Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Pregnant Women

Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Pregnant Women

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02319174
Enrollment
43
Registered
2014-12-18
Start date
2015-06-30
Completion date
2016-11-04
Last updated
2018-12-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Pre Eclampsia

Brief summary

The purpose of the study is to help make a lower cost automatic blood pressure monitor device for diagnosis and monitoring of pre-eclampsia in pregnant women, where automatic blood pressure monitoring is limited or not available. The study will compare this low cost device to a commercially available system used for pre-eclamptic women in many United States hospitals that the investigators will be bringing to Malawi as a part of this study. The team hopes to show that this lower cost blood pressure machine works well and can help women with pre-eclampsia. The study also aims to see if this machine is easy for the nurse to use. 70 pregnant women who are either at-risk or diagnosed with pre-eclampsia will be enrolled at University of Texas Health Science Center Houston. Patient arm circumference will be measured with measurement tape. They will be seated upright in a comfortable chair with arm at heart level and an arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm. The cuff will be inflated and then deflated until measurement concludes. Heart rate will be measured with tactile arterial palpation.The process will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement. The results of this study will help researchers understand the performance and usability of this device in Malawi and help decide if any design changes are needed.

Detailed description

The research team has developed an automatic blood pressure monitor to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited. This study aims to evaluate the accuracy of the device's blood pressure measurements in pregnant and pre-eclamptic women. This device must be evaluated with the above described population because blood pressure measurement devices are known to perform differently in pregnant and pre-eclamptic women than in normal healthy adults. This study will take place at the University of Texas Health Science Center (UTHSC) where up to 70 subjects will be recruited to participate during their regularly scheduled antenatal care checkups. 1. The Patient arm circumference will be measured with measurement tape. 2. The Patient will be seated upright in a comfortable chair with arm at heart level. 3. An arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm of the patient over the brachial artery. 4. Cuff will be inflated to \ 200 mmHg and then deflated until measurement concludes. 5. Heart rate will be measured with Tactile Arterial Palpation. 6. Steps 3-4 will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement.

Interventions

DEVICESphygmo

A team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals

Sponsors

The University of Texas Health Science Center, Houston
CollaboratorOTHER
William Marsh Rice University
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Eligibility

Sex/Gender
FEMALE
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Pregnant women receiving antenatal care at the UT OB/GYN clinic. * Over the age of 18

Exclusion criteria

* Women under the age of 18 * Women unable to provide informed consent

Design outcomes

Primary

MeasureTime frameDescription
Accuracy of Blood Pressure Readings by SphygmoBlood pressure was measured an average of 9 times for each participant and the average of these measurements was recorded. Each measurement period lasted approximately 30-45 minutesMean difference of systolic and diastolic blood pressures between Sphygmo blood pressure measurements and measurements from the gold standard sphygmomanometer.
Percent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Blood pressure was measured an average of 9 times for each participant during their single measurement period, and the average of those measurements was recorded. Each measurement period lasted approximately 30-45 minutesThe British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of \< 5 mmHg, \< 10 mmHg, and \< 15 mmHg. A letter grade of A requires that over 60%, 85%, and 95% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. A letter grade of B requires that over 50%, 75%, and 90% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. The test device achieved a grade of (A/A) with the validation data from this study.

Countries

United States

Participant flow

Participants by arm

ArmCount
Sphygmo
All subjects will have blood pressure measured by both the research device (Sphygmo) and the commercially available device. All clinical decisions will be made using measurements from the commercially available device. All subjects are pregnant women. Sphygmo: A team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals
42
Total42

Baseline characteristics

CharacteristicSphygmo
Age, Continuous29 years
Region of Enrollment
United States
42 participants
Sex: Female, Male
Female
42 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 43
other
Total, other adverse events
0 / 43
serious
Total, serious adverse events
0 / 43

Outcome results

Primary

Accuracy of Blood Pressure Readings by Sphygmo

Mean difference of systolic and diastolic blood pressures between Sphygmo blood pressure measurements and measurements from the gold standard sphygmomanometer.

Time frame: Blood pressure was measured an average of 9 times for each participant and the average of these measurements was recorded. Each measurement period lasted approximately 30-45 minutes

Population: Of 41 participants who completed the study, 5 participants were excluded on account of the observer reporting a difficulty in hearing, movement, or uncertainty with the measurement. This left 36 participants, 18 of which were used to train the device. The data presented here is from the remaining 18 participants.

ArmMeasureGroupValue (MEAN)Dispersion
SphygmoAccuracy of Blood Pressure Readings by SphygmoSystolic Mean114.5 mmHgStandard Deviation 10.9
SphygmoAccuracy of Blood Pressure Readings by SphygmoDiastolic Mean65.7 mmHgStandard Deviation 8.3
DinamapAccuracy of Blood Pressure Readings by SphygmoSystolic Mean115.6 mmHgStandard Deviation 12.5
DinamapAccuracy of Blood Pressure Readings by SphygmoDiastolic Mean66.3 mmHgStandard Deviation 10.6
Primary

Percent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.

The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of \< 5 mmHg, \< 10 mmHg, and \< 15 mmHg. A letter grade of A requires that over 60%, 85%, and 95% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. A letter grade of B requires that over 50%, 75%, and 90% were achieved in the \< 5 mmHg, \< 10 mmHg, and \<15 mmHg categories, respectively. The test device achieved a grade of (A/A) with the validation data from this study.

Time frame: Blood pressure was measured an average of 9 times for each participant during their single measurement period, and the average of those measurements was recorded. Each measurement period lasted approximately 30-45 minutes

Population: Of 41 participants who completed the study, 5 participants were excluded on account of the observer reporting a difficulty in hearing, movement, or uncertainty with the measurement. This left 36 participants, 18 of which were used to train the device. The data presented here is from the remaining 18 participants.

ArmMeasureGroupValue (NUMBER)
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Systolic Difference <5 mmHg67 percentage of measurements in range
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Systolic Difference <10 mmHg85 percentage of measurements in range
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Systolic Difference <15 mmHg98 percentage of measurements in range
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Diastolic Difference <5 mmHg70 percentage of measurements in range
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Diastolic Difference <10 mmHg89 percentage of measurements in range
SphygmoPercent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.Diastolic Difference <15 mmHg98 percentage of measurements in range