Rheumatoid Arthritis
Conditions
Brief summary
This study is designed to evaluate RLHFs concerning administration of the tocilizumab autoinjector AI-1000 G2 in adults with rheumatoid arthritis (RA) who have been receiving subcutaneous (SC) tocilizumab using the commercially available prefilled syringe and needle safety device (PFS-NSD). The study will enroll participants with RA, a subset of whom will be assigned to perform self-injection with the AI-1000 G2. Enrolled caregivers (CGs) and healthcare professionals (HCPs) will administer the AI-1000 G2 injection to the remaining study participants.
Interventions
Tocilizumab will be administered using the AI-1000 G2.
Participants will receive three doses of SC tocilizumab using the AI-1000 G2 device at Visits 1, 2, and 3 (Days 0, 14, and 28). The dose will remain at 162 milligrams (mg) unless changes are required per protocol for safety or efficacy reasons.
Sponsors
Study design
Eligibility
Inclusion criteria
* Participants with RA for greater than or equal to (\>/=6) months receiving 162 mg tocilizumab PFS-NSD for \>/=8 weeks and who are suitable for continued treatment at their currently prescribed dose * CGs and professionally qualified HCPs who are able and willing to administer the injection
Exclusion criteria
* RA: Functional status Class IV * RA: Neuropathies or other conditions that might interfere with pain evaluation * RA: Pregnant or breastfeeding * RA: Low neutrophil or platelet count at last laboratory assessment * RA: Elevated liver enzymes at last laboratory assessment * Current participation in another interventional clinical trial * Criteria that might give the participant/CG/HCP an advantage in injection tasks such as employment in the pharmaceutical industry, etc.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Day 14 | Safety-critical tasks included those tasks where errors would have a reasonably foreseeable potential for clinical impact/harm, potentially resulting in direct physical injury to the user and/or conditions that require medical intervention. Safety-critical tasks included the following: release activation button; check the expiry date; inspect device prior to use; and inspect medication prior to use. Essential tasks included those essential to the execution of the injection. Essential tasks included the following: open the carton, remove the device and associated documents; remove cap; start an injection by depressing the needle-shield at the injection site and pressing the activation button; and hold the autoinjector until the complete dose has been delivered. The percentage of participants who successfully performed safety-critical and essential tasks during the first unassisted use (Day 14) was reported. |
| Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Day 28 | Safety-critical tasks included those tasks where errors would have a reasonably foreseeable potential for clinical impact/harm, potentially resulting in direct physical injury to the user and/or conditions that require medical intervention. Safety-critical tasks included the following: release activation button; check the expiry date; inspect device prior to use; and inspect medication prior to use. Essential tasks included those essential to the execution of the injection. Essential tasks included the following: open the carton, remove the device and associated documents; remove cap; start an injection by depressing the needle-shield at the injection site and pressing the activation button; and hold the autoinjector until the complete dose has been delivered. The percentage of participants who successfully performed safety-critical and essential tasks during the second unassisted use (Day 28) was reported. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | 0 and 15 minutes after injection on Days 0, 14, 28 | Injection pain was assessed on a categorical 6-point Likert scale, ranging from no pain to severe and intolerable. The percentage of participants was reported by response at each assessment timepoint among participants with RA. |
| Percentage of Participants by Response to Device Satisfaction Questionnaire | Days 0, 14, 28 | Device satisfaction was assessed using twelve questions on a categorical 5-point Likert scale, ranging from strongly agree to strongly disagree. Question (Q) 1 (felt the autoinjection was easy to use), Q2 (felt comfortable while using the autoinjector), Q3 (felt the autoinjector was easy to hold), Q4 (able to tell when injection had completed), Q5 (felt he/she can inject properly with the autoinjector), Q6 (felt he/she had control over the injection process), Q7 (felt confident that he/she injected successfully), Q8 (felt it is easy to dispose of the autoinjector), Q9 (autoinjector would help to manage his/her injection schedule), Q10 (liked the look/feel of the autoinjector), Q11 (would recommend the autoinjector to someone else who needed to inject), Q12 (would continue having injections with the autoinjector). The percentage of participants was reported by response at each assessment among participants with RA, CGs, and HCPs. |
| Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | Days 14, 28 | Ancillary tasks included those tasks where the potential harm resulting from use error would be minor in severity, or the resultant harms were estimated to occur at sufficiently low levels. Ancillary tasks included the following: wash hands; clean the injection site with alcohol swab; wait for the alcohol to dry; dispose of the autoinjector cap; inspect full dose delivered after use; dispose of the autoinjector; and treatment of injection site after injection. The percentage of participants who succesffully performed ancillary tasks during the first (Day 14) and second unassisted use (Day 28) was reported. |
