Clinical Performance of Habitual Wearers of Bioclear 1-day Lenses When Refitted With Biomedics 1-day Extra for 1 Week

Clinical Performance of Habitual Wearers of Bioclear 1-day Lenses When Refitted With Biomedics 1-day Extra for 1 Week

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02406495
Enrollment
40
Registered
2015-04-02
Start date
2015-04-30
Completion date
2015-05-31
Last updated
2020-07-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Myopia, Hyperopia

Brief summary

Open label, 1-week daily disposable, dispensing study

Detailed description

Evaluate the clinical performance of habitual wearers of FILCON IV 1 sphere lenses when refitted with ocufilcon D asphere for 1 week of daily disposable wear.

Interventions

contact lens

DEVICEfilcon IV 1

contact lens

Sponsors

CooperVision, Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 40 Years
Healthy volunteers
No

Inclusion criteria

* Is between 18 and 40 years of age (inclusive) * Has had a self-reported visual exam in the last two years * Is an adapted Avaira sphere contact lens wearer (at least 1 week in Avaira sphere) * Has a contact lens spherical prescription between + 2.25 to - 8.00 (inclusive) * Has a spectacle cylinder up to 0.75D in each eye. * Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye. * Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses. * Has clear corneas and no active ocular disease * Has read, understood and signed the information consent letter. * Patient contact lens refraction should fit within the available parameters of the study lenses. * Is willing to comply with the wear schedule (at least 5 days per week, \> 8 hours/day assuming there are no contraindications for doing so). * Is willing to comply with the visit schedule

Exclusion criteria

* Is not a habitual wearer of Avaira sphere lenses * Has a CL prescription outside the range of the available parameters of the study lenses. * Has a spectacle cylinder ≥1.00D of cylinder in either eye. * Has a history of not achieving comfortable CL wear (5 days per week; \> 8 hours/day) * Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye. * Presence of clinically significant (grade 2-4) anterior segment abnormalities * Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear. * Slit lamp findings that would contraindicate contact lens wear such as: * Pathological dry eye or associated findings * Pterygium, pinguecula, or corneal scars within the visual axis * Neovascularization \> 0.75 mm in from of the limbus * Giant papillary conjunctivitis (GCP) worse than grade 1 * Anterior uveitis or iritis (past or present) * Seborrheic eczema, Seborrheic conjunctivitis * History of corneal ulcers or fungal infections * Poor personal hygiene * Has a known history of corneal hypoesthesia (reduced corneal sensitivity) * Has aphakia, keratoconus or a highly irregular cornea. * Has Presbyopia or has dependence on spectacles for near work over the contact lenses. * Has undergone corneal refractive surgery. * Is participating in any other type of eye related clinical or research study

Design outcomes

Primary

MeasureTime frameDescription
Lens Fit, Centration - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens fit, centration for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale: optimum, decentration acceptable, and decentration unacceptable
Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens fit, post-blink movement for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-4, 0=Insufficient, unacceptable movement, 1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).
Lens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens fit, lens tightness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 0%-100% continuous scale 0% - Falls from cornea without lid support 50% - Optimum 100% - No movement
Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens fit, overall fit acceptance for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 0-4, 0=Should not be worn, 4=Perfect.

Secondary

MeasureTime frameDescription
Vision Satisfaction (Subjective Ratings) - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekSubjective ratings of vision satisfaction for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=dissatisfied, 10=very satisfied).
Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon D1 WeekSubjective ratings of participant's lens preference for either filcon IV 1 or ocufilcon D on comfort, dryness, handling, vision, and overall. Forced choice: filcon IV 1 or ocufilcon D.
Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens satisfaction of comfort for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.
Visual Acuity - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekVisual acuity for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week using logMAR chart (a logMAR of 0.0=20/20 in Snellen notation and negative values indicate better visual acuity).
Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens satisfaction of handling forfilcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.
Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens satisfaction of vision for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.
Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens satisfaction overall for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.
Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekLens satisfaction of dryness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.
Comfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekSubjective ratings for comfort for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=could feel, 10=cannot feel).
Dryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekSubjective ratings of dryness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=dryness, 10=no dryness).
Handling (Subjective Ratings) - Filcon IV 1 and Ocufilcon DBaseline and 1 WeekSubjective ratings of handling for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=very difficult to handle, 10=very easy to handle).

