A Multi-Center Study of the CF Quantum® Sweat Test System

A Multi-Center Study of A New Method of Sweat Testing: The CF Quantum® Sweat Test

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02404714
Enrollment
55
Registered
2015-03-31
Start date
2017-02-08
Completion date
2018-07-01
Last updated
2018-08-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cystic Fibrosis

Brief summary

This is an observational comparative study to compare the standard-of-care methods for sweat testing used to support diagnosing cystic fibrosis (CF) against a new method of sweat testing called the CF Quantum Sweat Test System. The results of the new device in this study are for comparison ONLY and will NOT be used to diagnose, treat or mitigate the subject's condition.

Detailed description

The primary objective is to evaluate the diagnostic accuracy of a new sweat test method in both normal individuals and individuals with CF or cystic fibrosis related metabolic syndrome (CRMS). Subject participation is for one day and duration of study for each subject is one day. No follow-up will be required. Number of subjects; 300 split evenly between subjects with previously diagnosed CF or CRMS (cystic fibrosis related metabolic syndrome) and subjects referred to the sweat test lab for clinical reasons. Number of sites = 6, U of WI-Madison, U of WI-Milwaukee, U of MI-Ann Arbor, Dayton Children's Hospital (OH), U of MN, and Children's of Alabama. Primary endpoints: * Clinical and analytical sensitivity of the new test method * Clinical and analytical sensitivity of the standard-of-care methods Statistics Sensitivity and specificity will be calculated and reported along with the corresponding 95% confidence intervals. Bias assessment will be conducted according to the Clinical and Laboratory Standards Institute (CLSI) guidelines for method comparison and bias estimation. Pearson's correlation analysis will be conducted to test for adequate range of the replicate values for the new test method.

Interventions

DEVICEIontophoresis sweat test

The intervention is iontophoretic sweat testing which includes the stimulation, collection and analysis of sweat. The standard of care sweat testing used at each institution will be compared to a new iontophoresis sweat test called the CF Quantum.

Sponsors

Polychrome Medical, Inc.
Lead SponsorINDUSTRY

Study design

Observational model
CASE_ONLY
Time perspective
CROSS_SECTIONAL

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

1. Patients referred to the sweat test lab on a clinical basis. 2. Infants who require a sweat test as follow-up to an abnormal CF screening test. 3. Patients who have already been diagnosed with CF or CRMS. 4. Written informed consent (and assent when applicable) obtained from subject or subject's legal representative.

Exclusion criteria

1. Patient is receiving oxygen by open delivery. 2. Infants less than 48 hours of age. 3. Diffuse inflammation or rash on the collection site. 4. Patients who have had a reaction to a prior iontophoretic sweat test procedure. 5. Arm is too small for both the new and conventional sweat test.

Design outcomes

Primary

MeasureTime frame
Diagnostic sensitivity and specificity (true positive rate and the false negative rate to measure chloride in the sweat sample) of the new sweat test device compared to the standard-of-care sweat test deviceDay 1

Secondary

MeasureTime frame
Compare the sweat test's quantity not sufficient (QNS) rate of the new device to the standard-of-care device.Day 1

Countries

United States

Outcome results

None listed