Skip to content

The Esteem® Hearing Implant Post Approval Study

To Evaluate the Long-term Safety and Efficacy of the Esteem® Hearing Implant in Subjects Suffering From Moderate to Severe Hearing Loss

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02666222
Enrollment
51
Registered
2016-01-28
Start date
2010-08-31
Completion date
2015-03-31
Last updated
2016-05-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss

Keywords

hearing loss, sensorineural, hearing implant

Brief summary

The purpose of this study is to evaluate the long-term safety and efficacy of the Esteem® Hearing Implant in subjects suffering from moderate to severe hearing loss.

Detailed description

The Post Approval Study is an extension of the pivotal clinical trial, which was designed as a prospective, multi-center, non-randomized, clinical trial to evaluate the safety and efficacy of the Esteem® System. For this trial the subject acts as his or her own control. This trial has been designed to meet the United States' regulatory requirements. The subjects included in this study include the 57 subjects implanted under the original Investigational Device Exemption (IDE) study and the five (5) subjects implanted under the continued access expansion approved by the FDA. Of these 62 subjects, all but one (explant) were available for continued follow-up. These subjects were re-consented for the 5 year Post Approval Study. No new subjects were enrolled and implanted in this study. The goal was to monitor a minimum of 45 subjects through their five (5) year follow-up. Ultimately, 51 subjects completed the study. The following questions are to be answered: * Is the Esteem effective through 5 year follow-up? * Is the Esteem safe through 5 year follow-up?

Interventions

DEVICEThe Esteem Hearing Implant

The Esteem® Hearing Implant is designed to improve hearing in subjects suffering from moderate to severe hearing loss that is sensorineural in origin. The Esteem® has been approved by FDA for the US market under Pre-Market Approval P090018 and this study fulfills the requirement for one of the Post Approval Studies.

Sponsors

Envoy Medical Corporation
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 90 Years
Healthy volunteers
Yes

Inclusion criteria

Subject must meet all of the following criteria to be eligible for treatment in the trial: 1. Subject is at least 18 years old 2. Subject understands the nature of the procedure and has signed the Subject Informed Consent Form prior to the procedure 3. Subject is willing and able to comply with specified follow-up evaluations and understands the audiological test procedures and use of the Esteem® System. 4. Subject has moderate to severe sensorineural hearing loss between 500 and 4000 Hz in the ear to be implanted with pure tone air-conduction threshold levels within the limits of a Hearing Aid (HA) as follow: Freq (Hz) 500 1000 2000 3000 4000 Lower Level (dB HL) 30 35 35 35 35 Upper Level (dB HL) 100 100 100 100 100 5. Subject's air-bone gap is no greater than 10 dB at 4 of the 5 following frequencies: 500, 1000, 2000, 3000 and 4000 Hz. 6. Subject has an unaided maximum word recognition score of greater than or equal to 40% with recorded delivery using a phonetically balanced word list at Speech Reception Threshold (SRT) + 40 dB or at maximum tolerable presentation level. 7. Subject is a current user of a properly functioning and appropriately fit hearing aid. This is defined as the subject has used this aid for at least four (4) hours (average) per day (in the ear to be implanted) for at least three (3) months for a new aid or one (1) month for an adjusted aid. 8. Subject's hearing aid, in the ear to be implanted, shall appropriately fit optimally. 9. Subject has normally functioning eustachian tube 10. Subject has normal tympanic membrane 11. Subject has a normal middle ear anatomy 12. Subject has adequate space for Esteem® System implant determined via fine cut temporal bone CT scan 13. Subject is a native speaker of the English language. 14. Subject is a hearing aid user in the ear to be implanted

Exclusion criteria

Subjects will be excluded from the trial if any one of the following criteria is met: 1. Subject has a history of post-adolescent chronic middle ear infections, inner ear disorders or recurring vertigo requiring treatment, disorders such as mastoiditis, Hydrops or Meniere's syndrome or disease 2. Subject has a history of external otitis or eczema for the outer ear canal and the investigator believes this will affect the Esteem® System implantation 3. Subject has cholesteatoma or destructive middle ear disease 4. Subject has life expectancy of less than two (2) years due to other medical conditions 5. Subject has retrocochlear or central auditory disorders 6. Subject is known to be suffering from any psychological, developmental, physical, or emotional disorder that the investigator feels would interfere with the surgery or follow-up testing 7. Subject has a known history of fluctuating air conduction and/or bone conduction hearing loss over a one-year period of 15 dB in either direction at 2 or more frequencies (from 500 - 4000 Hz) 8. Subject has sudden hearing loss due to unknown cause 9. Subject has a history of disabling tinnitus, defined as tinnitus which required treatment. 10. Subject is unable to adequately perform audiological testing 11. Subject has a medical condition or undergoing a treatment that may affect healing and the investigator does not believe the subject is a good candidate for the trial. 12. Subject has diabetes that is not well controlled with medication or diet and the investigator does not believe in his best medical judgment that the subject would be a good candidate for the trial 13. Subject is pregnant at the time of device implant 14. Subject has a history of keloid formation 15. Subject has known hypersensitivity to silicone rubber, polyurethane, stainless steel, titanium and/or gold

