Wearable Lower Extremity Exoskeleton to Promote Walking in Persons With Multiple Sclerosis

Wearable Lower Extremity Exoskeleton to Promote Walking in Persons With Multiple Sclerosis

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02519244
Enrollment
11
Registered
2015-08-10
Start date
2015-05-31
Completion date
2017-02-06
Last updated
2019-05-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Multiple Sclerosis

Keywords

Multiple Sclerosis, Ambulation, Rehabilitation, Wearable Robotic Devices

Brief summary

The purpose of this study is to investigate whether a wearable robotic exoskeleton can help people with multiple sclerosis (MS) to walk again.

Detailed description

Loss of the ability to walk is one of the significant problems in persons with multiple sclerosis (MS). This limitation prevents the person from having an active and meaningful life at home and in the community. Also, more energy is needed to walk, which can cause fatigue and further leads to a non-active life style and poor quality of life. Unfortunately, walking training is often not available for this population because of the severity and progress of the disease; therefore, a new strategy is needed to help people with MS to walk again. A new technology, the wearable robotic exoskeleton, has been developed to help persons with spinal cord injuries stand up and walk. The robotic legs are powered by batteries and controlled by motors. This study will investigate the use of a wearable robotic exoskeleton to help persons with MS to walk. Participants will receive trainings (5 days per week for 3 weeks) to walk with the robotic exoskeleton (EKSO, EKSO Bionics, Richmond, CA). During the training, a physical therapist will teach the participant how to maintain balance, sit to stand, stand to sit, and walk with the wearable robotic exoskeleton. The hypothesis is that persons with MS will be able to use a robotic exoskeleton to walk better and with less energy costs.

Interventions

DEVICEEkso® (Wearable lower limb exoskeleton)

The wearable lower limb exoskeleton is a powered, robotic lower limb exoskeleton with actuated hips and knees. A control algorithm has been implemented in this device, which allows for provision of assistance to lower limb segments during movement, dependent on user needs.

Sponsors

National Multiple Sclerosis Society
CollaboratorOTHER
Texas Woman's University
CollaboratorOTHER
The University of Texas Health Science Center, Houston
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

Subjects will be enrolled if they meet the following inclusion criteria: * Age 18 years or older * Ambulatory with assistive devices (Ambulation status will be determined by the EDSS score) * With an Expanded Disability Status Scale (EDSS) score between 6 and 7.5 inclusive * Height and weight are between 160 and 188 cm (5'2'' to 6'2''), and less than 100 kg (220 lb), respectively (per exoskeleton manufacturer) * Able to follow simple 3 step commends * Able to understand the study procedure and consent form Subjects will be excluded if they have any of the following

Exclusion criteria

* History of severe neurologic injuries other than MS (Spinal Cord Injury, Cerebral Palsy, Amyotrophic Lateral Sclerosis, Traumatic Brain Injury, Cerebral Vascular Accident, etc.) * Severe comorbidities: * active infections * heart, lung, or circulatory conditions * pressure ulcers * Documented severe osteoporosis affecting the hip and spine * Severe spasticity in the lower extremities (Modified Ashworth ≥ 3) or uncontrolled clonus * Unhealed limb or pelvic fractures * Skin issues that would prevent wearing the device * Range of motion restrictions that would prevent subject from achieving a normal, reciprocal gait pattern, or would restrict a subject from completing normal sit to stand or stand to sit transitions. * Upper extremity strength deficits that limit ability to balance with a front rolling walker or crutches. * Heterotopic ossification that resists functional range of motion in lower extremities * Contractures (\>15 degrees at the hips or \>20 degrees at the knees) * Psychiatric or cognitive comorbidities resulting in motor planning or impulsivity concerns * Colostomy * Have received any physical therapy intervention within 3 months prior to enrolment in the study * Non-English speaking

