Positive Predictive Value of the MolecuLight i:X Imaging Device to Predict the Presence of Bacteria in Chronic Wounds

A Prospective, Single-blind Evaluation of the Positive Predictive Value (PPV) of the MolecuLight i:X(TM) Imaging Device to Predict the Presence of Bacteria in Chronic Wounds

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02682069
Enrollment
30
Registered
2016-02-15
Start date
2016-02-29
Completion date
2017-08-30
Last updated
2020-02-26

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Wounds

Keywords

Diabetic foot ulcer, venous leg ulcer, fluorescence imaging, wound imaging

Brief summary

This is a non-randomized evaluation for which 50 patients will be imaged at the Southwest Regional Wound Care Center, Lubbock, Texas, who present with a chronic wound and are receiving standard treatment. The MolecuLight i:X Imaging Device will be used as an adjunctive tool in the assessment of the wound and will be used to guide the targeted sampling of a wound (punch biopsy method).

Detailed description

MolecuLight Inc. (Toronto, Canada) has recently introduced to the Canadian market an innovative imaging device, the MolecuLight i:X™ Imaging Device, that offers real-time detection of important biological and molecular information of a chronic wound, and could have significant impact on improving conventional wound care and management. The MolecuLight i:X Imaging Device is intended to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device can capture and document either an image or video of the chronic wound where the presence of fluorescent bacteria appears under violet light illumination. This information can be used to guide a clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present. Though the MolecuLight i:X Imaging Device has been shown to be effective in controlled settings, this study is deploying the device in a larger population and evaluating the ability of the MolecuLight i:X device to positively predict the presence of bacteria in chronic wounds. The overall objective of this work is to evaluate the MolecuLight i:X Imaging Device in screening of chronic wounds for the presence of bacteria. The device is intended to guide the clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present. All imaging will be performed with the MolecuLight i:X Imaging Device at an appropriate distance from the wound of interest (8 - 12 cm), which is indicated by the range finder LED. Illumination is provided by two violet (405 nm) LEDs that produce a bright, but clinically safe, uniform illumination. Fluorescence imaging is performed on the camera in real-time while the device is in fluorescence mode and the room lights are turned off. If room ambient light cannot be eliminated to an acceptable level (indicated by the ambient light sensor), the MolecuLight drape must be used to achieve a dark environment. A standard measurement sticker will be placed adjacent to the wound within the field of view to act as a reference for size. All microbiological analysis will be performed as per standard practice at each local clinic. Tissue samples will be analyzed for culture and sensitivity, yielding the species of bacteria present in the sample (if any), the susceptibility to various antibiotics and a measure of the bacterial load.

Interventions

The intended use of the device is to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device will be used as part of the current clinical wound assessment process which may include examination for characteristic signs and symptoms of infection. The device can capture and document either an image or video of the chronic wound where the presence of florescent bacteria appears under violet light illumination. This information can be used to guide a clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present.

Sponsors

Southwest Regional Wound Care Center
CollaboratorOTHER
MolecuLight Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SCREENING
Masking
SINGLE (Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Male or female patients presenting with chronic wound of the lower limb (e.g. diabetic foot ulcer, venous leg ulcer, other acceptable aetiology) 2. 18 years or older

Exclusion criteria

1. Treatment with an investigational drug within 1 month before study enrolment 2. Use of systemic (oral or intravenous) antibiotics 3. Inability to consent 4. Any contra-indication to routine wound care and/or monitoring

Design outcomes

Primary

MeasureTime frameDescription
Positive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and GreaterBaselinePPV reflects the probability that a region of red fluorescence within or around a wound will contain bacteria. Meaning the number cases where qPCR analysis of wound tissue biopsies from red fluorescent region showed to have pathogen load ≥ 104 CFU/g divided by the total number of cases where red florescence was observed in the wound multiplied by 100,

Participant flow

Participants by arm

ArmCount
All Patients
There is one arm to this study and all patients will undergo the same procedures. The test of this technology is on a per wound basis where bacteria will fluoresce red and samples will be obtained from the discrete red locations. Microbiology results indicating the presence or absence of bacteria will be correlated to the fluorescence signal in the fluorescent images.MolecuLight i:X Imaging Device: The intended use of the device is to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device will be used as part of the current clinical wound assessment process which may include examination for characteristic signs and symptoms of infection. The device can capture and document either an image or video of the chronic wound where the presence of florescent bacteria appears under violet light illumination.
30
Total30

Baseline characteristics

CharacteristicAll Patients
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
30 Participants
Region of Enrollment
United States
30 Participants
Sex: Female, Male
Female
6 Participants
Sex: Female, Male
Male
24 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 30
serious
Total, serious adverse events
0 / 30

Outcome results

Primary

Positive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and Greater

PPV reflects the probability that a region of red fluorescence within or around a wound will contain bacteria. Meaning the number cases where qPCR analysis of wound tissue biopsies from red fluorescent region showed to have pathogen load ≥ 104 CFU/g divided by the total number of cases where red florescence was observed in the wound multiplied by 100,

Time frame: Baseline

ArmMeasureValue (NUMBER)
All PatientsPositive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and Greater100 Percentage of PPV