Optical System to Continuously Measure Arterial Blood Pressure

Comparison of an Optical System to Continuously Measure Arterial Blood Pressure Against a Gold-standard Arterial Catheter

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02651558
Acronym
BP
Enrollment
70
Registered
2016-01-11
Start date
2016-04-30
Completion date
2017-11-01
Last updated
2018-04-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure

Brief summary

Comparison of an optical method to continuously measure blood pressure against an invasive arterial catheter.

Detailed description

The measurement of blood pressure during anesthesia is commonly performed by the inflation of a brachial cuff providing only intermittent blood pressure measurements. In some case, it is required to add a continuous invasive monitoring of blood pressure, via the insertion of an arterial catheter. Unfortunately, the related morbidity is not negligible. To limit the consequences of such a gesture, some developments have been done in the past to allow for continuous non-invasive measurements of blood pressure during anesthesia. Unfortunately, most of these devices are subject to important limitations and constraints of use. A new system has been designed by CSEM consisting of an optical system fixed on patient's skin, absolutely non-invasive and easy to use. In order to study the reliability of this device, the investigators plan to compare its blood pressure estimates against gold-standard arterial catheter measurements during induction of general anesthesia on: * 40 patients wearing the optical system at the chest. * and 40 patients wearing the optical system at the fingertip

Interventions

DEVICEExperimental: OBPM 2015-MD-0022 - CHEST

At the onset of general anesthesia: 1. continuous measurements of radial arterial blood pressure values (arterial catheter, via a commercial PHILIPS IntelliVue MP50 monitor) 2. continuous estimates of optical blood pressure values at the chest (optical blood pressure system, via Swissmedic approved OBPM measuring device 2015-MD-0022 developed by CSEM) will be recorded.

DEVICEExperimental: OBPM 2015-MD-0022 - CHEST & FINGER

At the onset of general anesthesia: 1. continuous measurements of radial arterial blood pressure values (arterial catheter, via a commercial PHILIPS IntelliVue MP50 monitor) 2. continuous estimates of optical blood pressure values at the chest (optical blood pressure system, via Swissmedic approved OBPM measuring device 2015-MD-0022 developed by CSEM) 3. continuous estimates of optical blood pressure values at the fingertip (optical blood pressure system, via Swissmedic approved OBPM measuring device 2015-MD-0022 developed by CSEM) will be recorded.

DEVICEExperimental: OBPM 2015-MD-0022 - FINGER

At the onset of general anesthesia: 1. continuous measurements of radial arterial blood pressure values (arterial catheter, via a commercial PHILIPS IntelliVue MP50 monitor) 2. continuous estimates of optical blood pressure values at the fingertip (optical blood pressure system, via Swissmedic approved OBPM measuring device 2015-MD-0022 developed by CSEM) will be recorded.

Sponsors

Centre Hospitalier Universitaire Vaudois
CollaboratorOTHER
CSEM Centre Suisse d'Electronique et de Microtechnique SA - Recherche et Developpement
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* patient requiring the placement of an arterial catheter for general anesthesia * minimum of 18 years * written consent

Exclusion criteria

* minor patient or incapable of discernment * patient with AIMD * thoracic surgery

Design outcomes

Primary

MeasureTime frameDescription
Error (in mmHg) between paired determinations of arterial blood pressure from an optical device against reference measurements from an arterial catheter24 monthsThe signals recorded by the optical blood pressure device will be retrospectively analysed (offline analysis) in order to test the hypothesis that blood pressure values as derived by the new device is comparable to blood pressure values as measured by an arterial catheter within the limits of agreement in mmHg stated by ISO 81060-2 2013.

Countries

Switzerland

Outcome results

None listed