Evaluation of a Ascensia ONYX NEXT Investigational Blood Glucose Monitoring System

User Performance of the ONYX NEXT Blood Glucose Monitoring System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02390167
Enrollment
376
Registered
2015-03-17
Start date
2015-03-31
Completion date
2015-04-30
Last updated
2019-07-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study was to determine if untrained subjects with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Interventions

DEVICEONYX NEXT BGMS

Untrained Persons WITH Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested venous blood from subjects WITH Diabetes and BG results were compared to reference method results obtained from the subjects' venous plasma.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Males and females, 18 years of age and older * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a blood glucose monitoring study using the ONYX NEXT or ONYX PLUS BGMS * Working for a medical laboratory, hospital, or other clinical setting that involves training on or clinical use of blood glucose monitors * Working for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk General enrollment guidelines: * At least 70% of subjects will be younger than age 65 * At least 10% (approximately 10-15%) of subjects will be naive users * At least 20% of subjects with diabetes will have type 1 diabetes

Design outcomes

Primary

MeasureTime frameDescription
Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourUntrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).

Secondary

MeasureTime frameDescription
Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff1 hourStudy staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).
Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff1 hourStudy staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI venous plasma) and +/- 15% (\>=100 mg/dL YSI venous plasma).
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff1 hourStudy staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range.
Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourUntrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method1 hourUntrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 12.5% (\>=100 mg/dL YSI capillary plasma).
Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS1 hourStaff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.
Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS1 hourStaff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.
Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose1 hourStaff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'.
Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method1 hourUntrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).
Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method1 hourUntrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff1 hourStudy staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range.

Countries

United States

Participant flow

Participants by arm

ArmCount
Persons With and Without Diabetes
Untrained Subjects WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). The ONYX NEXT BGMS: Untrained Persons WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma.
376
Total376

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyLost to Follow-up1

Baseline characteristics

CharacteristicPersons With and Without Diabetes
Age, Continuous52.1 years
Race/Ethnicity, Customized
American Indian or Alaska Native
7 Number of responses
Race/Ethnicity, Customized
Asian
23 Number of responses
Race/Ethnicity, Customized
Black or African American
52 Number of responses
Race/Ethnicity, Customized
Ethnicity Unknown or Not Reported
5 Number of responses
Race/Ethnicity, Customized
Hispanic/Latino
50 Number of responses
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
2 Number of responses
Race/Ethnicity, Customized
Not Hispanic or Latino
321 Number of responses
Race/Ethnicity, Customized
Race Unknown or Not Reported
9 Number of responses
Race/Ethnicity, Customized
White
292 Number of responses
Region of Enrollment
United States
376 participants
Sex: Female, Male
Female
202 Participants
Sex: Female, Male
Male
174 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
8 / 375
serious
Total, serious adverse events
0 / 375

Outcome results

Primary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 329 (332 with diabetes-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method99.4 Percent of Results within 15mg/dL/15%
Secondary

Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Untrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 318 (332 with diabetes -14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method97.2 Percent of Results within 15mg/dL/15%
Secondary

Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS

Staff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find it easy to do a blood test with this meter.98.1 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSThe meter display was easy to see and read.98.4 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSIt was easy to understand my test results.99.7 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI like the overall meter design.97.6 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find the meter easy to use.98.1 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions (User Guide) were easy to understand.97.6 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain how to run a test.98.7 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain meter error messages.99.2 Percent of Subjects who responded
Secondary

Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS

Staff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'.

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find it easy to do a blood test with this meter.97.9 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSThe meter display was easy to see and read.98.2 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSIt was easy to understand my test results.99.7 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI like the overall meter design.97.3 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find the meter easy to use.97.9 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions (User Guide) were easy to understand.97.6 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain how to run a test.98.5 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain meter error messages.99.4 Percent of Subjects who responded
Secondary

Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose

Staff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'.

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help me talk w my HCP.100 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help w self-monitor satisfaction.99.70 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help w managing my diabetes.99.40 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help prevent low blood sugar99.69 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to understand food/exercise effects.99.70 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:help w using results to better control diabete99.40 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to achieve greater peace of mind.99.09 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI prefer the meter I just used to my regular meter92.09 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my meter - my care provider gave it to me.85.27 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my current meter -my insurance covers strips85.62 Percent of Subjects who responded
Persons With DiabetesPercent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my current meter-I think it is most accurate84.09 Percent of Subjects who responded
Secondary

Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method

Untrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 366 (375-9) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method97.0 Percent of Results within 15mg/dL/15%
Secondary

Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range96.7 Percent of Results within 15%
Secondary

Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range98.6 Percent of Results within 20%
Secondary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method

Untrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\< 75 mg/dL YSI capillary plasma) and +/- 15% (\>= 75 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 372 (375 - 3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method99.7 Percent of Results within 15mg/dL/15%
Secondary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method

Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 12.5% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 329 (332-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method98.5 Percent of Results within12.5mg/dL/12.5%
Secondary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 372 (375-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range99.5 Percent of Results within 15%
Secondary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 372 (375-3) Blood glucose results were analyzed. Three subjects did not contain meter BG results after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range99.7 Percent of Results within 20%
Secondary

Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff

Study staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/- 15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 332 Blood glucose results from subjects with diabetes were analyzed.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff99.7 Percent of Results within 15mg/dL/15%
Secondary

Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff

Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 375 Blood glucose results were analyzed.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff99.7 Percent of Results within 15%
Secondary

Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff

Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 375 Blood glucose results were analyzed.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff100 Percent of Results within 20%
Secondary

Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff

Study staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (\<100 mg/dL YSI venous plasma) and +/- 15% (\>=100 mg/dL YSI venous plasma).

Time frame: 1 hour

Population: 330 (332-2) Blood glucose results were analyzed. Two (2) subjects had unsuccessful venipuncture attempts, so no venous results were obtained for them.

ArmMeasureValue (NUMBER)
Persons With DiabetesPercent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff100 Percent of Results within 15mg/dL/15%