Skip to content

Three-way, Cross-over Closed-loop Exercise Study

A Randomized, Three-way, Cross-over Study to Assess the Efficacy of a Bi-hormonal Closed-loop System With vs Without Exercise Announcement vs Open-loop Therapy in Reducing Exercise-related Hypoglycemia

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02241889
Enrollment
23
Registered
2014-09-16
Start date
2015-01-31
Completion date
2015-09-30
Last updated
2017-05-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Type 1 Diabetes

Keywords

diabetes, closed-loop, bi-hormonal, exercise

Brief summary

Blood glucose control systems that utilize both insulin and glucagon to manage blood sugar are paving the way to revolutionize the management of this disease. The benefit of improved control of blood sugar levels compared to standard insulin pump therapy has already been demonstrated. However, the risk of low blood sugar in type 1 diabetes increases considerably during exercise. The investigators research group has shown that small doses of glucagon can prevent low blood sugar when used in a blood sugar control system in patients with type 1 diabetes. However, as insulin sensitivity changes occur very rapidly during exercise, the ability to recognize the onset of physical activity in order to appropriately adjust the insulin and glucagon delivery is vital to helping prevent exercise-induced hypoglycemia. The purpose of this study is to test how well a new modified blood sugar control system controls blood sugar during exercise compared to: 1) the current system without modifications and 2) standard insulin pump therapy.

Detailed description

Subjects will exercise during three study visits. During one visit, blood sugar will be controlled using standard insulin pump therapy. The investigators blood glucose control system will manage blood sugar for subjects during the other visits, with exercise announcement with insulin and glucagon delivery modifications during one visit. Treatment order will be randomized. Heart rate and accelerometry data will also be collected during each visit for the purposes of designing an exercise detection system for future studies. Each subject will be asked to undergo an eligibility screening and to sign a consent form before any study procedures take place. At the screening appointment, after the consent form is signed, blood will be drawn for laboratory tests and a brief physical exam will be performed. An EKG and VO2max test will be completed at screening. Subjects that are enrolled in the study will come in for a sensor insertion visit up to 72 hours before each study visit. Subjects will undergo three 21-hour studies at an OHSU clinic or the OHSU clinical research center. Subjects will arrive at the research center at 8pm. Approximately 2 hours after a low-carbohydrate breakfast on Day 2, subjects will exercise for 45 minutes at 60% of their VO2max determined at screening. Three hours after completion of exercise, subjects will eat lunch. The study will end \ 5 hours after lunch. Capillary blood glucose (CBG) will be measured every 2 hours during the day, every 3 hours at night and before and after exercise for studies using the blood glucose control system with subjects blinded to the CBG values. For studies using insulin pump therapy, the subject will sample CBG four times during the day as well as before and after exercise. Study staff will monitor studies remotely using a cloud server. An investigator or nurse practitioner with a specialty in endocrinology or internal medicine will be on site (located within the same complex) at all times.

Interventions

Subject's own insulin pump will be used to manage blood glucose.

DEVICEClosed-loop Artificial Pancreas Controller

Closed-loop Artificial Pancreas Controller includes insulin and glucagon delivery algorithm implemented on a smart phone, utilizing sensor glucose values from a Dexcom G4 sensor and sending delivery commands to two Tandem t:slim insulin pumps, one filled with insulin and one with glucagon.

Sponsors

Oregon Health and Science University
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
21 Years to 45 Years
Healthy volunteers
Yes

Inclusion criteria

* Diagnosis of type 1 diabetes mellitus for at least 1 year. * Male or female subjects 21 to 45 years of age. * Physically willing and able to perform 45 min of exercise (as determined by the investigator after reviewing the subjects activity level) * Current use of an insulin pump. * Willingness to follow all study procedures, including attending all clinic visits. * Willingness to sign informed consent and HIPAA documents.

