Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions

Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02614768
Enrollment
12
Registered
2015-11-25
Start date
2015-07-31
Completion date
2015-09-30
Last updated
2023-05-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes Mellitus, Type 1

Keywords

Glucose Monitoring System

Brief summary

The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom G4 Platinum Sensor in patients with type 1 Diabetes.

Detailed description

The study is a single-center open-label study in patients with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII) or multiple daily injections (MDI) therapy. The study will include a total of 12 subjects.

Interventions

The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled for investigative use only.

Sponsors

Medical University of Graz
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
BASIC_SCIENCE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Informed consent obtained after being advised of the nature of the study * Male or female aged ≥18 years * Type 1 diabetes for at least 6 months according to the World Health Organization (WHO) definition * Treatment with multiple daily injections (MDI) or continuous subcutaneous insulin infusion (CSII) for at least 3 months * Body Mass Index (BMI) \< 35 kg/m² * Willing and able to wear 5 CGM devices for the duration of the study and undergo all study procedures. * HbA1c ≤ 86 mmol/mol

Exclusion criteria

* Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the subject * Female of childbearing potential who is pregnant, breast-feeding or intend to become pregnant or is not using adequate contraceptive methods * Any mental condition rendering the subject incapable of giving his consent * Subject is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration * Subject may not use acetaminophen (paracetamol) while participating in the study * Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the subject's safety or successful participation in the study. * Subject is actively enrolled in another clinical trial * Known adrenal gland problem, pancreatic tumour, or insulinoma * Inability of the subject to comply with all study procedures * Inability of the subject to understand the patient information. * Subject donated blood in the last 3 months

Design outcomes

Primary

MeasureTime frameDescription
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%36 hoursThe primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.

Participant flow

Participants by arm

ArmCount
Glucose Sensor
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled for investigative use only.
12
Total12

Baseline characteristics

CharacteristicGlucose Sensor
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
12 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
12 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Region of Enrollment
Europe
12 participants
Sex: Female, Male
Female
5 Participants
Sex: Female, Male
Male
7 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 12
other
Total, other adverse events
3 / 12
serious
Total, serious adverse events
0 / 12

Outcome results

Primary

Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%

The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.

Time frame: 36 hours

ArmMeasureGroupValue (NUMBER)
Glucose SensorPercentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%Abbot73.2 Percentage of sensor values
Glucose SensorPercentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%Dexcom56.1 Percentage of sensor values
Glucose SensorPercentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%Medtronic52.0 Percentage of sensor values

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026