Countries

Spain

Participant flow

Participants by arm

ArmCount
Overall Participants
Habitual wearers of (FILCON IV 1) sphere lenses refitted with asphere ocufilcon D lenses. filcon IV 1: contact lens ocufilcon D: contact lens
40
Total40

Baseline characteristics

CharacteristicOverall Participants
Age, Continuous23.92 years
STANDARD_DEVIATION 5.43
Sex: Female, Male
Female
31 Participants
Sex: Female, Male
Male
9 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 40
serious
Total, serious adverse events
0 / 40

Outcome results

Primary

Lens Fit, Centration - Filcon IV 1 and Ocufilcon D

Lens fit, centration for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale: optimum, decentration acceptable, and decentration unacceptable

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Fit, Centration - Filcon IV 1 and Ocufilcon DOptimum58 percentage of eyes
Filcon IV 1Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration acceptable43 percentage of eyes
Filcon IV 1Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration unacceptable0 percentage of eyes
Ocufilcon DLens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration acceptable30 percentage of eyes
Ocufilcon DLens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration unacceptable0 percentage of eyes
Ocufilcon DLens Fit, Centration - Filcon IV 1 and Ocufilcon DOptimum70 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DOptimum98 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration unacceptable0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration acceptable3 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration unacceptable0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DOptimum98 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Centration - Filcon IV 1 and Ocufilcon DDecentration acceptable3 percentage of eyes
Primary

Lens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon D

Lens fit, lens tightness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 0%-100% continuous scale 0% - Falls from cornea without lid support 50% - Optimum 100% - No movement

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (MEAN)Dispersion
Filcon IV 1Lens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon DOD49.1 percentage of mean lens tightnessStandard Deviation 5.9
Filcon IV 1Lens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon DOS48.9 percentage of mean lens tightnessStandard Deviation 5.2
Ocufilcon DLens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon DOD50.1 percentage of mean lens tightnessStandard Deviation 0.79
Ocufilcon DLens Fit, Lens Tightness - Filcon IV 1 and Ocufilcon DOS50.0 percentage of mean lens tightnessStandard Deviation 1.9
Primary

Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon D

Lens fit, overall fit acceptance for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 0-4, 0=Should not be worn, 4=Perfect.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DNot perfect but OK38 percentage of eyes
Filcon IV 1Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DPerfect63 percentage of eyes
Filcon IV 1Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DShould not be worn0 percentage of eyes
Filcon IV 1Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DBorderline but acceptable0 percentage of eyes
Filcon IV 1Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DMin. acceptable0 percentage of eyes
Ocufilcon DLens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DBorderline but acceptable0 percentage of eyes
Ocufilcon DLens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DMin. acceptable0 percentage of eyes
Ocufilcon DLens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DShould not be worn0 percentage of eyes
Ocufilcon DLens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DNot perfect but OK35 percentage of eyes
Ocufilcon DLens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DPerfect65 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DNot perfect but OK0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DShould not be worn0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DPerfect100 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DMin. acceptable0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DBorderline but acceptable0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DShould not be worn0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DNot perfect but OK0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DPerfect100 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DMin. acceptable0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Overall Fit Acceptance - Filcon IV 1 and Ocufilcon DBorderline but acceptable0 percentage of eyes
Primary

Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D

Lens fit, post-blink movement for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-4, 0=Insufficient, unacceptable movement, 1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D2=Optimal movement23 percentage of eyes
Filcon IV 1Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D0=Insufficient, unacceptable movement0 percentage of eyes
Filcon IV 1Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D3=Moderate, but acceptable movement65 percentage of eyes
Filcon IV 1Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D4=Excessive, unacceptable movement3 percentage of eyes
Filcon IV 1Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D1=Minimal, but acceptable movement10 percentage of eyes
Ocufilcon DLens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D4=Excessive, unacceptable movement3 percentage of eyes
Ocufilcon DLens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D2=Optimal movement18 percentage of eyes
Ocufilcon DLens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D3=Moderate, but acceptable movement70 percentage of eyes
Ocufilcon DLens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D1=Minimal, but acceptable movement10 percentage of eyes
Ocufilcon DLens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D0=Insufficient, unacceptable movement0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D4=Excessive, unacceptable movement0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D3=Moderate, but acceptable movement90 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D1=Minimal, but acceptable movement0 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D2=Optimal movement10 percentage of eyes
Ocufilcon D OD (Oculus Dexter)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D0=Insufficient, unacceptable movement0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D1=Minimal, but acceptable movement0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D3=Moderate, but acceptable movement90 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D0=Insufficient, unacceptable movement0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D4=Excessive, unacceptable movement0 percentage of eyes
Ocufilcon D OS (Oculus Sinister)Lens Fit, Post-blink Movement - Filcon IV 1 and Ocufilcon D2=Optimal movement10 percentage of eyes
Secondary

Comfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon D

Subjective ratings for comfort for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=could feel, 10=cannot feel).

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (MEAN)Dispersion
Filcon IV 1Comfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort overall7.3 units on a scaleStandard Deviation 1.3
Filcon IV 1Comfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort on insertion7.4 units on a scaleStandard Deviation 1.1
Filcon IV 1Comfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort prior to removal7.3 units on a scaleStandard Deviation 1.4
Ocufilcon DComfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort overall8.1 units on a scaleStandard Deviation 1.4
Ocufilcon DComfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort on insertion8.1 units on a scaleStandard Deviation 1.3
Ocufilcon DComfort (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort prior to removal8.1 units on a scaleStandard Deviation 1.3
Secondary

Dryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon D

Subjective ratings of dryness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=dryness, 10=no dryness).

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (MEAN)Dispersion
Filcon IV 1Dryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness during day7.2 units on a scaleStandard Deviation 1.3
Filcon IV 1Dryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness prior to removal7.3 units on a scaleStandard Deviation 1.3
Filcon IV 1Dryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DOverall dryness7.3 units on a scaleStandard Deviation 1.3
Ocufilcon DDryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness during day8.0 units on a scaleStandard Deviation 1.5
Ocufilcon DDryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness prior to removal8.0 units on a scaleStandard Deviation 1.5
Ocufilcon DDryness (Subjective Ratings) - Filcon IV 1 and Ocufilcon DOverall dryness8.1 units on a scaleStandard Deviation 1.5
Secondary

Handling (Subjective Ratings) - Filcon IV 1 and Ocufilcon D

Subjective ratings of handling for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=very difficult to handle, 10=very easy to handle).

Time frame: Baseline and 1 Week

ArmMeasureValue (MEAN)Dispersion
Filcon IV 1Handling (Subjective Ratings) - Filcon IV 1 and Ocufilcon D7.4 units on a scaleStandard Deviation 1.2
Ocufilcon DHandling (Subjective Ratings) - Filcon IV 1 and Ocufilcon D8.2 units on a scaleStandard Deviation 1.3
Secondary

Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon D

Subjective ratings of participant's lens preference for either filcon IV 1 or ocufilcon D on comfort, dryness, handling, vision, and overall. Forced choice: filcon IV 1 or ocufilcon D.

Time frame: 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort20 percentage of participants
Filcon IV 1Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DHandling20 percentage of participants
Filcon IV 1Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness20 percentage of participants
Filcon IV 1Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DVision20 percentage of participants
Filcon IV 1Lens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DOverall20 percentage of participants
Ocufilcon DLens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DVision80 percentage of participants
Ocufilcon DLens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DOverall80 percentage of participants
Ocufilcon DLens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DComfort80 percentage of participants
Ocufilcon DLens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DDryness80 percentage of participants
Ocufilcon DLens Preference (Subjective Ratings) - Filcon IV 1 and Ocufilcon DHandling80 percentage of participants
Secondary

Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon D

Lens satisfaction of comfort for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DCompletely Satisfied20 percentage of participants
Filcon IV 1Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DSomewhat Satisfied68 percentage of participants
Filcon IV 1Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied13 percentage of participants
Filcon IV 1Lens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DCompletely Satisfied60 percentage of participants
Ocufilcon DLens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied5 percentage of participants
Ocufilcon DLens Satisfaction, Comfort - Filcon IV 1 and Ocufilcon DSomewhat Satisfied35 percentage of participants
Secondary

Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon D

Lens satisfaction of dryness for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DCompletely Satisfied20 percentage of participants
Filcon IV 1Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DSomewhat Satisfied65 percentage of participants
Filcon IV 1Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied15 percentage of participants
Filcon IV 1Lens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DCompletely Satisfied60 percentage of participants
Ocufilcon DLens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied5 percentage of participants
Ocufilcon DLens Satisfaction, Dryness - Filcon IV 1 and Ocufilcon DSomewhat Satisfied35 percentage of participants
Secondary

Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon D

Lens satisfaction of handling forfilcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied13 percentage of participants
Filcon IV 1Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DSomewhat Satisfied65 percentage of participants
Filcon IV 1Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Filcon IV 1Lens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DCompletely Satisfied23 percentage of participants
Ocufilcon DLens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DCompletely Satisfied60 percentage of participants
Ocufilcon DLens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DSomewhat Satisfied38 percentage of participants
Ocufilcon DLens Satisfaction, Handling - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied3 percentage of participants
Secondary

Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon D

Lens satisfaction overall for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DCompletely Satisfied20 percentage of participants
Filcon IV 1Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied13 percentage of participants
Filcon IV 1Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DSomewhat Satisfied68 percentage of participants
Filcon IV 1Lens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DSomewhat Satisfied35 percentage of participants
Ocufilcon DLens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DCompletely Satisfied60 percentage of participants
Ocufilcon DLens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Overall - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied5 percentage of participants
Secondary

Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon D

Lens satisfaction of vision for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. Scale 1-4, 1=completely satisfied, 4=completely dissatisfied.

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (NUMBER)
Filcon IV 1Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DSomewhat Satisfied65 percentage of participants
Filcon IV 1Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DCompletely Satisfied23 percentage of participants
Filcon IV 1Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied13 percentage of participants
Filcon IV 1Lens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DCompletely Dissatisfied0 percentage of participants
Ocufilcon DLens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DSomewhat Satisfied38 percentage of participants
Ocufilcon DLens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DSomewhat Dissatisfied3 percentage of participants
Ocufilcon DLens Satisfaction, Vision - Filcon IV 1 and Ocufilcon DCompletely Satisfied60 percentage of participants
Secondary

Vision Satisfaction (Subjective Ratings) - Filcon IV 1 and Ocufilcon D

Subjective ratings of vision satisfaction for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week. (Scale 0-10, 0=dissatisfied, 10=very satisfied).

Time frame: Baseline and 1 Week

ArmMeasureValue (MEAN)Dispersion
Filcon IV 1Vision Satisfaction (Subjective Ratings) - Filcon IV 1 and Ocufilcon D7.5 units on a scaleStandard Deviation 1.3
Ocufilcon DVision Satisfaction (Subjective Ratings) - Filcon IV 1 and Ocufilcon D8.2 units on a scaleStandard Deviation 1.3
Secondary

Visual Acuity - Filcon IV 1 and Ocufilcon D

Visual acuity for filcon IV 1 assessed at baseline and ocufilcon D assessed at 1 week using logMAR chart (a logMAR of 0.0=20/20 in Snellen notation and negative values indicate better visual acuity).

Time frame: Baseline and 1 Week

ArmMeasureGroupValue (MEAN)Dispersion
Filcon IV 1Visual Acuity - Filcon IV 1 and Ocufilcon DOD0 LogMARStandard Deviation 0
Filcon IV 1Visual Acuity - Filcon IV 1 and Ocufilcon DOS0 LogMARStandard Deviation 0
Filcon IV 1Visual Acuity - Filcon IV 1 and Ocufilcon DBinocular-0.01 LogMARStandard Deviation 0.03
Ocufilcon DVisual Acuity - Filcon IV 1 and Ocufilcon DOD0 LogMARStandard Deviation 0
Ocufilcon DVisual Acuity - Filcon IV 1 and Ocufilcon DOS0 LogMARStandard Deviation 0
Ocufilcon DVisual Acuity - Filcon IV 1 and Ocufilcon DBinocular-0.01 LogMARStandard Deviation 0.04

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026