| Swollen Joint Count (SJC) Among Participants With RA | Baseline (Day 0) | Sixty-six joints were assessed for swelling among participants with RA. The number of swollen joints at Baseline was reported. |
| Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Days 0, 14, 28 | Ability to perform daily living activities was assessed across eight component sets including dressing/grooming, arising, eating, walking, hygiene, reach, grip, and common activities. Twenty questions were scored on a 4-point Likert scale from 0 meaning without any difficulty to 3 meaning unable to do. The overall score was computed as the sum of domain scores divided by the number of domains answered. Therefore, the score range for HAQ-DI was the same as that of the individual questions, that is, from 0 to 3, where 0 indicates least difficulty and 3 indicates extreme difficulty. The mean change from baseline in HAQ-DI at each assessment was reported among participants with RA. |
| Tender Joint Count (TJC) Among Participants With RA | Baseline (Day 0) | Sixty-eight joints were assessed for tenderness among participants with RA. The number of tender joints at Baseline was reported. |
| Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | 0 and 15 minutes after injection on Days 0, 14, 28 | Injection pain was assessed on a continuous 100-millimeter (mm) VAS, where 0 mm represents no pain and 100 mm represents unbearable pain. The mean VAS response at each assessment timepoint was reported among participants with RA. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Caregivers CGs performed injection of 162 mg SC tocilizumab to a subset of participants with RA using the AI-1000 G2 device. Visit 1 (Day 0) was conducted for administration training, while Visits 2 and 3 (Days 14 and 28) were conducted for use/performance evaluation. | 17 |
| Healthcare Professionals HCPs performed injection of 162 mg SC tocilizumab to a subset of participants with RA using the AI-1000 G2 device. Because enrolled HCPs were to be professionally qualified to deliver SC injections, no administration training was provided. Visits 2 and 3 (Days 14 and 28) were conducted for use/performance evaluation. | 19 |
| RA Group 1 (Self-Administration) Participants with RA performed self-injection of 162 mg SC tocilizumab with the AI-1000 G2 device. Visit 1 (Day 0) was conducted for administration training, while Visits 2 and 3 (Days 14 and 28) were conducted for use/performance evaluation. | 18 |
| RA Group 2 (Administration by CG) CGs performed injection of 162 mg SC tocilizumab to participants with RA using the AI-1000 G2 device. Visit 1 (Day 0) was conducted for administration training, while Visits 2 and 3 (Days 14 and 28) were conducted for use/performance evaluation. | 17 |
| RA Group 3 (Administration by HCP) HCPs performed injection of 162 mg SC tocilizumab to participants with RA using the AI-1000 G2 device. Because enrolled HCPs were to be professionally qualified to deliver SC injections, no administration training was provided. Visit 1 (Day 0) was performed by the study nurse. Visits 2 and 3 (Days 14 and 28) were conducted by the HCP for use/performance evaluation. | 19 |
| Total | 90 |
Withdrawals & dropouts
| Period | Reason | FG000 | FG001 | FG002 | FG003 | FG004 |
|---|---|---|---|---|---|---|
| Overall Study | Adverse Event | 0 | 0 | 0 | 1 | 0 |
| Overall Study | Protocol Deviation | 0 | 1 | 0 | 0 | 0 |
| Overall Study | Withdrawal by Subject | 0 | 0 | 1 | 0 | 1 |
Baseline characteristics
| Characteristic | Caregivers | Healthcare Professionals | RA Group 1 (Self-Administration) | RA Group 2 (Administration by CG) | RA Group 3 (Administration by HCP) | Total |
|---|---|---|---|---|---|---|
| Age, Categorical <=18 years | 0 Participants | 0 Participants | 0 Participants | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical >=65 years | 2 Participants | 2 Participants | 4 Participants | 2 Participants | 3 Participants | 13 Participants |
| Age, Categorical Between 18 and 65 years | 15 Participants | 17 Participants | 14 Participants | 15 Participants | 16 Participants | 77 Participants |
| Sex: Female, Male Female | 12 Participants | 17 Participants | 13 Participants | 13 Participants | 13 Participants | 68 Participants |
| Sex: Female, Male Male | 5 Participants | 2 Participants | 5 Participants | 4 Participants | 6 Participants | 22 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk | EG002 affected / at risk |
|---|---|---|---|
| deaths Total, all-cause mortality | — / — | — / — | — / — |
| other Total, other adverse events | 4 / 18 | 4 / 17 | 3 / 19 |
| serious Total, serious adverse events | 0 / 18 | 0 / 17 | 0 / 19 |
Outcome results
Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use
Safety-critical tasks included those tasks where errors would have a reasonably foreseeable potential for clinical impact/harm, potentially resulting in direct physical injury to the user and/or conditions that require medical intervention. Safety-critical tasks included the following: release activation button; check the expiry date; inspect device prior to use; and inspect medication prior to use. Essential tasks included those essential to the execution of the injection. Essential tasks included the following: open the carton, remove the device and associated documents; remove cap; start an injection by depressing the needle-shield at the injection site and pressing the activation button; and hold the autoinjector until the complete dose has been delivered. The percentage of participants who successfully performed safety-critical and essential tasks during the first unassisted use (Day 14) was reported.
