017-01-31
Last updated
2026-01-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Stone Clearance

Brief summary

Twenty (20) subjects presenting with a renal stone and candidates for FURS (flexible ureteroscopic renal surgery) will undergo a single FURS procedure for a kidney stone as part of the study, using the study device. These subjects will be randomized into stone breaking treatment or stone dusting.

Detailed description

This is a randomized, two-arm, prospective, single center study. Subjects will undergo a single flexible ureteroscopic renal surgery (FURS) procedure for a kidney stone as part of this study, using the study device, and will be treated as randomized to either stone breaking treatment or stone dusting treatment. Follow up visits are scheduled for the day of release from the hospital and at 1 month post procedure. Additional treatment sessions, if required, will be at the discretion of the investigator, as part of the site standard of practice. Additional treatment sessions will not be included in this study.

Interventions

DEVICELumenis Pulse 120H holmium laser system and 200 DFL ball tip fiber

The Lumenis Pulse P120H will be utilized in high and low frequencies according to subjects group allocation.

Sponsors

Lumenis Be Ltd.
CollaboratorINDUSTRY
Boston Scientific Corporation
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Subject will be older than 18 years of age 2. Subject was diagnosed with renal calculi meeting the following criteria: * Largest stone measures no more than 2cm on a single diameter and is no less than 0.8cm as defined by X-ray or CT. * Accumulative stone burden (addition of all stone diameters) is no more than 2.5cm. 3. Subject is a candidate for FURS procedure 4. Subject is willing and has signed the Informed Consent Form

Exclusion criteria

1. Subject has undergone a previous treatment for stones in the same kidney 2. Subject requires another, concomitant procedure, other than FURS, to be performed during this treatment session. 3. Female subject is pregnant 4. Concomitant anticoagulant medication that cannot be suspended during surgery

Design outcomes

Primary

MeasureTime frameDescription
Stone ClearanceStone clearance based on 1 month follow-up imaging datastone clearance will be confirmed by CT or X-Ray at 1 month follow up.
Post-procedure CatheterizationPost-procedure through hospital discharge; typically discharge occurs on the same day or 1 day post procedurePost-procedure catheterization by group
Procedure Timeduring the procedureDuration (in minutes) of the procedure as an indicator for treatment efficiency.
Usage of Alternative Stone Extracting Devicesas measured during the procedureThe need of the surgeon to use baskets, forceps or saline flushing to achieve complete stone clearance. This outcome can influence both procedure time and further complication.
Hospital StayProcedure through hospital dischargeHospital duration was measured to evaluate treatment efficiency

Secondary

MeasureTime frameDescription
Number of Participants With Intraoperative ComplicationsIntraoperativeNumber of participants with complications that occur during the procedure.
Fiber TransmissionIntraoperativeFiber transmission (preserving energy transmission during lasing) by group
Visibility During ProcedureIntraproceduralVisibility during procedure by group, rating categories: Poor, Fair, Good, Very Good or Excellent
Number of Participants With Occurrence of RetropulsionIntraoperativeOccurrence of retropulsion by group
Ability to Reach Stone FragmentsIntraoperativePerformance rating of ability to reach stone fragments by group
Damage to Scope or FiberIntraoperativeDamage to scope or fiber by group

Countries

United States

Participant flow

Participants by arm

ArmCount
Stone Breaking
Lithotripsy will be performed using the Lumenis Pulse 120H holmium laser system with low frequency parameters. Lumenis Pulse 120H holmium laser system and 200 DFL ball tip fiber: The Lumenis Pulse P120H will be utilized in high and low frequencies according to subjects group allocation.
10
Stone Dusting
Lithotripsy will be performed using the Lumenis Pulse 120H holmium laser system with high frequency parameters. Lumenis Pulse 120H holmium laser system and 200 DFL ball tip fiber: The Lumenis Pulse P120H will be utilized in high and low frequencies according to subjects group allocation.
10
Total20

