Wounds
Conditions
Keywords
Diabetic foot ulcer, venous leg ulcer, fluorescence imaging, wound imaging
Brief summary
This is a non-randomized evaluation for which 50 patients will be imaged at the Southwest Regional Wound Care Center, Lubbock, Texas, who present with a chronic wound and are receiving standard treatment. The MolecuLight i:X Imaging Device will be used as an adjunctive tool in the assessment of the wound and will be used to guide the targeted sampling of a wound (punch biopsy method).
Detailed description
MolecuLight Inc. (Toronto, Canada) has recently introduced to the Canadian market an innovative imaging device, the MolecuLight i:X™ Imaging Device, that offers real-time detection of important biological and molecular information of a chronic wound, and could have significant impact on improving conventional wound care and management. The MolecuLight i:X Imaging Device is intended to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device can capture and document either an image or video of the chronic wound where the presence of fluorescent bacteria appears under violet light illumination. This information can be used to guide a clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present. Though the MolecuLight i:X Imaging Device has been shown to be effective in controlled settings, this study is deploying the device in a larger population and evaluating the ability of the MolecuLight i:X device to positively predict the presence of bacteria in chronic wounds. The overall objective of this work is to evaluate the MolecuLight i:X Imaging Device in screening of chronic wounds for the presence of bacteria. The device is intended to guide the clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present. All imaging will be performed with the MolecuLight i:X Imaging Device at an appropriate distance from the wound of interest (8 - 12 cm), which is indicated by the range finder LED. Illumination is provided by two violet (405 nm) LEDs that produce a bright, but clinically safe, uniform illumination. Fluorescence imaging is performed on the camera in real-time while the device is in fluorescence mode and the room lights are turned off. If room ambient light cannot be eliminated to an acceptable level (indicated by the ambient light sensor), the MolecuLight drape must be used to achieve a dark environment. A standard measurement sticker will be placed adjacent to the wound within the field of view to act as a reference for size. All microbiological analysis will be performed as per standard practice at each local clinic. Tissue samples will be analyzed for culture and sensitivity, yielding the species of bacteria present in the sample (if any), the susceptibility to various antibiotics and a measure of the bacterial load.
Interventions
The intended use of the device is to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device will be used as part of the current clinical wound assessment process which may include examination for characteristic signs and symptoms of infection. The device can capture and document either an image or video of the chronic wound where the presence of florescent bacteria appears under violet light illumination. This information can be used to guide a clinician to inspect, sample, debride or further evaluate areas within or around a wound where fluorescent bacteria are present.
Sponsors
Study design
Eligibility
Inclusion criteria
1. Male or female patients presenting with chronic wound of the lower limb (e.g. diabetic foot ulcer, venous leg ulcer, other acceptable aetiology) 2. 18 years or older
Exclusion criteria
1. Treatment with an investigational drug within 1 month before study enrolment 2. Use of systemic (oral or intravenous) antibiotics 3. Inability to consent 4. Any contra-indication to routine wound care and/or monitoring
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Positive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and Greater | Baseline | PPV reflects the probability that a region of red fluorescence within or around a wound will contain bacteria. Meaning the number cases where qPCR analysis of wound tissue biopsies from red fluorescent region showed to have pathogen load ≥ 104 CFU/g divided by the total number of cases where red florescence was observed in the wound multiplied by 100, |
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| All Patients There is one arm to this study and all patients will undergo the same procedures. The test of this technology is on a per wound basis where bacteria will fluoresce red and samples will be obtained from the discrete red locations. Microbiology results indicating the presence or absence of bacteria will be correlated to the fluorescence signal in the fluorescent images.MolecuLight i:X Imaging Device: The intended use of the device is to assist clinicians during care and management of patients with chronic wounds by screening for the presence of potentially harmful bacteria levels. The device will be used as part of the current clinical wound assessment process which may include examination for characteristic signs and symptoms of infection. The device can capture and document either an image or video of the chronic wound where the presence of florescent bacteria appears under violet light illumination. | 30 |
| Total | 30 |
Baseline characteristics
| Characteristic | All Patients |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 0 Participants |
| Age, Categorical Between 18 and 65 years | 30 Participants |
| Region of Enrollment United States | 30 Participants |
| Sex: Female, Male Female | 6 Participants |
| Sex: Female, Male Male | 24 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 0 / 30 |
| serious Total, serious adverse events | 0 / 30 |
Outcome results
Positive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and Greater
PPV reflects the probability that a region of red fluorescence within or around a wound will contain bacteria. Meaning the number cases where qPCR analysis of wound tissue biopsies from red fluorescent region showed to have pathogen load ≥ 104 CFU/g divided by the total number of cases where red florescence was observed in the wound multiplied by 100,
Time frame: Baseline
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| All Patients | Positive Predictive Value (PPV) - Ability of Fluorescence Imaging With the MolecuLight i:X to Predict Presence of Bacterial Loads of 104 CFU/g and Greater | 100 Percentage of PPV |