Quantitative Characterization of Safe Irrigation for Ventricular Shunt Catheters

Quantitative Characterization of Safe Irrigation for Ventricular Shunt Catheters

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02651337
Enrollment
4
Registered
2016-01-11
Start date
2016-04-01
Completion date
2016-06-14
Last updated
2020-09-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hydrocephaly

Brief summary

The objective of this pilot study is to perform a set of pressure and flow rate measurements in order to quantitatively characterize safe irrigation or flushing of a ventricular shunt catheter, a routine maneuver performed by a neurosurgeon during a shunt revision surgery. This maneuver involves injecting saline with a syringe to clear plugged shunt catheters. Based on data obtained in Stage I of the study, a device has been fabricated which provides the same small-volume pulse of fluid used by surgeons. In Stage II of the study, pressure and flow rate measurements generated by this specialized flusher device will be collected and measured.

Detailed description

Primary Objective: To collect pressure and flow rate measurements during shunt revision surgeries in human patients. Pressure recordings of the injected fluid will be obtained using disposable off-the-shelf sterile pressure transducers (which are typically used for intracranial and arterial pressures monitoring) readily available in the operating room. First, the surgeon will attempt to flush the catheter using the Alcyone in-line Flusher device. If this fails, the surgeon may elect to attempt to flush using a standard syringe attached to the catheter. The flow rates will be measured by video, recording the syringe during the flushing maneuver, and then extracting plunger motion information by applying image processing on the acquired videos. In the event the recording equipment is not available (in that these operations are usually unscheduled and sometimes urgent), the same irrigation techniques may be used without the pressure recordings. Secondary Objective: To analyze the collected data and establish maximal allowable injection pressure and flow rate. The collected data segments for each patient will be analyzed and maximal values for pressure and the saline injection rate will be recorded in a spreadsheet. Inter-patient variation in the results, presumably due to different degrees of occlusion, will be statistically evaluated. We will also determine if any of the catheters which appear occluded in the operating room resume flow with the flushing maneuvers.

Interventions

DEVICECerebrospinal Fluid (CSF) Drainage

Placement of Alivio in line flusher during CSF drainage

Sponsors

Alcyone Therapeutics, Inc
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
30 Days to 35 Years
Healthy volunteers
No

Inclusion criteria

1. Subjects must be 30 days to 35 years of age, of any race or gender. 2. Subjects must be scheduled to undergo a neurosurgical procedure potentially requiring a revision of neurosurgical catheters. 3. Subjects aged \>18 years of age must be willing and able to provide written informed consent. Parental consent will be required for minors.

Exclusion criteria

1. Inability or unwillingness of subject or parent/guardian to give informed consent/assent. 2. Judgment of the investigator, surgeon, or any member of the care team (including anesthesiology or operating room nursing staff) that participation in the study will interfere with, or be detrimental to, administration of optimal healthcare of the subject. 3. Inability or unwillingness of subject to endure the test or any other condition that would preclude or bias the results of the study according to the judgment of the investigator.

Design outcomes

Primary

MeasureTime frameDescription
To Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .Intra-proceduralCriteria for success: Investigator Determination if the Flusher can be properly primed. Investigator Determination if the Flusher can be easily connected to current shunt components. Investigator Determination if the Flusher can be activated. The following scenarios were also assessed: A noticeable resistance to compression/collapse of the dome to indicate that the Ventricular Catheter is too tightly blocked to be flushed. If the dome does not refill the Ventricular Catheter may still be non-flowing. If the dome refills, the flushing procedure may be repeated as needed and/or as determined reasonable by the surgeon.

Countries

United States

Participant flow

Recruitment details

5 participants consented for study participation, 4 patients were enrolled

Participants by arm

ArmCount
Drainage With Alivio In-line Flusher
Volume of saline needed to drain CSF using the Alivio in-line Flusher historically compared with volume of saline needed to drain CSF using a standard syringe Cerebrospinal Fluid (CSF) Drainage: Placement of Alivio in line flusher during CSF drainage
5
Total5

Baseline characteristics

CharacteristicDrainage With Alivio In-line Flusher
Age, Categorical
<=18 years
3 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
2 Participants
Age, Continuous15 years
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
5 Participants
Sex: Female, Male
Female
5 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 5
other
Total, other adverse events
0 / 5
serious
Total, serious adverse events
0 / 5

Outcome results

Primary

To Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .

Criteria for success: Investigator Determination if the Flusher can be properly primed. Investigator Determination if the Flusher can be easily connected to current shunt components. Investigator Determination if the Flusher can be activated. The following scenarios were also assessed: A noticeable resistance to compression/collapse of the dome to indicate that the Ventricular Catheter is too tightly blocked to be flushed. If the dome does not refill the Ventricular Catheter may still be non-flowing. If the dome refills, the flushing procedure may be repeated as needed and/or as determined reasonable by the surgeon.

Time frame: Intra-procedural

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Alivio Ventricular Flusher DeviceTo Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .Ability to Prime the Flusher4 Participants
Alivio Ventricular Flusher DeviceTo Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .Adequate Flusher Connection to Shunt4 Participants
Alivio Ventricular Flusher DeviceTo Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .System Flushed by dome compression4 Participants
Alivio Ventricular Flusher DeviceTo Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .Dome able to be Refilled4 Participants
Alivio Ventricular Flusher DeviceTo Demonstrate That the Flusher Can be Primed, Connected, and Activated in a Sterile Clinical Setting Per the Instructions for Use by the Treating Investigator .Dome able to be Evacuated3 Participants