Skip to content

Hybrid L24 New Enrollment Post Approval Study

Nucleus Hybrid L24 Implant System: New Enrollment Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02379819
Acronym
HNE
Enrollment
52
Registered
2015-03-05
Start date
2015-02-28
Completion date
2022-04-14
Last updated
2023-03-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

High Frequency Sensorineural Hearing Loss

Brief summary

This study evaluates the long term safety and effectiveness of the Nucleus Hybrid L24 Implant in a group of newly implanted adults.

Interventions

Sponsors

Cochlear
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

Ear to Be Implanted: * Normal to moderate hearing loss in the low frequencies with severe to profound mid to high frequency hearing loss in the ear to be implanted * CNC word recognition of 10-60% in the ear to be implanted Contralateral Ear: * Moderately severe to profound mid to high frequency hearing loss in the contralateral ear * CNC word recognition equal to or better than that in the ear to be implanted but not more than 80% correct

Exclusion criteria

* not proficient in English * unwilling and/or unable to comply with the test protocol * deafness due to lesions of the acoustic nerve or central auditory pathway. * active middle-ear disease, with or without tympanic membrane perforation. * absence of cochlear development. * duration of severe to profound hearing loss of 30 years or greater

Design outcomes

Primary

MeasureTime frameDescription
Change in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceFrom pre-operatively through 3 years post-activation of the deviceThe CNC Words test consists of 10 recorded lists of 50 monosyllabic words in CD format. For this study, two lists will be administered in quiet at a level equal to 60 dBA in the sound field and recorded as a total number of words correct, which will be expressed as a percentage correct for this study. The CNC word test has a score range of 0-100% with higher values indicating better scores.
Number of Device or Procedure Related Adverse EventsUp to 5 years post-activation of the device

Secondary

MeasureTime frameDescription
Change in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionPre-operatively, up to 3 years post-activation of the deviceThe AzBio Sentence Tests consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. Each list includes 5 sentences from 4 different male and female speakers. Each word in the sentence counts toward the overall score and the resulting score is presented in percent correct. The AzBio sentence test has a score range of 0-100% with higher values indicating better scores.

Countries

United States

Participant flow

Participants by arm

ArmCount
Nucleus Hybrid L24 Implant
Adults age 18 and over who meet FDA criteria for unilateral implantation with the Nucleus Hybrid L24 Implant. Nucleus Hybrid L24 Implant
52
Total52

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyDeath2
Overall StudyLack of Efficacy1
Overall StudyLost to Follow-up1
Overall StudyWithdrawal by Subject10

Baseline characteristics

CharacteristicNucleus Hybrid L24 Implant
Age, Continuous59.9 years
STANDARD_DEVIATION 15.6
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
52 participants
Sex: Female, Male
Female
27 Participants
Sex: Female, Male
Male
25 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
2 / 52
other
Total, other adverse events
35 / 52
serious
Total, serious adverse events
1 / 52

Outcome results

Primary

Change in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the Device

The CNC Words test consists of 10 recorded lists of 50 monosyllabic words in CD format. For this study, two lists will be administered in quiet at a level equal to 60 dBA in the sound field and recorded as a total number of words correct, which will be expressed as a percentage correct for this study. The CNC word test has a score range of 0-100% with higher values indicating better scores.

Time frame: From pre-operatively through 3 years post-activation of the device

Population: Data was not collected for all 52 participants at all of the time points. At 2 years post-activation, one participant doesn't have a bilateral speech score.

ArmMeasureGroupValue (MEAN)
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceBilateral condition at 6 months post-activation29.7 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceBilateral condition at 1 year post-activation31.1 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceBilateral condition at 2 years post-activation32.4 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceUnilateral condition at 3 months post-activation31.3 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceUnilateral condition at 6 months post-activation35.2 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceUnilateral condition at 1 year post-activation35.8 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceUnilateral condition at 2 years post-activation35.3 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceUnilateral condition at 3 years post-activation36.8 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceBilateral condition at 3 months post-activation26.8 percentage of correct words
Nucleus Hybrid L24 ImplantChange in Consonant-Nucleus-Consonant (CNC) Word Recognition in Quiet for the Two Listening Conditions: Implant Ear Alone and When Using Both Ears Together From Pre-operatively Through 3 Years Post-activation of the DeviceBilateral condition at 3 years post-activation32.5 percentage of correct words
Primary

Number of Device or Procedure Related Adverse Events

Time frame: Up to 5 years post-activation of the device

Population: Related AEs were reported up to 3 years with some participants up to 5 years as the protocol was amended from 5 to 3 year follow-up in 2022.~The data include all related AEs reported.

ArmMeasureGroupValue (NUMBER)
Nucleus Hybrid L24 ImplantNumber of Device or Procedure Related Adverse EventsTotal number of device or procedure related adverse event98 Number of events
Nucleus Hybrid L24 ImplantNumber of Device or Procedure Related Adverse EventsNumber of serious device or procedure related adverse event1 Number of events
Secondary

Change in AzBio Sentence Recognition in Noise in the Best Unilateral Condition

The AzBio Sentence Tests consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. Each list includes 5 sentences from 4 different male and female speakers. Each word in the sentence counts toward the overall score and the resulting score is presented in percent correct. The AzBio sentence test has a score range of 0-100% with higher values indicating better scores.

Time frame: Pre-operatively, up to 3 years post-activation of the device

Population: Data was not collected for all 52 participants at all of the time points. At 1 year post-activation, one participant completed the testing at +10 dB SNR (decibel signal to noise ratio) instead of +5 dB SNR which is what is reported on.

ArmMeasureGroupValue (MEAN)
Nucleus Hybrid L24 ImplantChange in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionAt 3 months post-activation18 percentage of correct words
Nucleus Hybrid L24 ImplantChange in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionAt 6 months post-activation24.4 percentage of correct words
Nucleus Hybrid L24 ImplantChange in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionAt 1 year post-activation21.5 percentage of correct words
Nucleus Hybrid L24 ImplantChange in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionAt 2 years post-activation23.7 percentage of correct words
Nucleus Hybrid L24 ImplantChange in AzBio Sentence Recognition in Noise in the Best Unilateral ConditionAt 3 years post-activation17.8 percentage of correct words

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026