Exclusion criteria

* Female of childbearing potential who is pregnant or intending to become pregnant or breast-feeding, or is not using adequate contraceptive methods. Acceptable contraception includes birth control pill / patch / vaginal ring, Depo-Provera, Norplant, an IUD, the double barrier method (the woman uses a diaphragm and spermicide and the man uses a condom), or abstinence. * Any cardiovascular disease, defined as a clinically significant EKG abnormality at the time of screening or any history of: stroke, heart failure, myocardial infarction, angina pectoris, or coronary arterial bypass graft or angioplasty. Diagnosis of 2nd or 3rd degree heart block or any non-physiological arrhythmia judged by the investigator to be exclusionary. * Renal insufficiency (GFR \< 60 ml/min, using the MDRD equation as report by the OHSU laboratory). * Impaired liver function, defined as AST or ALT ≥2.5 times upper limit of normal, according to OHSU laboratory reference ranges. * Hematocrit of less than or equal to 34%. * History of severe hypoglycemia during the past 12 months prior to screening visit or hypoglycemia unawareness as judged by the investigator. * Adrenal insufficiency. * Any active infection. * Known or suspected abuse of alcohol, narcotics, or illicit drugs. * Seizure disorder. * Active foot ulceration. * Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication. * Major surgical operation within 30 days prior to screening. * Use of an investigational drug within 30 days prior to screening. * Chronic usage of any immunosuppressive medication (such as cyclosporine, azathioprine, sirolimus, or tacrolimus). * Bleeding disorder, treatment with warfarin, or platelet count below 50,000. * Allergy to aspart insulin. * Allergy to glucagon. * Insulin resistance requiring more than 200 units per day. * Need for uninterrupted treatment of acetaminophen. * Current administration of oral or parenteral corticosteroids. * Any life threatening disease, including malignant neoplasms and medical history of malignant neoplasms within the past 5 years prior to screening (except basal and squamous cell skin cancer). * C peptide level of ≥0.5 ng/ml * Any concurrent illness, other than diabetes, that is not controlled by a stable therapeutic regimen. * Beta blockers or non-dihydropyridine calcium channel blockers. * A positive response to any of the questions from the Physical Activity Readiness Questionnaire. * Any clinically significant disease or disorder which in the opinion of the Investigator may jeopardize the subject's safety or compliance with the protocol.

Design outcomes

Primary

MeasureTime frameDescription
Percent of Time With Sensed Glucose < 70 mg/dlfrom start of exercise (~hour 12) until study completion (hour 21)Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values less than 70 mg/dl using Dexcom sensor downloads.
Percent of Time With Sensed Glucose Between 70-180 mg/dlfrom start of exercise (~hour 12) until study completion (hour 21)Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values between 70-180 mg/dl using Dexcom sensor downloads.

Secondary

MeasureTime frameDescription
Percent of Time With Sensed Glucose < 50 mg/dlEntire 21 hour study duration excluding the first four hoursAssess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values less than 50 mg/dl using Dexcom sensor downloads.
Percent of Time With Sensed Glucose > 180 mg/dlEntire 21 hour study duration excluding the first four hoursAssess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values greater than 180 mg/dl using Dexcom sensor downloads.
Number of Events Capillary Blood Glucose (CBG) <70 mg/dlEntire 21 hour study duration excluding the first four hoursNumber of events measured with capillary blood glucose \<70 mg/dl.
Mean of the Mean Sensed Glucose Per ParticipantEntire 21 hour study duration excluding the first four hoursAssess the mean sensed glucose per participant using Dexcom sensor downloads.
Number of Events With CBG <50 mg/dlEntire 21 hour study duration excluding the first four hoursAssess the total number of events that the Contour Next BG meter reported blood glucose values less than 50 mg/dl across all participants in each group.
Percent of Time of CBG>180 mg/dlEntire 21 hour study duration excluding the first four hoursAssess the percent of time that the Contour Next BG meter reported blood glucose values greater than 180 mg/dl using meter downloads.
Number of Events With CBG Between 70 - 180 mg/dlEntire 21 hour study duration excluding the first four hoursAssess the number of events that the Contour Next BG meter reported blood glucose values between 70-180 mg/dl using meter downloads.
Number of Carbohydrate TreatmentsEntire 21 hour study duration excluding the first four hoursAssess the number of rescue carbohydrate treatments.

Countries

United States

Participant flow

Participants by arm

ArmCount
All Study Participants
All study participants underwent three 21 hour study visits using either standard of care insulin pump therapy, closed loop control, or closed loop control with adjustments to insulin and glucagon after exercise announcement. Treatment order was randomized.
23
Total23

Withdrawals & dropouts

PeriodReasonFG000FG001FG002FG003FG004FG005
Overall Studydevice issues001000
Overall StudyWithdrawal by Subject000001

Baseline characteristics

CharacteristicAll Study Participants
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
23 Participants
Age, Continuous32 years
STANDARD_DEVIATION 7
Region of Enrollment
United States
23 participants
Sex: Female, Male
Female
14 Participants
Sex: Female, Male
Male
9 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
deaths
Total, all-cause mortality
0 / 230 / 230 / 23
other
Total, other adverse events
1 / 233 / 233 / 23
serious
Total, serious adverse events
0 / 230 / 230 / 23

Outcome results

Primary

Percent of Time With Sensed Glucose < 70 mg/dl

Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values less than 70 mg/dl using Dexcom sensor downloads.

Time frame: from start of exercise (~hour 12) until study completion (hour 21)

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyPercent of Time With Sensed Glucose < 70 mg/dl0.9 percentage of time
Closed-loop Without AdjustmentPercent of Time With Sensed Glucose < 70 mg/dl3.2 percentage of time
Closed-loop With AdjustmentPercent of Time With Sensed Glucose < 70 mg/dl0.3 percentage of time
Primary

Percent of Time With Sensed Glucose Between 70-180 mg/dl

Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values between 70-180 mg/dl using Dexcom sensor downloads.