Time frame: Day 14
Population: Safety Population; only those who performed the administration task(s) on Day 14 were included because the endpoint was not applicable to non-injecting participants. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Remove cap | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Start injection | 93.75 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Release activation button | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Check expiry date | 88.24 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect device prior to use | 76.47 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect medication prior to use | 82.35 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Start injection | 94.12 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect device prior to use | 47.06 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Release activation button | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Check expiry date | 58.82 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Remove cap | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect medication prior to use | 82.35 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Start injection | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Remove cap | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect device prior to use | 66.67 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Check expiry date | 72.22 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Release activation button | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During First Unassisted Use | Inspect medication prior to use | 88.89 percentage of participants |
Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use
Safety-critical tasks included those tasks where errors would have a reasonably foreseeable potential for clinical impact/harm, potentially resulting in direct physical injury to the user and/or conditions that require medical intervention. Safety-critical tasks included the following: release activation button; check the expiry date; inspect device prior to use; and inspect medication prior to use. Essential tasks included those essential to the execution of the injection. Essential tasks included the following: open the carton, remove the device and associated documents; remove cap; start an injection by depressing the needle-shield at the injection site and pressing the activation button; and hold the autoinjector until the complete dose has been delivered. The percentage of participants who successfully performed safety-critical and essential tasks during the second unassisted use (Day 28) was reported.
Time frame: Day 28
Population: Safety Population; only those who performed the administration task(s) on Day 28 were included because the endpoint was not applicable to non-injecting participants.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Remove cap | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Start injection | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Release activation button | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Check expiry date | 94.44 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect device prior to use | 77.78 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect medication prior to use | 88.89 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Start injection | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect device prior to use | 58.82 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Release activation button | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Check expiry date | 76.47 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Remove cap | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect medication prior to use | 70.59 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Start injection | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Remove cap | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Open carton, remove device/documents | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Hold until complete dose delivered | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect device prior to use | 83.33 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Check expiry date | 94.44 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Release activation button | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Safety-Critical and Essential Tasks During Second Unassisted Use | Inspect medication prior to use | 94.44 percentage of participants |
Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA
Ability to perform daily living activities was assessed across eight component sets including dressing/grooming, arising, eating, walking, hygiene, reach, grip, and common activities. Twenty questions were scored on a 4-point Likert scale from 0 meaning without any difficulty to 3 meaning unable to do. The overall score was computed as the sum of domain scores divided by the number of domains answered. Therefore, the score range for HAQ-DI was the same as that of the individual questions, that is, from 0 to 3, where 0 indicates least difficulty and 3 indicates extreme difficulty. The mean change from baseline in HAQ-DI at each assessment was reported among participants with RA.
