User Performance Evaluation for DARIO™ BGMS

User Performance Evaluation for DARIO™ Blood Glucose Monitoring System (BGMS)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02357043
Enrollment
368
Registered
2015-02-06
Start date
2014-12-31
Completion date
2015-03-31
Last updated
2021-06-15

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The study aims to evaluate the accuracy of the blood glucose level results obtained from fingertip using the Dario Blood Glucose Monitoring System (BGMS) compared to Yellow Springs Instrument (YSI) as well as to evaluate the usability of the Dario Blood Glucose Monitoring System (BGMS) by laypersons using only device labeling.

Detailed description

Each subject enrolled in the study will be requested to follow the device instructions and perform his/her own finger-stick test using the Dario Blood Glucose Monitoring System (BGMS). Immediately after the subject tested him/herself a trained nurse/technician will collect another blood sample from the subject for testing by the Yellow Springs Instrument (YSI). The subject will complete a usability questionnaire on his/her experience using the Dario Blood Glucose Monitoring System (BGMS) and the nurse/technician will complete a questionnaire assessing the subject's performance. A venous blood sample for Complete Blood Count (CBC) will be taken by the nurse/technician and sent for analysis.

Interventions

Finger-stick obtained with BGMS

DEVICEYSI

Finger-stick sample obtained by nurse/technician to be tested with YSI

Sponsors

LabStyle Innovations Ltd.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

1. Subject has Type I or Type II diabetes 2. Subject speaks and reads English proficiently 3. Subjects owns and uses Android and iOS based smart mobile device for at least 1 year 4. Subject is able and agrees to sign the informed consent form (minor subjects must be accompanied by a parent or legal guardian to provide consent for their participation in the study).

Exclusion criteria

1. Subject has medical training or works in the field of Blood Glucose Monitoring System (BGMS) 2. Subject is critically ill 3. Subject has an impairment that prevents him/her to follow the study procedures 4. Subject has any condition that the Principal Investigator believes may interfere in the subject's participation in the study

Design outcomes

Primary

MeasureTime frameDescription
Accuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician5 minutesBlood Glucose Level obtained by layperson on Dario meter versus Blood Glucose Level by Yellow Springs Instrument (obtained by nurse/technician)

Countries

United States

Participant flow

Participants by arm

ArmCount
Dario Blood Glucose Monitoring System
Subject will be requested to follow the device instructions and perform his/her own finger-stick test using the Dario Glucose Monitoring System (BGMS). Dario Blood Glucose Monitoring System: Finger-stick obtained with BGMS YSI: Finger-stick sample obtained by nurse/technician to be tested with YSI
368
Total368

Baseline characteristics

CharacteristicDario Blood Glucose Monitoring System
Age, Categorical
<=18 years
3 Participants
Age, Categorical
>=65 years
65 Participants
Age, Categorical
Between 18 and 65 years
300 Participants
Age, Continuous52.4 years
Race/Ethnicity, Customized
Americam Indian or Alaska Natice
3 participants
Race/Ethnicity, Customized
Asian
7 participants
Race/Ethnicity, Customized
Black or African America
81 participants
Race/Ethnicity, Customized
Hispanic or Latino
1 participants
Race/Ethnicity, Customized
Native Hawaiian or other Pacific Ilander
1 participants
Race/Ethnicity, Customized
Other
8 participants
Race/Ethnicity, Customized
White
267 participants
Region of Enrollment
United States
368 participants
Sex: Female, Male
Female
169 Participants
Sex: Female, Male
Male
199 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 368
serious
Total, serious adverse events
0 / 368

Outcome results

Primary

Accuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician

Blood Glucose Level obtained by layperson on Dario meter versus Blood Glucose Level by Yellow Springs Instrument (obtained by nurse/technician)

Time frame: 5 minutes

ArmMeasureGroupValue (NUMBER)
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 75 mg/dl Within ±5mg/dL12 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>=75 mg/dl Within ±15%328 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 75 mg/dl Within ±10mg/dL22 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 75 mg/dl Within ±15mg/dL24 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>=75 mg/dl Within ±5%134 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>=75 mg/dl Within ±10%257 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>=75 mg/dl Within ±20%339 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 100 mg/dl Within ±5mg/dL42 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 100 mg/dl Within ±10mg/dL73 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician< 100 mg/dl Within ±15mg/dL91 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>= 100 mg/dl Within ±5%111 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>= 100 mg/dl Within ±10%211 Participants
Dario Blood Glucose Monitoring SystemAccuracy - Blood Glucose Level Obtained by Layperson With the Dario Meter Compared to Blood Glucose Level on Yellow Springs Instrument Obtained by Nurse/Technician>= 100 mg/dl Within ±15%265 Participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026