Design outcomes

Primary

MeasureTime frameDescription
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (With Exoskeleton)3 weeksPhysical demands during the Six-Minute Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)baselinePhysical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)3 weeksPhysical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)baselinePhysical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)3 weeksPhysical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (Without Exoskeleton)baselinePhysical demands during the Six-Minute Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.
Speed in the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)baselineThis task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at his/her comfortable pace without losing balance. A research team member will walk next to the subject for safety.
Speed in the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)3 weeksThis task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at his/her comfortable pace without losing balance. A research team member will walk next to the subject for safety.
Speed in the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)baselineThis task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at fast speed. A research team member will walk next to the subject for safety.
Speed in the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)3 weeksThis task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at fast speed. A research team member will walk next to the subject for safety.
Distance Walked During the Six-minute Walk Test (Without Exoskeleton)baselineSubjects will be asked to walk back and forth in a hallway with or without exoskeleton for 6 minutes. The objective is to cover as much space as possible in 6 minutes. Subjects can slow down or stop to rest if they feel like, but should start walking when they feel they are able. A research team member will walk behind the subject to prevent loss of balance during the test.
Distance Walked During the Six-minute Walk Test (With Exoskeleton)3 weeksSubjects will be asked to walk back and forth in a hallway with or without exoskeleton for 6 minutes. The objective is to cover as much space as possible in 6 minutes. Subjects can slow down or stop to rest if they feel like, but should start walking when they feel they are able. A research team member will walk behind the subject to prevent loss of balance during the test.

Secondary

MeasureTime frameDescription
Cognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (With Exoskeleton)3 weeksCognitive demands during the Timed 25 Feet Walk Test at self-selected pace with and without exoskeleton will be determined by reaction time using a dual-task paradigm. We chose a simple reaction time (RT) task, in which the response will be biting on a pressure sensor to make the response pathways as independent as possible from the motor pathways of locomotion. The secondary RT task consists of biting a pressure transducer placed in the mouth in response to an unpredictable sensory (will not cause pain) electrical stimulation applied by an electrode on the back of the neck without changing walking speed and pattern. The stimulation intensity will be adjusted for each individual before data collection. Shorter RT indicates that reduced amount of attentional resources are required.
Amount of Time Taken to Complete the Time Up and Go Test (Without Exoskeleton)baselineThis task will be performed with and without exoskeleton. This task involves subject to stand from the standard chair, walk straight for 3 meters, turn around, walk back to the chair and sit down with shoes and assistive devices if any. A standard chair with arm rests will be place at the start of the testing course. A mark will be placed on the floor at the 3 meter distance.
Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Physical Health Composite ScorebaselineMSQOL-54 physical health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life
Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Mental Health Composite ScorebaselineMSQOL-54 mental health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life
Cognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (Without Exoskeleton)3 weeksCognitive demands during the Timed 25 Feet Walk Test at self-selected pace with and without exoskeleton will be determined by reaction time using a dual-task paradigm. We chose a simple reaction time (RT) task, in which the response will be biting on a pressure sensor to make the response pathways as independent as possible from the motor pathways of locomotion. The secondary RT task consists of biting a pressure transducer placed in the mouth in response to an unpredictable sensory (will not cause pain) electrical stimulation applied by an electrode on the back of the neck without changing walking speed and pattern. The stimulation intensity will be adjusted for each individual before data collection. Shorter RT indicates that reduced amount of attentional resources are required.

Countries

United States

Participant flow

Pre-assignment details

11 were enrolled; however, one person did not initiate the intervention because they were scheduled for knee surgery.