Time frame: Days 0, 14, 28
Population: Safety Population; only participants with RA were included because the endpoint was not applicable to CGs and HCPs. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| RA Group 1 (Self-Administration) | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D14 | 0.0962 units on a scale | Standard Deviation 0.2166 |
| RA Group 1 (Self-Administration) | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | D0 | 0.7222 units on a scale | Standard Deviation 0.6974 |
| RA Group 1 (Self-Administration) | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D28 | 0.0139 units on a scale | Standard Deviation 0.2674 |
| Caregivers | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D14 | -0.0368 units on a scale | Standard Deviation 0.1863 |
| Caregivers | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | D0 | 0.8676 units on a scale | Standard Deviation 0.6532 |
| Caregivers | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D28 | -0.0078 units on a scale | Standard Deviation 0.2519 |
| Healthcare Professionals | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | D0 | 1.2434 units on a scale | Standard Deviation 0.6541 |
| Healthcare Professionals | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D28 | -0.0903 units on a scale | Standard Deviation 0.3398 |
| Healthcare Professionals | Change From Baseline in Health Assessment Questionnaire-Disability Index (HAQ-DI) Among Participants With RA | Change at D14 | -0.0441 units on a scale | Standard Deviation 0.2575 |
Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA
Injection pain was assessed on a categorical 6-point Likert scale, ranging from no pain to severe and intolerable. The percentage of participants was reported by response at each assessment timepoint among participants with RA.
Time frame: 0 and 15 minutes after injection on Days 0, 14, 28
Population: Safety Population; only participants with RA were included because the endpoint was not applicable to CGs and HCPs. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, no pain | 70.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, no pain | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, minimal but tolerable | 23.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, no pain | 66.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, mild but tolerable | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, mild but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, minimal but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, mild but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, no pain | 88.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, minimal but tolerable | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, no pain | 88.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, minimal but tolerable | 33.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, no pain | 100.0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, minimal but tolerable | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, mild but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, mild but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, mild but tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, severe and intolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate-severe/tolerable | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, minimal but tolerable | 27.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, minimal but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, no pain | 47.1 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, minimal but tolerable | 41.2 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, mild but tolerable | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, severe and intolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, no pain | 100.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, minimal but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, mild but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, severe and intolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, no pain | 64.7 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, minimal but tolerable | 35.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, mild but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, severe and intolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, no pain | 100.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, minimal but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, mild but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, severe and intolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, no pain | 68.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, minimal but tolerable | 18.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, mild but tolerable | 12.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, severe and intolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, no pain | 100.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, mild but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate but tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, no pain | 52.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, no pain | 55.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, mild but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, minimal but tolerable | 33.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, mild but tolerable | 21.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, mild but tolerable | 5.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, moderate but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, no pain | 47.4 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate but tolerable | 5.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, mild but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, moderate but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, moderate-severe/tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, minimal but tolerable | 5.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, minimal but tolerable | 21.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 0min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 15min, no pain | 94.7 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, no pain | 82.4 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, no pain | 88.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, minimal but tolerable | 17.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate-severe/tolerable | 11.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, severe and intolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, mild but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, moderate but tolerable | 5.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate-severe/tolerable | 5.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate but tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, mild but tolerable | 11.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D28, 15min, minimal but tolerable | 11.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, moderate-severe/tolerable | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 0min, minimal but tolerable | 17.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D0, 0min, moderate but tolerable | 5.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Categorical Scale of Injection Pain Among Participants With RA | D14, 15min, severe and intolerable | 0 percentage of participants |
Percentage of Participants by Response to Device Satisfaction Questionnaire
Device satisfaction was assessed using twelve questions on a categorical 5-point Likert scale, ranging from strongly agree to strongly disagree. Question (Q) 1 (felt the autoinjection was easy to use), Q2 (felt comfortable while using the autoinjector), Q3 (felt the autoinjector was easy to hold), Q4 (able to tell when injection had completed), Q5 (felt he/she can inject properly with the autoinjector), Q6 (felt he/she had control over the injection process), Q7 (felt confident that he/she injected successfully), Q8 (felt it is easy to dispose of the autoinjector), Q9 (autoinjector would help to manage his/her injection schedule), Q10 (liked the look/feel of the autoinjector), Q11 (would recommend the autoinjector to someone else who needed to inject), Q12 (would continue having injections with the autoinjector). The percentage of participants was reported by response at each assessment among participants with RA, CGs, and HCPs.
