Skip to content

Accuracy Validation of a Pulse Oximetry Monitor

SpO2 Accuracy Validation of a USB Pulse Oximetry Monitor Interface Cable Via Reference Co-Oximetry

Status
Withdrawn
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02506010
Enrollment
0
Registered
2015-07-22
Start date
2016-11-30
Completion date
2017-01-31
Last updated
2016-10-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Keywords

Hypoxia

Brief summary

The purpose of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.

Detailed description

The description of this study is to conduct an oxygen saturation (SpO2) accuracy validation comparing prototype pulse oximeter oxygen saturation measurements directly to arterial blood saturation during non-motion conditions over the range of 60-100% SaO2 as measured by Reference CO-Oximetry. Additionally, the prototype pulse oximeter pulse rate measurements will be compared to 3-lead ECG heart rate measurements made via a multi-parameter monitor.

Interventions

None listed

Sponsors

Clinimark, LLC
CollaboratorOTHER
Medtronic - MITG
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to 45 Years
Healthy volunteers
Yes

Inclusion criteria

* healthy * non-smoking, or who have refrained from smoking for 2 days * male or female of any race * 18-45 years. * Subjects must have the ability to understand and provide written informed consent as well as be willing to comply with study procedures. * To be eligible to participate in the study, the subject must clear a recent physical (12-lead ECG, medical history and blood test for sickle cell disease.) prior to enrollment

Exclusion criteria

* \- morbid obesity (defined as BMI \> 39.5), * injuries, deformities or abnormalities or piercings that may prevent proper application of the device under test, compromised circulation, * subjects with known respiratory conditions (smokers with COHb levels \>3%, flu, pneumonia, bronchitis, shortness of breath/ respiratory distress, severe asthma, emphysema, COPD), * subjects with known heart or cardiovascular conditions (uncontrolled hypertension, heart arrhythmias other than sinus arrhythmia, previous heart attack, blocked artery, history of myocardial ischemia, angina, myocardial infarction, congestive heart failure or cardiomyopathy, history of stroke, transient ischemic attack or carotid artery disease), * other known health condition (diabetes, thyroid disease, kidney disease / chronic renal impairment, history of seizures (except childhood febrile seizures) or epilepsy, or unexplained syncope, recent history of frequent migraine headaches, recent head injury, cancer / chemotherapy * women actively trying to get pregnant or are pregnant, * clotting disorders (history of bleeding disorders or current use of blood thinners, Hemophilia, blood clots) * Sickle cell disease * Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors * Unwillingness or inability to remove colored nail polish or artificial nails from test digit(s) * Prior or known allergies to: lidocaine (or similar pharmacological agents, e.g., Novocain) or heparin * Recent arterial cannulation (i.e., less than 30 days prior to study date), Six or more arterial cannulations of each (right & left) radial artery, history of clinically significant complications from previous arterial cannulation

Design outcomes

Primary

MeasureTime frame
The OxiCable pulse rate measurement performance is an ARMS of 3 or better over the full range 60-100% SaO2.2.5 hours