Participants by arm

ArmCount
Robot-assisted Rehabilitation
Subjects will participate in individualized locomotion training sessions using wearable lower limb exoskeleton, Ekso®. Each training session will last up to 60 minutes, 5 days per week for 3 weeks, for a total of 15 sessions. During the training, subjects will wear a lower extremity exoskeleton robotic walking device. Subjects will participate in individualized treatment sessions which may include: sit to stand, static and dynamic standing balance, weight shifting, walking, turning, and stand to sit. Ekso® (Wearable lower limb exoskeleton): The wearable lower limb exoskeleton is a powered, robotic lower limb exoskeleton with actuated hips and knees. A control algorithm has been implemented in this device, which allows for provision of assistance to lower limb segments during movement, dependent on user needs.
10
Total10

Baseline characteristics

CharacteristicRobot-assisted Rehabilitation
Age, Continuous54.3 years
STANDARD_DEVIATION 12.4
Height1.68 meters
STANDARD_DEVIATION 0.09
Kurtzke Expanded Disability Status Scale (EDSS)6.55 units on a scale
STANDARD_DEVIATION 0.44
Multiple Sclerosis Type
Primary Progressive Multiple Sclerosis (PPMS)
3 Participants
Multiple Sclerosis Type
Relapsing-Remitting Multiple Sclerosis (RRMS)
5 Participants
Multiple Sclerosis Type
Secondary Progressive Multiple Sclerosis (SPMS)
2 Participants
Race/Ethnicity, Customized
African American
4 Participants
Race/Ethnicity, Customized
Asian
1 Participants
Race/Ethnicity, Customized
Caucasian
4 Participants
Race/Ethnicity, Customized
Hispanic
1 Participants
Region of Enrollment
United States
10 Participants
Sex: Female, Male
Female
8 Participants
Sex: Female, Male
Male
2 Participants
Weight70 kilograms
STANDARD_DEVIATION 12
Years Since Diagnosis15 years
STANDARD_DEVIATION 7.1

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 10
other
Total, other adverse events
0 / 10
serious
Total, serious adverse events
0 / 10

Outcome results

Primary

Distance Walked During the Six-minute Walk Test (With Exoskeleton)

Subjects will be asked to walk back and forth in a hallway with or without exoskeleton for 6 minutes. The objective is to cover as much space as possible in 6 minutes. Subjects can slow down or stop to rest if they feel like, but should start walking when they feel they are able. A research team member will walk behind the subject to prevent loss of balance during the test.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationDistance Walked During the Six-minute Walk Test (With Exoskeleton)108.09 metersStandard Deviation 26.26
Primary

Distance Walked During the Six-minute Walk Test (Without Exoskeleton)

Subjects will be asked to walk back and forth in a hallway with or without exoskeleton for 6 minutes. The objective is to cover as much space as possible in 6 minutes. Subjects can slow down or stop to rest if they feel like, but should start walking when they feel they are able. A research team member will walk behind the subject to prevent loss of balance during the test.

Time frame: 3 weeks

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationDistance Walked During the Six-minute Walk Test (Without Exoskeleton)138.90 metersStandard Deviation 66.89
Primary

Distance Walked During the Six-minute Walk Test (Without Exoskeleton)

Subjects will be asked to walk back and forth in a hallway with or without exoskeleton for 6 minutes. The objective is to cover as much space as possible in 6 minutes. Subjects can slow down or stop to rest if they feel like, but should start walking when they feel they are able. A research team member will walk behind the subject to prevent loss of balance during the test.

Time frame: baseline

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationDistance Walked During the Six-minute Walk Test (Without Exoskeleton)121.15 metersStandard Deviation 64.66
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (With Exoskeleton)

Physical demands during the Six-Minute Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (With Exoskeleton)7.77 ml/min/KgStandard Deviation 1.12
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (Without Exoskeleton)

Physical demands during the Six-Minute Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: baseline

Population: One subject did not perform the Timed 25 Feet Walk Test at Fast Speed due to fatigue. For another subject, there was a technical error during the assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (Without Exoskeleton)5.76 ml/min/KgStandard Deviation 1.25
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (Without Exoskeleton)