Time frame: Days 0, 14, 28
Population: Safety Population. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, agree | 38.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, agree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly agree | 55.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly agree | 66.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, agree | 27.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, disagree (n=17,0,0,16,17) | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, neutral | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, neutral | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, agree | 33.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly agree | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, agree | 27.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly agree | 55.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, agree | 23.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, disagree | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly agree | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly agree | 64.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, neutral | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, agree | 27.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, strongly agree | 61.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, agree | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, agree | 17.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly agree | 76.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly agree | 70.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, disagree | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, agree | 44.4 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, strongly agree | 44.4 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, neutral | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, agree | 17.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, agree | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly agree | 58.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, agree | 22.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly agree | 77.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, neutral | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, agree | 22.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, strongly agree | 66.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, neutral | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, agree | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly agree | 76.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, agree | 22.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, strongly agree | 77.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, disagree | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, strongly agree | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, agree | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly agree | 76.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly disagree | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, agree | 23.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, disagree | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, strongly agree | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, neutral | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly agree | 70.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, agree | 17.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly agree | 70.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, disagree | 5.9 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, agree | 17.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly agree | 76.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, agree | 11.1 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, neutral | 5.6 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly agree | 82.4 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, agree | 35.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly agree | 83.3 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly agree | 58.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, agree | 16.7 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, neutral | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, agree | 27.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly agree | 76.5 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly agree | 72.2 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, agree | 11.8 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly disagree | 0 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, agree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly agree | 76.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, strongly agree | 88.2 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly agree | 76.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly agree | 75.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, agree | 18.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, agree | 17.6 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, neutral | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, strongly agree | 76.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, agree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, neutral | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly agree | 88.2 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly agree | 87.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, agree | 12.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, agree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly agree | 100.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly agree | 76.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, agree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly agree | 58.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, agree | 43.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D0, strongly agree | 94.1 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, agree | 35.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly agree | 56.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, agree | 6.7 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly agree | 81.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, agree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D0, strongly agree | 88.2 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly agree | 93.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, agree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, agree | 12.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, neutral | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D0, strongly agree | 88.2 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, agree | 18.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly agree | 68.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly agree | 81.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, neutral | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, strongly agree | 94.1 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, disagree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, agree | 12.5 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly agree | 75.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, agree | 17.6 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D0, strongly agree | 70.6 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, disagree | 6.7 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, agree | 13.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly agree | 80.0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, neutral | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D0, disagree | 5.9 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly disagree | 6.3 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, neutral | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, agree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D0, agree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly agree | 82.4 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, disagree | 11.8 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, agree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D0, strongly disagree | 0 percentage of participants |
