Effectiveness and Safety Study of the Abbott Sensor Based Glucose Monitoring Systems

Effectiveness and Safety Study of the Abbott Sensor Based Glucose Monitoring System - Personal and Abbott Sensor Based Glucose Monitoring System - Professional

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02283411
Enrollment
156
Registered
2014-11-05
Start date
2014-11-30
Completion date
2015-03-31
Last updated
2018-02-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2

Brief summary

The purpose of the study is to evaluate the safety and effectiveness of the Abbott Sensor Based Glucose Monitoring System Personal (System-P) and Abbott Sensor Based Glucose Monitoring System Professional (System-Pro) when used as an adjunct to blood glucose testing over a 14-day wear period in adult subjects. The Abbott Sensor Based Glucose Monitoring System - Professional Sensors will be worn by adult subjects over a 14-day wear period. The primary objective is to characterize the Systems performance with respect to Yellow Spring Instrument (YSI) reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the Abbott Sensor Based Glucose Monitoring Systems in reference to YSI. Safety of the Abbott Sensor Based Glucose Monitoring Systems will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants. 240 subjects will be enrolled and additional healthy subjects may also be enrolled in the study as training subjects.

Detailed description

Sensors are inserted into the back of the subject's upper arm for up to fourteen (14) days. The subjects are expected to perform capillary BG tests during the 14-day sensor wear. Three in clinic visits are scheduled during the 14-day sensor wear period. During each in-clinic visit, study staff performed IV blood draws to obtain venous blood for YSI reference glucose measurements.

Interventions

Subjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes.

Sponsors

Abbott Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Subject must be at least 18 years of age. 2. Subject must have a diagnosis of type 1 or type 2 diabetes mellitus. 3. Use one of the following for their diabetes management: * intensive insulin therapy (insulin therapy through an insulin pump and/or multiple daily insulin injections) or * non-intensive insulin therapy (Basal insulin, injectable non-insulin anti-diabetic agents such as a pramlintide or GLP-1 agonists, NPH insulin either alone or in combination with regular insulin or fast acting insulin analogue, or oral medication) 4. Intensive-insulin using (IIT) subjects only - During one of the in-clinic sessions, subject is willing to have their blood glucose levels manipulated into high and low glucose levels. Insulin-to-carbohydrate ratio and insulin sensitivity formula must be documented prior to having blood glucose levels manipulated. NIT subjects will only be observed during each in-clinic session. 5. For subjects that exercise routinely (at least 3 times per week), willing to exercise during at least one of the in-clinic sessions, if asked. 6. Willing to perform a minimum of 8 finger sticks per day during the study. 7. Subject must be able to read and understand English. 8. In the investigator's opinion, the subject must be able to follow the instructions provided to him/her by the study site and perform all study tasks as specified by the protocol. 9. Subject must be available to participate in all study visits. 10. Subject must be willing and able to provide written signed and dated informed consent.

Exclusion criteria

1. Subject has known allergy to medical grade adhesive or isopropyl alcohol used to disinfect skin. 2. Subject is pregnant, is attempting to conceive or is not willing and able to practice birth control during the study duration (applicable to female subjects only). 3. Subject has extensive skin changes/diseases at the proposed application sites that could interfere with device placement or the accuracy of interstitial glucose measurements. Such conditions include, but are not limited to extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, dermatitis herpetiformis, skin lesions, redness, infection or edema. 4. Subject is on dialysis. 5. Subject has a history of untreated hypothyroidism, adrenal gland failure or insufficiency, celiac disease, renal failure, cystic fibrosis, unstable coronary heart disease, or serious psychiatric disorders. 6. Subjects with a history of hypoglycemia unawareness, hypoglycemic seizures or unconsciousness. 7. Subjects who have experienced diabetes-related complications requiring assistance from another person in the last six months. 8. Subject currently is participating in another clinical trial. 9. Subject has donated blood within 112 days (3.7 months) prior to the beginning of the study activities. 10. Subject is anemic. 11. Subject does not have a known insulin to carbohydrate ratio at the time of enrollment (IIT subjects only). 12. Subject has concomitant medical condition which, in the opinion of the investigator, could interfere with the study or present a risk to the safety or welfare of the subject or study staff. Such conditions include but are not limited to: \- History of HIV, Hepatitis B or C 13. Subject has X-ray, MRI or CT appointment scheduled during the period of study participation, and the appointment cannot be rescheduled for a time before study participation starts or after study participation ends. 14. Subject has an abnormal EKG, unless cleared for study participation by a cardiologist. 15. Subject is unsuitable for participation due to any other cause as determined by the Investigator. Additional Criteria 16. The following subjects may participate in the study but will be excluded from all glycemic and exercise challenges: * Subject's taking medications known to mask symptoms or hypoglycemia, such as beta blockers; * Subjects with any of the following: clinically significant history of cardiovascular or cerebrovascular disease, cardiac arrhythmia, neurological disorders such as seizures, CVA, or syncope, or hypokalemia.