Time frame: from start of exercise (~hour 12) until study completion (hour 21)

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyPercent of Time With Sensed Glucose Between 70-180 mg/dl67.6 percent of time
Closed-loop Without AdjustmentPercent of Time With Sensed Glucose Between 70-180 mg/dl72.5 percent of time
Closed-loop With AdjustmentPercent of Time With Sensed Glucose Between 70-180 mg/dl67.3 percent of time
Secondary

Mean of the Mean Sensed Glucose Per Participant

Assess the mean sensed glucose per participant using Dexcom sensor downloads.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyMean of the Mean Sensed Glucose Per Participant8.6 mg/dl
Closed-loop Without AdjustmentMean of the Mean Sensed Glucose Per Participant8.0 mg/dl
Closed-loop With AdjustmentMean of the Mean Sensed Glucose Per Participant8.6 mg/dl
Secondary

Number of Carbohydrate Treatments

Assess the number of rescue carbohydrate treatments.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyNumber of Carbohydrate Treatments0.33 carbohydrate treatments
Closed-loop Without AdjustmentNumber of Carbohydrate Treatments0.43 carbohydrate treatments
Closed-loop With AdjustmentNumber of Carbohydrate Treatments0.29 carbohydrate treatments
Secondary

Number of Events Capillary Blood Glucose (CBG) <70 mg/dl

Number of events measured with capillary blood glucose \<70 mg/dl.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyNumber of Events Capillary Blood Glucose (CBG) <70 mg/dl0.43 occurrences of blood glucose < 70 mg/dl
Closed-loop Without AdjustmentNumber of Events Capillary Blood Glucose (CBG) <70 mg/dl0.33 occurrences of blood glucose < 70 mg/dl
Closed-loop With AdjustmentNumber of Events Capillary Blood Glucose (CBG) <70 mg/dl0.24 occurrences of blood glucose < 70 mg/dl
Secondary

Number of Events With CBG <50 mg/dl

Assess the total number of events that the Contour Next BG meter reported blood glucose values less than 50 mg/dl across all participants in each group.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (NUMBER)
Standard Insulin Pump TherapyNumber of Events With CBG <50 mg/dl1 occurrences of blood glucose < 50 mg/dl
Closed-loop Without AdjustmentNumber of Events With CBG <50 mg/dl0 occurrences of blood glucose < 50 mg/dl
Closed-loop With AdjustmentNumber of Events With CBG <50 mg/dl1 occurrences of blood glucose < 50 mg/dl
Secondary

Number of Events With CBG Between 70 - 180 mg/dl

Assess the number of events that the Contour Next BG meter reported blood glucose values between 70-180 mg/dl using meter downloads.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyNumber of Events With CBG Between 70 - 180 mg/dl1.6 # of events
Closed-loop Without AdjustmentNumber of Events With CBG Between 70 - 180 mg/dl3.1 # of events
Closed-loop With AdjustmentNumber of Events With CBG Between 70 - 180 mg/dl3.0 # of events
Secondary

Percent of Time of CBG>180 mg/dl

Assess the percent of time that the Contour Next BG meter reported blood glucose values greater than 180 mg/dl using meter downloads.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyPercent of Time of CBG>180 mg/dl1.1 percentage of time
Closed-loop Without AdjustmentPercent of Time of CBG>180 mg/dl0.7 percentage of time
Closed-loop With AdjustmentPercent of Time of CBG>180 mg/dl1.1 percentage of time
Secondary

Percent of Time With Sensed Glucose > 180 mg/dl

Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values greater than 180 mg/dl using Dexcom sensor downloads.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyPercent of Time With Sensed Glucose > 180 mg/dl26.9 percent of time
Closed-loop Without AdjustmentPercent of Time With Sensed Glucose > 180 mg/dl16.9 percent of time
Closed-loop With AdjustmentPercent of Time With Sensed Glucose > 180 mg/dl25.2 percent of time
Secondary

Percent of Time With Sensed Glucose < 50 mg/dl

Assess the percent of time that the Dexcom G4 or G4 Share reported sensor glucose values less than 50 mg/dl using Dexcom sensor downloads.

Time frame: Entire 21 hour study duration excluding the first four hours

Population: Data was analyzed from 21 subjects that completed all 3 visits and visits went to completion. No data was analyzed from the two subjects that did not complete all visits.

ArmMeasureValue (MEAN)
Standard Insulin Pump TherapyPercent of Time With Sensed Glucose < 50 mg/dl0.001 percent of time
Closed-loop Without AdjustmentPercent of Time With Sensed Glucose < 50 mg/dl0.001 percent of time
Closed-loop With AdjustmentPercent of Time With Sensed Glucose < 50 mg/dl0.0 percent of time

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026