| Caregivers | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D0, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, neutral | 5.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, agree | 36.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D14, strongly agree | 57.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, agree | 27.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q1, D28, strongly agree | 72.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, agree | 42.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D14, strongly agree | 57.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, agree | 27.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q2, D28, strongly agree | 72.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, agree | 31.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D14, strongly agree | 68.4 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, agree | 22.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q3, D28, strongly agree | 77.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, agree | 31.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D14, strongly agree | 68.4 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, agree | 16.7 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q4, D28, strongly agree | 83.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, agree | 15.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D14, strongly agree | 84.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, agree | 22.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q5, D28, strongly agree | 77.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, agree | 26.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D14, strongly agree | 73.7 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, agree | 22.2 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q6, D28, strongly agree | 77.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, agree | 21.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D14, strongly agree | 78.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, agree | 11.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q7, D28, strongly agree | 88.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, agree | 10.5 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D14, strongly agree | 89.5 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, agree | 11.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q8, D28, strongly agree | 88.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, agree | 31.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly agree | 68.4 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, neutral | 5.6 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, agree | 27.8 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly agree | 66.7 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, agree | 21.1 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly agree | 78.9 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, disagree | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, neutral | 0 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, agree | 33.3 percentage of participants |
| Healthcare Professionals | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly agree | 66.7 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, disagree | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly agree | 68.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, neutral | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly disagree | 14.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, disagree | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly agree | 64.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, neutral | 7.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, agree | 21.4 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, agree | 21.4 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly agree | 57.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, neutral | 7.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, agree | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly disagree | 7.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, neutral | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, neutral | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly agree | 68.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, agree | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, agree | 18.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly agree | 81.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, neutral | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, agree | 25.0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly disagree | 7.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, neutral | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly agree | 68.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly agree | 68.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, agree | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, agree | 18.8 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly agree | 81.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, neutral | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly disagree | 14.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, disagree (n=17,0,0,16,17) | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, neutral | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, agree | 21.4 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly agree | 64.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly agree | 57.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, neutral | 6.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, agree | 28.6 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, agree | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly agree | 81.3 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, neutral | 7.1 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly disagree | 0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, neutral | 12.5 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly agree | 75.0 percentage of participants |
| RA Group 2 (Administration by CG) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, agree | 18.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly agree | 52.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, neutral | 11.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly agree | 52.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, strongly agree | 50.0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, disagree | 5.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, neutral | 5.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, neutral | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly agree | 47.1 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, agree | 23.5 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, agree | 44.4 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, disagree | 5.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, neutral | 5.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, agree | 41.2 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, agree | 33.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, agree | 29.4 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, agree | 35.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, neutral | 23.5 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, neutral | 11.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly agree | 55.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D0, strongly agree | 58.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D28, disagree | 5.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, disagree | 5.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly agree | 55.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, neutral | 5.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D0, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, agree | 41.2 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, disagree (n=17,0,0,16,17) | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly agree | 44.4 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q10, D14, strongly agree | 41.2 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, neutral | 11.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, agree | 35.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, agree | 33.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, neutral | 5.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, neutral | 17.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, agree | 35.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, neutral | 5.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, neutral | 16.7 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D14, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D14, strongly agree | 52.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, agree | 35.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D14, strongly agree | 52.9 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, strongly disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D0, disagree | 0 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q9, D28, strongly disagree | 5.6 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D28, agree | 33.3 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q12, D0, strongly agree | 58.8 percentage of participants |
| RA Group 3 (Administration by HCP) | Percentage of Participants by Response to Device Satisfaction Questionnaire | Q11, D28, disagree | 5.6 percentage of participants |
Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use
Ancillary tasks included those tasks where the potential harm resulting from use error would be minor in severity, or the resultant harms were estimated to occur at sufficiently low levels. Ancillary tasks included the following: wash hands; clean the injection site with alcohol swab; wait for the alcohol to dry; dispose of the autoinjector cap; inspect full dose delivered after use; dispose of the autoinjector; and treatment of injection site after injection. The percentage of participants who succesffully performed ancillary tasks during the first (Day 14) and second unassisted use (Day 28) was reported.