Design outcomes

Primary

MeasureTime frameDescription
The Device Performance Was Evaluated in Terms of Point Accuracy of the Abbott Sensor Based Glucose Monitoring Systems in Reference to Yellow Spring Instrument (YSI).14 daysPoint accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels \<80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had up to 136 YSI samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels \<80 mg/dL.
Safety of the Abbott Sensor Based Glucose Monitoring Systems Was Characterized by Adverse Device Effects and Serious Adverse Device Effects Experienced by Study Participants.Safety was evaluated throughout the subject's study participation.Safety of the Abbott Sensor Based Glucose Monitoring Systems, including adverse device effects and serious adverse device effects were assessed for all participants enrolled in the study.

Secondary

MeasureTime frameDescription
The Trend Accuracy of the System P Device Performance Was Evaluated.14 daysTrend accuracy of the device performance was assessed by comparing of the glucose rate of change results between the sensor glucose readings and YSI reference results for System P only.
The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated.14 daysThe Continuous Glucose Error Grid Analysis (CG-EGA) in the Hypoglycemic range (YSI ≤ 70 mg/dL), in the Euglycemic range (70 \< YSI ≤ 180 mg/dL) and in the Hyperglycemic range (YSI \> 180 mg/dL) were evaluated.
The Point Accuracy of the Device Performance Was Evaluated in Reference to Capillary Blood Glucose (BG).14 daysPoint accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the capillary blood glucose (BG) value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for capillary blood glucose (BG) levels \<80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had approximately112 capillary blood glucose (BG) samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the BG value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for BG levels \<80 mg/dL.

Countries

United States

Participant flow

Participants by arm

ArmCount
Diabetes Mellitus, Type 1 and Type 2
Subjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes. Abbott Sensor Based Glucose Monitoring System: Subjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes.
125
Total125

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyScreen Failed28
Overall StudyWithdrawal by Subject3

Baseline characteristics

CharacteristicDiabetes Mellitus, Type 1 and Type 2
Age, Continuous46.7 years
STANDARD_DEVIATION 14.4
Region of Enrollment
United States
125 Participants
Sex: Female, Male
Female
64 Participants
Sex: Female, Male
Male
61 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 125
other
Total, other adverse events
6 / 125
serious
Total, serious adverse events
0 / 125

Outcome results

Primary

Safety of the Abbott Sensor Based Glucose Monitoring Systems Was Characterized by Adverse Device Effects and Serious Adverse Device Effects Experienced by Study Participants.

Safety of the Abbott Sensor Based Glucose Monitoring Systems, including adverse device effects and serious adverse device effects were assessed for all participants enrolled in the study.

Time frame: Safety was evaluated throughout the subject's study participation.

Population: All 125 subjects who participated in the study are included in this outcome measure.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Diabetes Mellitus, Type 1 and Type 2Safety of the Abbott Sensor Based Glucose Monitoring Systems Was Characterized by Adverse Device Effects and Serious Adverse Device Effects Experienced by Study Participants.6 Participants
Primary

The Device Performance Was Evaluated in Terms of Point Accuracy of the Abbott Sensor Based Glucose Monitoring Systems in Reference to Yellow Spring Instrument (YSI).

Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels \<80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had up to 136 YSI samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels \<80 mg/dL.

Time frame: 14 days

Population: A total number of 14284 paired points between sensor glucose readings and YSI glucose are used as the denominator to calculate the percentage of sensor glucose readings that are within ±20% of the YSI reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for YSI glucose levels \<80 mg/dL.

ArmMeasureGroupValue (NUMBER)
Diabetes Mellitus, Type 1 and Type 2The Device Performance Was Evaluated in Terms of Point Accuracy of the Abbott Sensor Based Glucose Monitoring Systems in Reference to Yellow Spring Instrument (YSI).System Personal82.9 percentage of paired points
Diabetes Mellitus, Type 1 and Type 2The Device Performance Was Evaluated in Terms of Point Accuracy of the Abbott Sensor Based Glucose Monitoring Systems in Reference to Yellow Spring Instrument (YSI).System Professional86.4 percentage of paired points
Secondary

The Point Accuracy of the Device Performance Was Evaluated in Reference to Capillary Blood Glucose (BG).

Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the capillary blood glucose (BG) value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for capillary blood glucose (BG) levels \<80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had approximately112 capillary blood glucose (BG) samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the BG value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for BG levels \<80 mg/dL.

Time frame: 14 days

Population: A total number of 11444 paired points between sensor glucose readings and BG glucose are used as the denominator to calculate the percentage of sensor glucose readings that are within ±20% of the BG reference value for glucose levels ≥ 80 mg/dL and within ±20 mg/dL for BG glucose levels \<80 mg/dL.

ArmMeasureGroupValue (NUMBER)
Diabetes Mellitus, Type 1 and Type 2The Point Accuracy of the Device Performance Was Evaluated in Reference to Capillary Blood Glucose (BG).System P83.6 Percentage of paired points
Diabetes Mellitus, Type 1 and Type 2The Point Accuracy of the Device Performance Was Evaluated in Reference to Capillary Blood Glucose (BG).System Pro87.8 Percentage of paired points
Secondary

The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated.

The Continuous Glucose Error Grid Analysis (CG-EGA) in the Hypoglycemic range (YSI ≤ 70 mg/dL), in the Euglycemic range (70 \< YSI ≤ 180 mg/dL) and in the Hyperglycemic range (YSI \> 180 mg/dL) were evaluated.

Time frame: 14 days

Population: A total of 13566 paired points between sensor glucose readings and YSI glucose were used to calculate and evaluate the percentage of sensor glucose readings at different glucose rates of change and different glucose ranges (hypoglycemic, euglycemic, and hyperglycemic ranges) in System P only.

ArmMeasureGroupValue (NUMBER)
Diabetes Mellitus, Type 1 and Type 2The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated.in the Hypoglycemic range (YSI ≤ 70 mg/dL)86.6 % of paired points
Diabetes Mellitus, Type 1 and Type 2The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated.In the Euglycemic range (70 < YSI ≤ 180 mg/dL)97.8 % of paired points
Diabetes Mellitus, Type 1 and Type 2The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated.In the Hyperglycemic range (YSI > 180 mg/dL)94.5 % of paired points
Secondary

The Trend Accuracy of the System P Device Performance Was Evaluated.

Trend accuracy of the device performance was assessed by comparing of the glucose rate of change results between the sensor glucose readings and YSI reference results for System P only.

Time frame: 14 days

Population: A total number of 13566 paired points between sensor glucose readings and YSI glucose are used as the denominator to calculate the absolute difference between the rates of change corresponding to Sensor Reading (GM) and YSI results.

ArmMeasureValue (NUMBER)
Diabetes Mellitus, Type 1 and Type 2The Trend Accuracy of the System P Device Performance Was Evaluated.82.8 Percentage of paired points