Validation of Continuous Noninvasive Haemoglobin Monitoring by the Radical 7 Pulse CO-Oximetry Monitor After in Vivo Adjustment.

Validation of a Continuous Noninvasive Haemoglobin Determination Monitor (Radical 7 Pulse CO-Oximeter, SpHb) Compared to the Central Laboratory Coulter (LabHb) and a Point of Care (POC) Invasive Cooximeter (COoxHb) After in Vivo Adjustment.

Status
UNKNOWN
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02399488
Enrollment
60
Registered
2015-03-26
Start date
2012-02-29
Completion date
2015-09-30
Last updated
2015-08-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Sphb Haemoglobin in Vivo Validation

Brief summary

Continuous haemoglobin determination could improve the outcomes of several patients in risk of bleeding during surgery. An in vivo calibration has been introduced to the device after our team found its need. The uncalilbrated validation of the Masimo Radical 7 continuous haemoglobin monitor has been made in various papers, although it has been compared to point of care devices, thus introducing an error as such point of care devices although validated for haemoglobin determination, are not as accurate as the gold standard. The validation of the gold standard calibrated device has not been described yet. When compared to the accepted gold standard (cyanmethemoglobin method, Coulter), accuracy and precision of the continuous haemoglobin monitor could be good enough to be used interchangeably with the gold standard.

Detailed description

Continuous haemoglobin determination could improve the outcomes of several patients in risk of bleeding during surgery. The in vivo calibration has not been adjusted to the gold standard haemoglobin determination in the literature yet. Two blood samples will be withdrawn at the same time from an arterial line, one for the central laboratory and the other one for the POC device. At the same exact time the SpHb measurement will be recorded. At least eight samples are expected per patient. A maximum of 20 samples will be withdrawn per patient. To compare bias, precision and limits of agreement among SpHb, LabHb and COoxHb the Bland Altman method will be used with multiple observations per subject when the true value varies. An error grid analysis will be made to know how it could affect to clinical decision making.

Interventions

DEVICEMasimo Radical 7

Simultaneous determination of haemoglobin with the continuous haemoglobin monitor, the central lanboratory coulter and a POC:

Sponsors

Hospital Universitario Doctor Peset
Lead SponsorOTHER

Study design

Observational model
CASE_ONLY
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Adult patients. * Surgery duration of at least two hours. * Patients scheduled to surgery under general anaesthesia and muscle relaxation. * Radial arterial line needed for blood pressure control or other reason not related to the study.

Exclusion criteria

* Radial arterial line not indicated / not possible to place. * Sampling difficult or impossible. * Need to reposition arterial line arm during surgery. * Any illness that could influence the haemoglobin measurement.

Design outcomes

Primary

MeasureTime frameDescription
Central coulter bias and limits of agreement after in vivo calibration3 monthsBias and limits of agreement of SpHb compared to the central laboratory coulter after in vivo Radical 7 calibration
Central coulter bias and limits of agreement3 monthsBias and limits of agreement of SpHb compared to the central laboratory coulter

Secondary

MeasureTime frameDescription
POC Cooximeter bias and limits of agreement3 monthsBias and limits of agreement of SpHb compared to the POC cooximeter
POC Cooximeter Coulter interchangeability3 monthsBias and limits of agreement of POC Cooximeter Hb determination compared to the central laboratory coulter

Other

MeasureTime frameDescription
Comparison of bias and limits of agreement3 monthsComparison of bias and limits of agreement among central laboratory, POC cooximeter and SpHb continuous monitor.
Clinical decision making3 monthsError grid analysis plot

Countries

Spain

Contacts

Primary ContactJuan Soliveres, MD, PhD
+34670758871

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026