Baseline characteristics

CharacteristicStone DustingTotalStone Breaking
Age, Continuous60.9 years
STANDARD_DEVIATION 15.4
61.6 years
STANDARD_DEVIATION 13.2
62.2 years
STANDARD_DEVIATION 11.2
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
10 participants20 participants10 participants
Sex: Female, Male
Female
5 Participants7 Participants2 Participants
Sex: Female, Male
Male
5 Participants13 Participants8 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 100 / 10
other
Total, other adverse events
0 / 101 / 10
serious
Total, serious adverse events
0 / 100 / 10

Outcome results

Primary

Hospital Stay

Hospital duration was measured to evaluate treatment efficiency

Time frame: Procedure through hospital discharge

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingHospital StaySame day hospital discharge10 Participants
Stone BreakingHospital StayNext day hospital discharge0 Participants
Stone DustingHospital StaySame day hospital discharge8 Participants
Stone DustingHospital StayNext day hospital discharge2 Participants
Primary

Post-procedure Catheterization

Post-procedure catheterization by group

Time frame: Post-procedure through hospital discharge; typically discharge occurs on the same day or 1 day post procedure

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingPost-procedure CatheterizationUreteric catheter not placed at the end of procedure9 Participants
Stone BreakingPost-procedure CatheterizationUreteric catheter placed at the end of procedure and not removed at hospital discharge1 Participants
Stone BreakingPost-procedure CatheterizationUreteric catheter placed at the end of procedure; removal at hospital discharge unknown0 Participants
Stone DustingPost-procedure CatheterizationUreteric catheter not placed at the end of procedure9 Participants
Stone DustingPost-procedure CatheterizationUreteric catheter placed at the end of procedure and not removed at hospital discharge0 Participants
Stone DustingPost-procedure CatheterizationUreteric catheter placed at the end of procedure; removal at hospital discharge unknown1 Participants
Primary

Procedure Time

Duration (in minutes) of the procedure as an indicator for treatment efficiency.

Time frame: during the procedure

ArmMeasureValue (MEAN)Dispersion
Stone BreakingProcedure Time46.3 MinutesStandard Deviation 24.8
Stone DustingProcedure Time36.1 MinutesStandard Deviation 16.5
Primary

Stone Clearance

stone clearance will be confirmed by CT or X-Ray at 1 month follow up.

Time frame: Stone clearance based on 1 month follow-up imaging data

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingStone ClearanceComplete clearance6 Participants
Stone BreakingStone ClearanceEvidence of fragments not requiring intervention3 Participants
Stone BreakingStone ClearanceEvidence of fragments requiring an additional intervention1 Participants
Stone DustingStone ClearanceComplete clearance6 Participants
Stone DustingStone ClearanceEvidence of fragments not requiring intervention4 Participants
Stone DustingStone ClearanceEvidence of fragments requiring an additional intervention0 Participants
Primary

Usage of Alternative Stone Extracting Devices

The need of the surgeon to use baskets, forceps or saline flushing to achieve complete stone clearance. This outcome can influence both procedure time and further complication.

Time frame: as measured during the procedure

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingUsage of Alternative Stone Extracting DevicesAny alternate stone extracting devices used0 Participants
Stone BreakingUsage of Alternative Stone Extracting DevicesNo alternate stone extracting devices used10 Participants
Stone DustingUsage of Alternative Stone Extracting DevicesAny alternate stone extracting devices used1 Participants
Stone DustingUsage of Alternative Stone Extracting DevicesNo alternate stone extracting devices used9 Participants
Secondary

