Electromagnetic Interference
Conditions
Brief summary
This study is designed to compare a new prototype surgical magnetic drape made with bottom-isolated ferrite magnets to 4 commercially available surgical magnetic drapes.
Detailed description
Magnetic fields may potentially interfere with the function of cardiovascular implantable electronic devices. Sterile magnetic drapes used to hold surgical instruments are often placed on the patient's thorax, and they are likely to interfere with the function of these devices. This study compares the magnetic interference of a new prototype surgical magnetic drape made with bottom-isolated ferrite magnets to the Covidien magnetic drape and 3 other commercially available surgical magnetic drapes.
Interventions
New prototype surgical magnetic drape made with bottom-isolated ferrite magnets applied to the patient's thorax
Green Covidien Magnetic Drape applied to the patient's thorax
Jaa-Cell Medic Reusable Magnetic Pad applied to the patient's thorax
Magnetic Instrument Pad #25-002, size 20 x 16, DeRoyal Industries Inc applied to the patient's thorax
Magnetic Instrument Pad #25-001, size 10 x 16, DeRoyal Industries Inc applied to the patient's thorax
Sponsors
Study design
Eligibility
Inclusion criteria
* Patients with a pacemaker * 18 years or older
Exclusion criteria
* Patients with implantable cardioverter-defibrillator (ICD) * Patients in whom the magnet rate cardiac rhythm is indistinguishable from their baseline rhythm * Refusal/Inability to consent * Language other than French or English
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Percentage of magnetic interference | Within 2 minutes after placement of surgical magnetic drape to the patient's chest | Percentage of patients experiencing magnetic interference with the application of each surgical magnetic drape. Magnetic interference defined as a cardiac rhythm change to asynchronous and identical to that produced by a round magnet rhythm over the pacemaker. |