Physical demands during the Six-Minute Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: One subject did not perform the Timed 25 Feet Walk Test at Fast Speed due to fatigue. For another subject, there was a technical error during the assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Six-Minute Walk Test (Without Exoskeleton)10.20 ml/min/KgStandard Deviation 0.9
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)7.59 ml/min/KgStandard Deviation 1.32
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: baseline

Population: One subject did not perform the Timed 25 Feet Walk Test at Fast Speed due to fatigue. For another subject, data was not collected for the the Timed 25 Feet Walk Test at Fast Speed due to a technical error.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)3.6 ml/min/KgStandard Deviation 1.31
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: One subject did not perform the Timed 25 Feet Walk Test at Fast Speed due to fatigue. For another subject, there was a technical error during the assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)6.87 ml/min/KgStandard Deviation 1.44
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)6.33 ml/min/KgStandard Deviation 1.32
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: baseline

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)2.83 ml/min/KgStandard Deviation 1.12
Primary

Physical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)

Physical demands during the Timed 25 Feet Walk Test will be indicated by energy expenditure. Energy Expenditure (as indicated by VO2-max) will be measured by the K4 b2 Cosmed as follows: Oxygen cost will be calculated from oxygen consumption as the product of gait speed and body weight. Oxygen consumption will be collected on a breath-by-breath basis measured by a portable metabolic system (K4 b2 Cosmed). Prior to the testing, the system will be calibrated using room air and reference gas mixture. During the testing, the subject will wear a face mask and a heart rate monitor at all times and will be asked to breathe normally. VO2-max, also known as maximal oxygen uptake, is the measurement of the maximum amount of oxygen a person can utilize during exercise.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationPhysical Demands as Assessed by Energy Expenditure (Which is Indicated by VO2-max as Measured by the Cosmed K4b2) During the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)5.72 ml/min/KgStandard Deviation 1.06
Primary

Speed in the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at fast speed. A research team member will walk next to the subject for safety.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Fast Speed (With Exoskeleton)0.33 meters per second (m/s)Standard Deviation 0.11
Primary

Speed in the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at fast speed. A research team member will walk next to the subject for safety.

Time frame: baseline

Population: One subject did not perform the Timed 25 Feet Walk Test at fast speed because they were only able to walk at the self-selected speed.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)0.51 meters per second (m/s)Standard Deviation 0.24
Primary

Speed in the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at fast speed. A research team member will walk next to the subject for safety.

Time frame: 3 weeks

Population: One subject did not perform the Timed 25 Feet Walk Test at fast speed because they were only able to walk at the self-selected speed.For one of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Fast Speed (Without Exoskeleton)0.61 meters per second (m/s)Standard Deviation 0.32
Primary

Speed in the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at his/her comfortable pace without losing balance. A research team member will walk next to the subject for safety.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Self-selected Speed (With Exoskeleton)0.30 meters per second (m/s)Standard Deviation 0.09
Primary

Speed in the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at his/her comfortable pace without losing balance. A research team member will walk next to the subject for safety.

Time frame: baseline

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)0.35 meters per second (m/s)Standard Deviation 0.18
Primary

Speed in the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)

This task will ask the subject to walk for 25 feet at comfortable pace with and without exoskeleton. Subjects will wear their exercise or walking shoes and are allowed to use assistive device such as cane or walker if necessary. The start and finish line of the 25 feet test course will be marked with tape on the floor. Additional 5 feet at the end of start and finish will be used for subject to turn around. A chair will be provided next to the start area so the subject may rest. During the test, the subject will walk at his/her comfortable pace without losing balance. A research team member will walk next to the subject for safety.

Time frame: 3 weeks

Population: For two of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationSpeed in the Timed 25 Feet Walk Test at Self-selected Speed (Without Exoskeleton)0.46 meters per second (m/s)Standard Deviation 0.24
Secondary

Amount of Time Taken to Complete the Time Up and Go Test (Without Exoskeleton)

This task will be performed with and without exoskeleton. This task involves subject to stand from the standard chair, walk straight for 3 meters, turn around, walk back to the chair and sit down with shoes and assistive devices if any. A standard chair with arm rests will be place at the start of the testing course. A mark will be placed on the floor at the 3 meter distance.

