The purpose of this study was to determine if untrained subjects with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Interventions
Untrained Persons WITH Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested venous blood from subjects WITH Diabetes and BG results were compared to reference method results obtained from the subjects' venous plasma.
Sponsors
Study design
Eligibility
Inclusion criteria
* Males and females, 18 years of age and older * Able to speak, read, and understand English * Willing to complete all study procedures
Exclusion criteria
* Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a blood glucose monitoring study using the ONYX NEXT or ONYX PLUS BGMS * Working for a medical laboratory, hospital, or other clinical setting that involves training on or clinical use of blood glucose monitors * Working for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk General enrollment guidelines: * At least 70% of subjects will be younger than age 65 * At least 10% (approximately 10-15%) of subjects will be naive users * At least 20% of subjects with diabetes will have type 1 diabetes
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma). |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | 1 hour | Study staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma). |
| Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | 1 hour | Study staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI venous plasma) and +/- 15% (\>=100 mg/dL YSI venous plasma). |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. |
| Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff | 1 hour | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range. |
| Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma). |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 12.5% (\>=100 mg/dL YSI capillary plasma). |
| Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | 1 hour | Staff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. |
| Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | 1 hour | Staff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. |
| Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | 1 hour | Staff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'. |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma). |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma). |
| Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff | 1 hour | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Persons With and Without Diabetes Untrained Subjects WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System).
The ONYX NEXT BGMS: Untrained Persons WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. | 376 |
| Total | 376 |
Withdrawals & dropouts
| Period | Reason | FG000 |
|---|---|---|
| Overall Study | Lost to Follow-up | 1 |
Baseline characteristics
| Characteristic | Persons With and Without Diabetes |
|---|---|
| Age, Continuous | 52.1 years |
| Race/Ethnicity, Customized American Indian or Alaska Native | 7 Number of responses |
| Race/Ethnicity, Customized Asian | 23 Number of responses |
| Race/Ethnicity, Customized Black or African American | 52 Number of responses |
| Race/Ethnicity, Customized Ethnicity Unknown or Not Reported | 5 Number of responses |
| Race/Ethnicity, Customized Hispanic/Latino | 50 Number of responses |
| Race/Ethnicity, Customized Native Hawaiian or Other Pacific Islander | 2 Number of responses |
| Race/Ethnicity, Customized Not Hispanic or Latino | 321 Number of responses |
| Race/Ethnicity, Customized Race Unknown or Not Reported | 9 Number of responses |
| Race/Ethnicity, Customized White | 292 Number of responses |
| Region of Enrollment United States | 376 participants |
| Sex: Female, Male Female | 202 Participants |
| Sex: Female, Male Male | 174 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 8 / 375 |
| serious Total, serious adverse events | 0 / 375 |
Outcome results
Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 329 (332 with diabetes-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 99.4 Percent of Results within 15mg/dL/15% |
Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Untrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 318 (332 with diabetes -14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 97.2 Percent of Results within 15mg/dL/15% |
Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
Staff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I find it easy to do a blood test with this meter. | 98.1 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | The meter display was easy to see and read. | 98.4 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | It was easy to understand my test results. | 99.7 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I like the overall meter design. | 97.6 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I find the meter easy to use. | 98.1 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions (User Guide) were easy to understand. | 97.6 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions clearly explain how to run a test. | 98.7 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions clearly explain meter error messages. | 99.2 Percent of Subjects who responded |
Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS
Staff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I find it easy to do a blood test with this meter. | 97.9 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | The meter display was easy to see and read. | 98.2 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | It was easy to understand my test results. | 99.7 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I like the overall meter design. | 97.3 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | I find the meter easy to use. | 97.9 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions (User Guide) were easy to understand. | 97.6 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions clearly explain how to run a test. | 98.5 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Instructions clearly explain meter error messages. | 99.4 Percent of Subjects who responded |
Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose
Staff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'.
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to help me talk w my HCP. | 100 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to help w self-monitor satisfaction. | 99.70 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to help w managing my diabetes. | 99.40 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to help prevent low blood sugar | 99.69 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to understand food/exercise effects. | 99.70 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:help w using results to better control diabete | 99.40 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Acc:important to achieve greater peace of mind. | 99.09 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | I prefer the meter I just used to my regular meter | 92.09 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | I use my meter - my care provider gave it to me. | 85.27 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | I use my current meter -my insurance covers strips | 85.62 Percent of Subjects who responded |
| Persons With Diabetes | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | I use my current meter-I think it is most accurate | 84.09 Percent of Subjects who responded |
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method
Untrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 366 (375-9) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | 97.0 Percent of Results within 15mg/dL/15% |
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 96.7 Percent of Results within 15% |
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 98.6 Percent of Results within 20% |
Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method
Untrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 372 (375 - 3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | 99.7 Percent of Results within 15mg/dL/15% |
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method
Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 12.5% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 329 (332-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method | 98.5 Percent of Results within12.5mg/dL/12.5% |
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 372 (375-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 99.5 Percent of Results within 15% |
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 372 (375-3) Blood glucose results were analyzed. Three subjects did not contain meter BG results after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 99.7 Percent of Results within 20% |
Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff
Study staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 332 Blood glucose results from subjects with diabetes were analyzed.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | 99.7 Percent of Results within 15mg/dL/15% |
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff
Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 375 Blood glucose results were analyzed.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff | 99.7 Percent of Results within 15% |
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff
Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 375 Blood glucose results were analyzed.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff | 100 Percent of Results within 20% |
Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff
Study staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI venous plasma) and +/- 15% (\>=100 mg/dL YSI venous plasma).
Time frame: 1 hour
Population: 330 (332-2) Blood glucose results were analyzed. Two (2) subjects had unsuccessful venipuncture attempts, so no venous results were obtained for them.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | 100 Percent of Results within 15mg/dL/15% |