Time frame: Days 14, 28
Population: Safety Population; only those who performed the administration task(s) were included because the endpoint was not applicable to non-injecting participants. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device | 93.33 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wait for alcohol to dry | 82.35 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, clean/swab injection site | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, inspect full dose delivered | 94.44 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device cap | 94.12 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wash hands | 88.89 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, inspect full dose delivered | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device cap | 94.44 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device | 94.44 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, clean/swab injection site | 94.44 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, treatment of site post injection | 100.00 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | Day (D) 14, wash hands | 82.35 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, wait for alcohol to dry | 94.12 percentage of participants |
| RA Group 1 (Self-Administration) | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, treatment of site post injection | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | Day (D) 14, wash hands | 82.35 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wash hands | 94.12 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, clean/swab injection site | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, clean/swab injection site | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, wait for alcohol to dry | 94.12 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wait for alcohol to dry | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device cap | 86.67 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device cap | 88.24 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, inspect full dose delivered | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, inspect full dose delivered | 94.12 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device | 94.12 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, treatment of site post injection | 100.00 percentage of participants |
| Caregivers | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, treatment of site post injection | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, treatment of site post injection | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, inspect full dose delivered | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, wait for alcohol to dry | 88.89 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, clean/swab injection site | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, clean/swab injection site | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | Day (D) 14, wash hands | 77.78 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wash hands | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, dispose of device cap | 94.44 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, dispose of device cap | 83.33 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, treatment of site post injection | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D14, inspect full dose delivered | 100.00 percentage of participants |
| Healthcare Professionals | Percentage of Participants Who Successfully Performed Ancillary Tasks During First and Second Unassisted Use | D28, wait for alcohol to dry | 83.33 percentage of participants |
Swollen Joint Count (SJC) Among Participants With RA
Sixty-six joints were assessed for swelling among participants with RA. The number of swollen joints at Baseline was reported.
Time frame: Baseline (Day 0)
Population: Safety Population; only participants with RA were included because the endpoint was not applicable to CGs and HCPs.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Swollen Joint Count (SJC) Among Participants With RA | 4.61 swollen joints | Standard Deviation 4.1 |
| Caregivers | Swollen Joint Count (SJC) Among Participants With RA | 3.71 swollen joints | Standard Deviation 4.45 |
| Healthcare Professionals | Swollen Joint Count (SJC) Among Participants With RA | 8.89 swollen joints | Standard Deviation 7.88 |
Tender Joint Count (TJC) Among Participants With RA
Sixty-eight joints were assessed for tenderness among participants with RA. The number of tender joints at Baseline was reported.
Time frame: Baseline (Day 0)
Population: Safety Population; only participants with RA were included because the endpoint was not applicable to CGs and HCPs.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| RA Group 1 (Self-Administration) | Tender Joint Count (TJC) Among Participants With RA | 11.17 tender joints | Standard Deviation 9.32 |
| Caregivers | Tender Joint Count (TJC) Among Participants With RA | 8.12 tender joints | Standard Deviation 11.29 |
| Healthcare Professionals | Tender Joint Count (TJC) Among Participants With RA | 16.32 tender joints | Standard Deviation 16.17 |
Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA
Injection pain was assessed on a continuous 100-millimeter (mm) VAS, where 0 mm represents no pain and 100 mm represents unbearable pain. The mean VAS response at each assessment timepoint was reported among participants with RA.
Time frame: 0 and 15 minutes after injection on Days 0, 14, 28
Population: Safety Population; only participants with RA were included because the endpoint was not applicable to CGs and HCPs. Here, n refers to number evaluable for the specified assessment, respectively.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 0 minutes (min) | 2.8 mm | Standard Deviation 4 |
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 15min | 1.2 mm | Standard Deviation 2.3 |
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 0min | 5.4 mm | Standard Deviation 12.9 |
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 15min | 1.5 mm | Standard Deviation 2.8 |
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 0min | 4.5 mm | Standard Deviation 5.2 |
| RA Group 1 (Self-Administration) | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 15min | 0.6 mm | Standard Deviation 1.5 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 15min | 1.9 mm | Standard Deviation 5.7 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 0 minutes (min) | 9.9 mm | Standard Deviation 13.7 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 15min | 0.5 mm | Standard Deviation 0.9 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 0min | 5.8 mm | Standard Deviation 9 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 15min | 0.9 mm | Standard Deviation 1.9 |
| Caregivers | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 0min | 5.0 mm | Standard Deviation 7 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 15min | 0.8 mm | Standard Deviation 1.9 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 0min | 11.4 mm | Standard Deviation 17.3 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 15min | 0.7 mm | Standard Deviation 1.2 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D14, 15min | 0.7 mm | Standard Deviation 1.4 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D0, 0 minutes (min) | 13.3 mm | Standard Deviation 22.9 |
| Healthcare Professionals | Visual Analog Scale (VAS) Score for Injection Pain Among Participants With RA | D28, 0min | 14.5 mm | Standard Deviation 20.1 |