Time frame: baseline

Population: One subject was not able to perform the Timed Up and Go test because they required support to sit and stand.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationAmount of Time Taken to Complete the Time Up and Go Test (Without Exoskeleton)33.38 secondsStandard Deviation 12.23
Secondary

Amount of Time Taken to Complete the Time Up and Go Test (Without Exoskeleton)

This task will be performed with and without exoskeleton. This task involves subject to stand from the standard chair, walk straight for 3 meters, turn around, walk back to the chair and sit down with shoes and assistive devices if any. A standard chair with arm rests will be place at the start of the testing course. A mark will be placed on the floor at the 3 meter distance.

Time frame: 3 weeks

Population: One subject was not able to perform the Timed Up and Go test because they required support to sit and stand.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationAmount of Time Taken to Complete the Time Up and Go Test (Without Exoskeleton)28.64 secondsStandard Deviation 13.74
Secondary

Cognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (With Exoskeleton)

Cognitive demands during the Timed 25 Feet Walk Test at self-selected pace with and without exoskeleton will be determined by reaction time using a dual-task paradigm. We chose a simple reaction time (RT) task, in which the response will be biting on a pressure sensor to make the response pathways as independent as possible from the motor pathways of locomotion. The secondary RT task consists of biting a pressure transducer placed in the mouth in response to an unpredictable sensory (will not cause pain) electrical stimulation applied by an electrode on the back of the neck without changing walking speed and pattern. The stimulation intensity will be adjusted for each individual before data collection. Shorter RT indicates that reduced amount of attentional resources are required.

Time frame: 3 weeks

Population: For three of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationCognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (With Exoskeleton)0.67 secondsStandard Deviation 0.43
Secondary

Cognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (Without Exoskeleton)

Cognitive demands during the Timed 25 Feet Walk Test at self-selected pace with and without exoskeleton will be determined by reaction time using a dual-task paradigm. We chose a simple reaction time (RT) task, in which the response will be biting on a pressure sensor to make the response pathways as independent as possible from the motor pathways of locomotion. The secondary RT task consists of biting a pressure transducer placed in the mouth in response to an unpredictable sensory (will not cause pain) electrical stimulation applied by an electrode on the back of the neck without changing walking speed and pattern. The stimulation intensity will be adjusted for each individual before data collection. Shorter RT indicates that reduced amount of attentional resources are required.

Time frame: 3 weeks

Population: For three of the subjects not reported, there were technical issues during assessment.

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationCognitive Demands as Indicated by Reaction Time in Dual Task Paradigm (Without Exoskeleton)0.51 secondsStandard Deviation 0.15
Secondary

Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Mental Health Composite Score

MSQOL-54 mental health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life

Time frame: 3 weeks

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationQuality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Mental Health Composite Score78.8 scoreStandard Deviation 13.2
Secondary

Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Mental Health Composite Score

MSQOL-54 mental health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life

Time frame: baseline

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationQuality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Mental Health Composite Score80.1 scoreStandard Deviation 10.7
Secondary

Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Physical Health Composite Score

MSQOL-54 physical health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life

Time frame: baseline

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationQuality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Physical Health Composite Score58.3 scoreStandard Deviation 12.3
Secondary

Quality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Physical Health Composite Score

MSQOL-54 physical health composite scores range from 0 to 100, with a higher scale score indicating improved quality of life

Time frame: 3 weeks

ArmMeasureValue (MEAN)Dispersion
Robot-assisted RehabilitationQuality of Life as Assessed by the Multiple Sclerosis Quality of Life-54 (MSQOL-54) Questionnaire - Physical Health Composite Score57.6 scoreStandard Deviation 13.6

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026