Design outcomes

Primary

MeasureTime frameDescription
Change From Baseline (Pre-implant Aided Condition) at Year 5 in Speech Reception Threshold (SRT)Baseline through Year 5 of Follow UpENDPOINT #1: SRT at baseline minus SRT at Year 5. Positive difference (i.e. lower value of SRT with the Esteem) indicates better outcome.
Change From Baseline (Pre-implant Aided Condition) at Year 5 in Word Recognition Score (WRS) at 50 dB HLBaseline through Year 5 of Follow UpENDPOINT #2: WRS at Year 5 minus WRS at baseline. Positive difference (in % correct) indicates better outcome.
Incidence of Serious Adverse Device Events (SADEs) and Device Failures and Replacements at Each Follow-up.SADEs, PAS phase through Year 5 of Follow UpENDPOINT #3: The analysis of the incidence of SADEs and device failures and replacements at each follow-up.
Bone Conduction StabilityBaseline through 5 Year Follow-UpENDPOINT #5: Difference between Baseline and 5 Year Pure-Tone Average (PTA; average of 500, 1000, 2000 Hz thresholds); calculated as PTA at Year 5 minus PTA at baseline. Smaller magnitude dB difference indicates better outcome.

Secondary

MeasureTime frameDescription
Improvement in Quality of Life as Reflected by Abbreviated Profile of Hearing Aid Benefit (APHAB) QuestionnaireBaseline through Year 5 of Follow UpAPHAB results were obtained from baseline aided condition through Year 5 of follow up. The APHAB responses are in terms of percent of time an individual experiences problems, on a scale of 0-100%; lower scores indicate fewer problems.Compared scores with Esteem to scores in baseline aided condition, calculated as APHAB Global score at baseline minus APHAB Global score at Year 5, giving a difference in benefit score. The Global Score is the mean of the scores (% of problems) for Ease of Communication (EC), Reverberation (RV), and Background Noise (BN) subscales of the APHAB. A positive difference in benefit score indicates more benefit with Esteem.

Countries

United States

Participant flow

Participants by arm

ArmCount
Esteem Implant
Subjects implanted with the Esteem Totally Implantable Hearing System
51
Total51

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyDeath2

Baseline characteristics

CharacteristicEsteem Implant
Age, Continuous56.1 years
STANDARD_DEVIATION 15.6
Race/Ethnicity, Customized
Asian
1 participants
Race/Ethnicity, Customized
Black/Non-Hispanic
0 participants
Race/Ethnicity, Customized
Hispanic
0 participants
Race/Ethnicity, Customized
White/Caucasian
50 participants
Sex: Female, Male
Female
17 Participants
Sex: Female, Male
Male
34 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
42 / 51
serious
Total, serious adverse events
14 / 51

Outcome results

Primary

Bone Conduction Stability

ENDPOINT #5: Difference between Baseline and 5 Year Pure-Tone Average (PTA; average of 500, 1000, 2000 Hz thresholds); calculated as PTA at Year 5 minus PTA at baseline. Smaller magnitude dB difference indicates better outcome.

Time frame: Baseline through 5 Year Follow-Up

Population: Difference between Baseline and 5 Year PTA (average of 500, 1000, 2000 Hz thresholds)

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantBone Conduction Stability3.7 dBStandard Error 1.1
Primary

Change From Baseline (Pre-implant Aided Condition) at Year 5 in Speech Reception Threshold (SRT)

ENDPOINT #1: SRT at baseline minus SRT at Year 5. Positive difference (i.e. lower value of SRT with the Esteem) indicates better outcome.

Time frame: Baseline through Year 5 of Follow Up

Population: 5 Year Follow-up

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantChange From Baseline (Pre-implant Aided Condition) at Year 5 in Speech Reception Threshold (SRT)4.8 dBStandard Error 2
Primary

Change From Baseline (Pre-implant Aided Condition) at Year 5 in Word Recognition Score (WRS) at 50 dB HL

ENDPOINT #2: WRS at Year 5 minus WRS at baseline. Positive difference (in % correct) indicates better outcome.

Time frame: Baseline through Year 5 of Follow Up

Population: 5 Year Follow-up Visit

ArmMeasureValue (MEAN)Dispersion
Esteem ImplantChange From Baseline (Pre-implant Aided Condition) at Year 5 in Word Recognition Score (WRS) at 50 dB HL17 percentage of correct responsesStandard Error 4.2
Primary

Incidence of Serious Adverse Device Events (SADEs) and Device Failures and Replacements at Each Follow-up.

ENDPOINT #3: The analysis of the incidence of SADEs and device failures and replacements at each follow-up.

Time frame: SADEs, PAS phase through Year 5 of Follow Up

Population: Cumulative through 5-Year Follow-up

ArmMeasureValue (NUMBER)
Esteem ImplantIncidence of Serious Adverse Device Events (SADEs) and Device Failures and Replacements at Each Follow-up.3 participants
Secondary

Improvement in Quality of Life as Reflected by Abbreviated Profile of Hearing Aid Benefit (APHAB) Questionnaire

APHAB results were obtained from baseline aided condition through Year 5 of follow up. The APHAB responses are in terms of percent of time an individual experiences problems, on a scale of 0-100%; lower scores indicate fewer problems.Compared scores with Esteem to scores in baseline aided condition, calculated as APHAB Global score at baseline minus APHAB Global score at Year 5, giving a difference in benefit score. The Global Score is the mean of the scores (% of problems) for Ease of Communication (EC), Reverberation (RV), and Background Noise (BN) subscales of the APHAB. A positive difference in benefit score indicates more benefit with Esteem.

Time frame: Baseline through Year 5 of Follow Up

Population: 5 Year Follow-up

ArmMeasureValue (MEAN)
Esteem ImplantImprovement in Quality of Life as Reflected by Abbreviated Profile of Hearing Aid Benefit (APHAB) Questionnaire5.6 difference score in units on a scale

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026