Evaluation of an Ascensia Diabetes Care ONYX PLUS Investigational Blood Glucose Monitoring System

User Performance of the ONYX PLUS Blood Glucose Monitoring System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02371876
Enrollment
134
Registered
2015-02-26
Start date
2015-02-28
Completion date
2015-03-31
Last updated
2019-10-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study was to determine if untrained subjects with diabetes can operate the Investigational ONYX PLUS Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Interventions

DEVICEONYX PLUS Investigational Blood Glucose Monitoring System

Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood. Study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Males and Females, 18 years of age and older * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Hemophilia or other bleeding disorder * Pregnancy * Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a BG monitor study using the ONYX PLUS BGMS * Working for a medical laboratory, hospital, or other clinical setting that involves training on or clinical use of blood glucose monitors * Working for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk The enrollment goal for the intended use population: 1. At least 70% of subjects will be younger than age 65 2. At least 20% of subjects will have type 1 diabetes

Design outcomes

Primary

MeasureTime frameDescription
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).

Secondary

MeasureTime frameDescription
Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method When Tested by Study Staff1 hourStudy staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
Percent of Venous Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method1 hourStudy staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI venous plasma) and +/- 15% (\>= 100mg/dL YSI venous plasma).
Percent of Alternate Site Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS1 hourStaff obtained responses from persons with diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree'. The percent of subjects who provided responses that were 'Strongly Agree','Agree','Neutral' about each statement was calculated.
Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose1 hourStaff obtained responses from persons with diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'No Answer or Not Applicable'. The percents (of subjects who responded) that were 'Strongly Agree','Agree','Neutral' about views/behaviors related to self-monitoring blood glucose were calculated.
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100mg/dL) of Laboratory Glucose Method1 hourUntrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5mg/dL (\<100mg/dL YSI capillary plasma) and +/- 12.5% (\>= 100mg/dL YSI capillary plasma).

Countries

United States

Participant flow

Participants by arm

ArmCount
Users of the Monitoring System
Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System. ONYX PLUS Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX PLUS Investigational Blood Glucose Monitoring System. All BG results were compared to reference method results obtained from subject capillary blood. Study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
134
Total134

Baseline characteristics

CharacteristicUsers of the Monitoring System
Age, Continuous54.0 years
Ethnicity (NIH/OMB)
Hispanic or Latino
14 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
119 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
3 participants
Race/Ethnicity, Customized
Asian
4 participants
Race/Ethnicity, Customized
Black or African American
14 participants
Race/Ethnicity, Customized
More than one race
3 participants
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
2 participants
Race/Ethnicity, Customized
Unknown or Not Reported
2 participants
Race/Ethnicity, Customized
White
112 participants
Region of Enrollment
United States
134 participants
Sex: Female, Male
Female
70 Participants
Sex: Female, Male
Male
64 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
4 / 134
serious
Total, serious adverse events
0 / 134

Outcome results

Primary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).

Time frame: 1 hour

ArmMeasureValue (NUMBER)
Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method99.2 Percentage of BG Results
Secondary

Percent of Alternate Site Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 126 (134-8) Blood glucose results were analyzed. Four subjects with low blood sugar did not attempt palm testing per protocol. Four subjects had low blood sugar; their palm results were not evaluable per protocol.

ArmMeasureValue (NUMBER)
Users of the Monitoring SystemPercent of Alternate Site Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method99.2 Percentage of BG Results
Secondary

Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS

Staff obtained responses from persons with diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree'. The percent of subjects who provided responses that were 'Strongly Agree','Agree','Neutral' about each statement was calculated.

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find it easy to do a blood test with this meter.99.2 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSThe meter display was easy to see and read.98.5 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSIt was easy to understand my test results.100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI like the overall meter design.97.8 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSI find the meter easy to use.100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSThe instructions were easy to understand.96.3 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain meter error messages.99.2 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMSInstructions clearly explain how to run a test.97.8 Percentage of Participants Who Responded
Secondary

Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose

Staff obtained responses from persons with diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'No Answer or Not Applicable'. The percents (of subjects who responded) that were 'Strongly Agree','Agree','Neutral' about views/behaviors related to self-monitoring blood glucose were calculated.

Time frame: 1 hour

Population: Percent (of those who responded) who 'Strongly Agree', 'Agree', 'Neutral' was calculated for each statement. Abbreviations used: 'HCP' is HealthCare Professional and 'Acc' is Accuracy

ArmMeasureGroupValue (NUMBER)
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help me talk w my HCP.99.25 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help w self-monitor satisfaction.99.25 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help w managing my diabetes.100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to help prevent low blood sugar100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to understand food/exercise effects.100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:help w using results to better control diabete100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseAcc:important to achieve greater peace of mind.100 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI prefer the meter I just used to my regular meter86.5 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my current meter -my insurance covers strips80 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my current meter-I think it is most accurate86.9 Percentage of Participants Who Responded
Users of the Monitoring SystemPercent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood GlucoseI use my meter - my care provider gave it to me.84.1 Percentage of Participants Who Responded
Secondary

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100mg/dL) of Laboratory Glucose Method

Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5mg/dL (\<100mg/dL YSI capillary plasma) and +/- 12.5% (\>= 100mg/dL YSI capillary plasma).

Time frame: 1 hour

ArmMeasureValue (NUMBER)
Users of the Monitoring SystemPercent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100mg/dL) of Laboratory Glucose Method98.5 Percentage of BG Results
Secondary

Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method When Tested by Study Staff

Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).

Time frame: 1 hour

ArmMeasureValue (NUMBER)
Users of the Monitoring SystemPercent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method When Tested by Study Staff99.2 Percentage of BG Results
Secondary

Percent of Venous Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method

Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI venous plasma) and +/- 15% (\>= 100mg/dL YSI venous plasma).

Time frame: 1 hour

Population: 132 (134-2) Blood glucose results were analyzed. Venipunctures were unsuccessful for two subjects.

ArmMeasureValue (NUMBER)
Users of the Monitoring SystemPercent of Venous Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method100 Percentage of BG Results