Development and Assessment of Feasibility of Non-invasive Multiple Sensor Hypo-Sense as a Tool for Detection of Hypoglycemia

Development and Assessment of Feasibility of Non-invasive Multiple Sensor Hypo-Sense Sensor as a Tool for Detection of Hypoglycemia - Exploratory Study for Prototype Rev 1

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02225379
Acronym
Hypo-Sense
Enrollment
3
Registered
2014-08-26
Start date
2014-09-30
Completion date
2016-07-31
Last updated
2017-01-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Nocturnal Hypoglycemia, Type 1 Diabetes

Keywords

Type 1 Diabetes, Hypoglycemia detection, Non-invasive sensor, Nocturnal hypoglycemia

Brief summary

Hypo Sense is a non- invasive method for detection of hypoglycemia. The Hypo Sense combines an array of non-invasive sensors which monitors the patient's physiological parameters (heart & respiration rate, perspiration, skin temperature and arm motion) designed as a wrist watch device. The Hypo sense is intended for monitoring symptoms of hypoglycemia in diabetic patients in hospital environment among type 1 and type 2 diabetes adults as an adjunctive device to reference methods The proposed study will be consisting of two main segments: The primary aim of segment 1 of the study is data collection and calibration of the Hypo Sense sensor prototype compared to standard invasive reference glucometer. The primary aim of segment 2 of the study is to validate the Hypo Sense prototype performance in detecting hypoglycemic events. During the first segment of the study we intend to collect in parallel measurements of blood glucose using reference method (capillary glucometer) and continuous data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced. The reference and study device data will be analyzed using multivariate regression model to formulate a calibration algorithm model. This model will translate the set of the physiological recorded parameters into detection of hypoglycemic events. During the second segment of the study we intend to evaluate the validity of the Hypo Sense sensor ability to detect hypoglycemic events compared to standard invasive reference method (capillary glucometer).

Interventions

DEVICEHypo-Sense (non invasive sensor)

Parallel measurements of capillary blood glucose using reference methods (both capillary glucometer and continuous sensor) and data generated by the non- invasive study device during approximately 4 hours, in which a hypoglycemic event will be induced.

Sponsors

Night Sense Ltd
CollaboratorUNKNOWN
Rabin Medical Center
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
PREVENTION
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

* Signing an inform consent form prior to any trial related procedure * Type 1 diabetes diagnosed at least 12 months prior to study inclusion * Age \> 18 years old

Exclusion criteria

* Participating in other device or drug study * Any significant disease or condition, including psychiatric disorders that in the opinion of the investigator is likely to affect patient's compliance or ability to complete the study * Patients with one or more of the following diseases: malignancy, myocardial insufficiency, nephrologic disease or any other chronic disease * Chronic skin problem in the lower inner arm * Pregnant or breast feeding women

Design outcomes

Primary

MeasureTime frameDescription
SensitivityAt final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)Sensitivity of the Hypo Sense-the proportion of hypoglycemia events detected by the Hypo Sense sensor from total number of hypoglycemic events detected by reference capillary glucometer
Positive Predictive Value of the hypo-SenseAt final visit ( week 4 for participants at segment 1 of the study and week 6 for participants at segment 2 of the study)Positive Predictive Value of the hypo sense- the proportion of validated (true) hypoglycemia events detected by the hypo sense sensor from total alerts detected by the Hypo-Sense sensor

Countries

Israel

Outcome results

None listed