Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis

Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02531646
Enrollment
66
Registered
2015-08-24
Start date
2015-11-30
Completion date
2016-04-30
Last updated
2017-03-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Solitary Pulmonary Nodules

Keywords

pulmonary nodule

Brief summary

The purpose of this study is to evaluate the imaging performance of Dual Energy and Digital Tomosynthesis (DE-DT) application against the commercially available predicate devices.

Detailed description

The acquisition system will consist of using site equipment including the CARESTREAM DRX-Evolution x-ray system, and the predicate CARESTREAM DRX Plus detector and the investigational DE-DT application. DE Study Design Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue). DT Study Design Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images. 11 phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.

Interventions

RADIATIONRadiation

Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure. Phantoms were imaged using linear tomography and digital tomography.

Sponsors

Carestream Health, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
DIAGNOSTIC
Masking
DOUBLE (Subject, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Age 18 years and older. Subject has provided informed consent. Study participant is scheduled to have a chest CT exam. Study participant is able to stand and be still during the exams.

Exclusion criteria

Human subject specific: Not able or willing to provide Informed Consent, or consent is withdrawn. Study participant is a female of child bearing age.

Design outcomes

Primary

MeasureTime frameDescription
Radlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Composite3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume3 month1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Radlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Participant flow

Participants by arm

ArmCount
Predicate & Invest. DE - Human Subjects
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue). Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
37
Predicate & Invest. DT - Human Subjects
Radiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images. Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
18
Predicate & Invest. DT - Phantom Images
Radiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational. Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
11
Total66

Withdrawals & dropouts

PeriodReasonFG000FG001FG002
Overall Studyequip. image processing issue510

Baseline characteristics

CharacteristicPredicate & Invest. DE - Human SubjectsPredicate & Invest. DT - Human SubjectsPredicate & Invest. DT - Phantom ImagesTotal
Age, Customized
> 18 years
37 years18 years11 years66 years
Sex/Gender, Customized
unknown
37 participants18 participants11 participants66 participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
— / —— / —— / —
other
Total, other adverse events
0 / 320 / 170 / 11
serious
Total, serious adverse events
0 / 320 / 170 / 11

Outcome results

Primary

Radlex Scale for Diagnostic Quality Ratings - DE Investigational Composite

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DE Investigational Composite3.3214 units on a scaleStandard Error 0.0242
p-value: 0Equivalence test
Primary

Radlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy3.2911 units on a scaleStandard Error 0.0178
p-value: 0Equivalence test
Primary

Radlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy3.5504 units on a scaleStandard Error 0.0163
p-value: 0t-test, 1 sided
Primary

Radlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

Population: 224 image ratings from 7 radiologist readers (32 images rated x 7 readers) for Predicate \& Invest. DE - Human Subjects arm above.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest3.4540 units on a scaleStandard Error 0.0183
Primary

Radlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

Population: 77 phantom image ratings from 7 radiologist readers (11 images rated x 7 readers) for Predicate \& Invest. DT - Phantom Images arm above.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images3.7156 units on a scaleStandard Error 0.0268
Primary

Radlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 month

Population: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for DT Investigational - Scout \& DT Volume.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume3.6277 units on a scaleStandard Error 0.023
Primary

Radlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

Population: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for Predicate \& Invest. DT Reference 1 - PA Chest arm above.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest3.4832 units on a scaleStandard Error 0.0228
p-value: 0t-test, 1 sided
Primary

Radlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

Population: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for Predicate \& Invest. DT Reference 2 - PA and LAT Chest arm above.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest3.5546 units on a scaleStandard Error 0.019
p-value: 0.007t-test, 1 sided
Primary

Radlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images

1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

Time frame: 3 months

Population: 77 phantom image ratings from 7 radiologist readers (11 images rated x 7 readers) for Predicate \& Invest. DT - Phantom Images arm above.

ArmMeasureValue (MEAN)Dispersion
Predicate & Invest. DE - Human SubjectsRadlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images3.3494 units on a scaleStandard Error 0.0433
p-value: 0t-test, 1 sided