Lesion Detection of Automated Breast Ultrasound Compared With Handheld Physician-performed Breast Ultrasound

Breast Lesion Detection Using the Acuson S2000 Automated Breast Volume Scanner (ABVS) Compared With Handheld (HH) Physician-performed Ultrasound (US) Examinations

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02310776
Enrollment
501
Registered
2014-12-08
Start date
2012-06-30
Completion date
2015-10-31
Last updated
2016-02-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Breast Diseases

Brief summary

Compare lesion detection of the automated breast ultrasound (ABVS) exam with the conventional handheld (HH) breast ultrasound (US) exam.

Detailed description

This is a research study to test a FDA-approved automated breast ultrasound (ABVS) machine in subjects recommended for an ultrasound of the breast. The main purpose of the study is to compare images obtained from the automated breast ultrasound exam with the standard physician performed handheld breast ultrasound exam for lesion detection equivalence. Each subject will have a bilateral hand-held breast ultrasound and a bilateral automated breast ultrasound, both exams preferably done on the same day. The hand-held ultrasound will be performed by a breast imaging physician participating in the study and the automated ultrasound exam will be performed by a sonographer. There is no radiation risk from ultrasound. The physician performing the handheld breast ultrasound is blinded to the study, will not know the subject's history, but only the reason for the breast ultrasound exam. The automated scan requires gentle pressure on the breast during a scan. Three basic views will be acquired for each breast, each scan lasting approximately one minute. Once all necessary views are acquired, the images will be sent to a remote workstation, where a second radiologist, also blinded to the results of the handheld study, will interpret the automated ultrasound. Once all images and patient information are gathered, the study radiologists will review the previous images, if available, to provide the final result that will determine clinical management. Therefore, it is possible that the hand-held or automated ultrasound scan will lead to further evaluation of the breast and possibly a biopsy.

Interventions

DEVICESupine automated breast ultrasound scanner

Automated wide-field-of-view breast ultrasound volume scan performed by a sonographer and interpreted by a breast imaging radiologist.

DEVICEHigh-resolution handheld breast ultrasound

Standard of care, small field-of-view 2D breast ultrasound performed and interpreted by a breast imaging radiologist

Sponsors

Siemens Medical Solutions
CollaboratorINDUSTRY
Northwestern University
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 90 Years
Healthy volunteers
Yes

Inclusion criteria

* anyone who is referred for an ultrasound of the breast

Exclusion criteria

* Subjects whose breast thickness is greater than 6 cm.The maximum depth of image filed for the ABVS acquisition is 6 cm

Design outcomes

Primary

MeasureTime frameDescription
Ratio of Lesions Detected With Automated Breast Ultrasound (ABVS) Compared to the Standard of Care Handheld (HH) Breast Ultrasound (US)Time for performance of ABVS was measured separately from interpretation time. Time for performance of real time HH whole breast US was also measured. For HH US physician performed, performance included interpretation.Solid lesions found on whole breast 3D supine automated breast ultrasound are compared in number and identity with solid lesions found using high resolution standard of care handheld breast ultrasound performed by physicians

Countries

United States

Participant flow

Participants by arm

ArmCount
Automated & Handheld Beast US Exams
1. Automated breast ultrasound exam: 3D supine automated breast ultrasound scanner; Supine automated breast ultrasound scanner: Automated wide-field-of-view breast ultrasound volume scan performed by a sonographer and interpreted by a breast imaging radiologist. 2. Handheld breast ultrasound exam: High-resolution handheld breast ultrasound; Standard of care, small field-of-view 2D breast ultrasound performed and interpreted by a breast imaging radiologist
501
Total501

Baseline characteristics

CharacteristicAutomated & Handheld Beast US Exams
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
40 Participants
Age, Categorical
Between 18 and 65 years
461 Participants
Region of Enrollment
United States
501 participants
Sex: Female, Male
Female
501 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 501
serious
Total, serious adverse events
0 / 501

Outcome results

Primary

Ratio of Lesions Detected With Automated Breast Ultrasound (ABVS) Compared to the Standard of Care Handheld (HH) Breast Ultrasound (US)

Solid lesions found on whole breast 3D supine automated breast ultrasound are compared in number and identity with solid lesions found using high resolution standard of care handheld breast ultrasound performed by physicians

Time frame: Time for performance of ABVS was measured separately from interpretation time. Time for performance of real time HH whole breast US was also measured. For HH US physician performed, performance included interpretation.

Population: Each participant was evaluated based on Breast Imaging-Reporting and Data System (BI-RADS) assessment for Automated and Handheld breast ultrasound exams. If participant had more than one lesion, each lesion was assessed and listed separately and if a participant did not have a lesion, she was entered once.

ArmMeasureValue (NUMBER)
Automated & Handheld Breast US ExamsRatio of Lesions Detected With Automated Breast Ultrasound (ABVS) Compared to the Standard of Care Handheld (HH) Breast Ultrasound (US)1.19 Detection ratio of lesions