Brief summary
The purpose of this study was to determine if untrained subjects with diabetes could perform basic tasks with the Styx lancing device to obtain adequate blood volume from fingerstick and palm lancings for Blood Glucose Meter testing.
Interventions
Untrained Persons With Diabetes used the Styx Lancing Device with 28 and 30 Gauge lancets to obtain fingerstick and Alternate Site palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all.
Sponsors
Study design
Eligibility
Inclusion criteria
* Males and females, 18 years of age and older * People with type 1 or type 2 diabetes * People who regularly perform self-tests (at least once/day) with their own lancing device * Able to speak, read, and understand English * Willing to complete all study procedures
Exclusion criteria
* Those with missing tips of the fingers indicated in the protocol or physically unable to lance the palm areas and all the fingers indicated in the protocol. * Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual or neurological impairments that would make the person unable to perform testing * Persons who have used the Styx device in a previous clinical evaluation * Working for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 28 Gauge Lancets) | 1 hour | Untrained subjects with Diabetes operated the Styx Lancing Device with 28 Gauge lancets to obtain fingerstick capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 28 Gauge Lancets) | 1 hour | Untrained subjects with Diabetes operated the Styx Lancing Device with 28 Gauge lancets to obtain AST palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined. |
| Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 30 Gauge Lancets) | 1 hour | Untrained subjects with Diabetes operated the Styx Lancing Device with 30 Gauge lancets to obtain fingerstick capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined. |
| Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 30 Gauge Lancets) | 1 hour | Untrained subjects with Diabetes operated the Styx Lancing Device with 30 Gauge lancets to obtain AST palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined. |
| Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | 1 hour | Staff obtained responses from persons with Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the Styx Lancing Device. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. The percent of subjects who provided responses that were 'Strongly Agree' or 'Agree' or 'Neutral' about each statement was calculated. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Persons With Diabetes Untrained Persons with Diabetes used the Styx Lancing Device System to obtain fingerstick and Alternate Site palm capillary blood.
Styx Lancing Device: Untrained Persons With Diabetes used the Styx Lancing Device with 28 and 30 Gauge lancets to obtain fingerstick and Alternate Site palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. | 119 |
| Total | 119 |
Baseline characteristics
| Characteristic | Persons With Diabetes |
|---|---|
| Age, Continuous | 55.1 years |
| Ethnicity (NIH/OMB) Hispanic or Latino | 3 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 110 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 6 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 3 Participants |
| Race (NIH/OMB) Asian | 1 Participants |
| Race (NIH/OMB) Black or African American | 14 Participants |
| Race (NIH/OMB) More than one race | 1 Participants |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 0 Participants |
| Race (NIH/OMB) Unknown or Not Reported | 3 Participants |
| Race (NIH/OMB) White | 97 Participants |
| Region of Enrollment United States | 119 participants |
| Sex: Female, Male Female | 61 Participants |
| Sex: Female, Male Male | 58 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 1 / 119 |
| serious Total, serious adverse events | 0 / 119 |
Outcome results
Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 28 Gauge Lancets)
Untrained subjects with Diabetes operated the Styx Lancing Device with 28 Gauge lancets to obtain fingerstick capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined.
Time frame: 1 hour
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 28 Gauge Lancets) | 119 participants |
Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 28 Gauge Lancets)
Untrained subjects with Diabetes operated the Styx Lancing Device with 28 Gauge lancets to obtain AST palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined.
Time frame: 1 hour
Population: 118 (119-1) Subject results were analyzed. One subject result was not evaluable because staff deviated from protocol by interfering with subject operation of the device.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 28 Gauge Lancets) | 116 participants |
Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 30 Gauge Lancets)
Untrained subjects with Diabetes operated the Styx Lancing Device with 30 Gauge lancets to obtain AST palm capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined.
Time frame: 1 hour
Population: 117 (119-2) Subject results were analyzed. Two subject results were not evaluable because staff deviated from protocol by interfering with subjects' operation of the device.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subjects With Numeric Meter Results When Users Obtain Alternate Site (AST) Palm Blood Using the Styx Lancing Device ( 30 Gauge Lancets) | 114 participants |
Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 30 Gauge Lancets)
Untrained subjects with Diabetes operated the Styx Lancing Device with 30 Gauge lancets to obtain fingerstick capillary blood. Study Staff tested the capillary blood using Contour NEXT Blood Glucose Meters (BGMs) and recorded the results. Meter results could be numeric, non-numeric (ie, error messages), or not obtained at all. The number of subjects with numeric BGMS results was determined.
Time frame: 1 hour
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subjects With Numeric Meter Results When Users Obtain Fingerstick Capillary Blood Using the Styx Lancing Device ( 30 Gauge Lancets) | 119 participants |
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device
Staff obtained responses from persons with Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the Styx Lancing Device. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. The percent of subjects who provided responses that were 'Strongly Agree' or 'Agree' or 'Neutral' about each statement was calculated.
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I can follow the device instructional insert | 98.3 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I find the lancing device easy to use | 96.6 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I find it easy to insert the lancet into the devic | 99.2 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I can attach the Black Adjustable Endcap | 100 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I am able to do a fingerstick with this device | 100 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I can remove the Clear AST endcap | 100 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | Easy to remove lancet without removing w fingers | 97.5 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | I like the overall design of the lancing device | 99.2 percentage of participants |
| Persons With Diabetes | Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding the Styx Lancing Device | Easy / intuitive to do a fingerstick w this device | 97.5 percentage of participants |