Evaluation of the Accuracy of the Hemochron Signature Elite Point of Care Device Compared With Laboratory Assay

Evaluation of the Accuracy of the Hemochron Signature Elite Point of Care Micro Coagulation Instrument Compared With Laboratory Assay

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02255214
Enrollment
24
Registered
2014-10-02
Start date
2013-10-01
Completion date
2015-05-01
Last updated
2026-01-30

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Coagulation Tests

Brief summary

Central Neuraxial Blockade (CNB) is commonly used in anaesthetics. It has several advantages over a general anaesthetic however contraindications include coagulopathy. Guidelines suggest a CNB should not be undertaken with an International Normalised Ratio (INR) of 1.5 or greater. A common cause of an elevated INR is warfarin therapy and regular monitoring of its effects is required. The Hemochron Signature Elite is a portable micro coagulation system that can be used as a point of care (POC) anticoagulation monitoring device. Marketing is focussed on its accuracy, convenience and efficiency in delivering a coagulation result. This study will assess the accuracy of this device in calculating the International Normalised Ratio (INR) result. This will be compared to another INR result obtained on the same sample via the current gold standard testing equipment in the Royal Infirmary Edinburgh (RIE) laboratories. The study will include patients presenting for surgery who have had their warfarin therapy stopped within the previous 7 days. Patients on warfarin therapy requiring surgery are currently identified in preoperative assessment clinics, on average 12 weeks prior to their date of surgery. Anecdotally there appears to be a difference between the 'expected' and actual INR when a POC device is used. Concern exists that patients may be inappropriately denied the potential benefits of central neural blockade (CNB) if a POC device provides a falsely high result, or conversely the investigators may be proceeding with CNB in patients with an unacceptably high INR if POC provides a falsely low result. There is also the possibility of improving hospital efficiency using this device as it provides a result much faster than the laboratory.

Interventions

None listed

Sponsors

NHS Lothian
Lead SponsorOTHER_GOV

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* Adult patients presenting for elective surgery who have stopped their warfarin therapy within the last 7 days.

Exclusion criteria

* Adults with incapacity

Design outcomes

Primary

MeasureTime frameDescription
The International Normalised Ratio (INR) Will be Measured by the Laboratory Assay and the Hemochron Signature Elite Device.18 monthsThe correlation between the international normalised ratio measured by the laboratory assay and the Hemochron Signature Elite device was assessed.Number of participants with a higher INR measurement by the Hemochron Signature Elite device than laboratory assay

Countries

United Kingdom

Contacts

PRINCIPAL_INVESTIGATORCameron R Ferguson, MBChB MRCP

NHS Lothian

Baseline characteristics

Characteristic
Age, Customized
Age not recorded
0 Participants
Region of Enrollment
United Kingdom
24 participants
Sex/Gender, Customized
Sex not recorded
0 Participants

Adverse events

Event typeEG000
affected / at risk
other
Total, other adverse events
0 / 24
serious
Total, serious adverse events
0 / 24

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026