Ability to Reach Stone Fragments

Performance rating of ability to reach stone fragments by group

Time frame: Intraoperative

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingAbility to Reach Stone FragmentsEasily performed6 Participants
Stone BreakingAbility to Reach Stone FragmentsPerformed well4 Participants
Stone BreakingAbility to Reach Stone FragmentsPerformed with difficulty0 Participants
Stone BreakingAbility to Reach Stone FragmentsImpossible to perform0 Participants
Stone DustingAbility to Reach Stone FragmentsImpossible to perform0 Participants
Stone DustingAbility to Reach Stone FragmentsEasily performed9 Participants
Stone DustingAbility to Reach Stone FragmentsPerformed with difficulty0 Participants
Stone DustingAbility to Reach Stone FragmentsPerformed well1 Participants
Secondary

Damage to Scope or Fiber

Damage to scope or fiber by group

Time frame: Intraoperative

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingDamage to Scope or FiberScope damageYes0 Participants
Stone BreakingDamage to Scope or FiberScope damageNo10 Participants
Stone BreakingDamage to Scope or FiberFiber damageYes0 Participants
Stone BreakingDamage to Scope or FiberFiber damageNo10 Participants
Stone DustingDamage to Scope or FiberFiber damageNo10 Participants
Stone DustingDamage to Scope or FiberScope damageYes0 Participants
Stone DustingDamage to Scope or FiberFiber damageYes0 Participants
Stone DustingDamage to Scope or FiberScope damageNo10 Participants
Secondary

Fiber Transmission

Fiber transmission (preserving energy transmission during lasing) by group

Time frame: Intraoperative

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingFiber TransmissionEasily performed8 Participants
Stone BreakingFiber TransmissionPerformed well2 Participants
Stone BreakingFiber TransmissionPerformed with difficulty0 Participants
Stone BreakingFiber TransmissionImpossible to perform0 Participants
Stone DustingFiber TransmissionImpossible to perform0 Participants
Stone DustingFiber TransmissionEasily performed9 Participants
Stone DustingFiber TransmissionPerformed with difficulty0 Participants
Stone DustingFiber TransmissionPerformed well1 Participants
Secondary

Number of Participants With Intraoperative Complications

Number of participants with complications that occur during the procedure.

Time frame: Intraoperative

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Stone BreakingNumber of Participants With Intraoperative Complications0 Participants
Stone DustingNumber of Participants With Intraoperative Complications0 Participants
Secondary

Number of Participants With Occurrence of Retropulsion

Occurrence of retropulsion by group

Time frame: Intraoperative

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingNumber of Participants With Occurrence of RetropulsionDid not occur8 Participants
Stone BreakingNumber of Participants With Occurrence of RetropulsionOccurred several times but could be controlled2 Participants
Stone BreakingNumber of Participants With Occurrence of RetropulsionOccurred several times and was difficult to control0 Participants
Stone BreakingNumber of Participants With Occurrence of RetropulsionStone lost to retropulsion0 Participants
Stone DustingNumber of Participants With Occurrence of RetropulsionStone lost to retropulsion0 Participants
Stone DustingNumber of Participants With Occurrence of RetropulsionDid not occur10 Participants
Stone DustingNumber of Participants With Occurrence of RetropulsionOccurred several times and was difficult to control0 Participants
Stone DustingNumber of Participants With Occurrence of RetropulsionOccurred several times but could be controlled0 Participants
Secondary

Visibility During Procedure

Visibility during procedure by group, rating categories: Poor, Fair, Good, Very Good or Excellent

Time frame: Intraprocedural

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Stone BreakingVisibility During ProcedureFair0 Participants
Stone BreakingVisibility During ProcedureVery Good7 Participants
Stone BreakingVisibility During ProcedurePoor0 Participants
Stone BreakingVisibility During ProcedureExcellent3 Participants
Stone BreakingVisibility During ProcedureGood0 Participants
Stone DustingVisibility During ProcedureExcellent3 Participants
Stone DustingVisibility During ProcedurePoor0 Participants
Stone DustingVisibility During ProcedureFair0 Participants
Stone DustingVisibility During ProcedureGood1 Participants
Stone DustingVisibility During ProcedureVery Good6 Participants