Radiofrequency Ablation in Patients With Spinal Hardware
Conditions
Keywords
radiofrequency ablation
Brief summary
The goal of the study was to determine if patients undergoing radiofrequency neurotomy (RFN) of lumbar facet joints that are adjacent to previously placed pedicle screws from a fusion, would lead to an increase in the temperature of the pedicle screws.
Detailed description
This study was a prospective clinical trial, examining outcomes related to temperature increases in pedicle screws. It was not possible to blind the investigators or the patient. Outcomes related temperature differences were collected at each pedicle screw adjacent to a zygapophyseal joint being treated. A conventional 18-gauge, 1-cm curved active tip radiofrequency cannula was placed under fluoroscopic guidance to the area to be treated (the junction of the transverse process and the superior articulating process of the facet joint). This position placed the tip of the radiofrequency cannula immediately adjacent to, but not in direct contact with, the pedicle screws. The radiofrequency cannula was connected to a radiofrequency generator, and the conventional settings for clinical medial branch neurotomy was used, heating the tissue to 80°C for 90 seconds. The procedure described up to this point is the normal procedure and protocol for RFN performed at the Mayo Clinic. The experimental portion of the procedure is as follows: A separate 22 gauge radiofrequency cannula was intentionally placed in direct contact with the dorsal surface of the pedicle screws and was connected to a radiofrequency generator solely for the purpose of measuring temperature, as the RF cannula are dual purpose probes capable of both conducting energy for the radiofrequency lesion and monitoring tissue temperature. In other words, this small gauge needle was placed for the sole purpose of acting as a thermistor probe/temperature monitor. This process was repeated for each level necessary for treatment.
Interventions
Subjects with spinal hardware that are undergoing radiofrequency ablation procedures will have an additional radiofrequency cannula placed at the site of adjacent pedicle screws for temperature measurement only.
Sponsors
Study design
Eligibility
Inclusion criteria
\- Scheduled for Radiofrequency procedure AND has spinal hardware placed from a previous spinal surgery
Exclusion criteria
* Active infection over skin where needle will be placed * Coagulopathy * Pregnancy
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Procedures in Which Temperature of Screws Increased | 2 hours | The study will measure for increased temperature in pedicle screws adjacent to lumbar facet joints during radiofrequency neurotomy (RFN). Temperatures will be recorded by placing a thermistor probe on the surface of the adjacent hardware. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Radiofrequency Neurotomy Subjects Subjects with spinal hardware that are undergoing radiofrequency ablation procedures will have an additional radiofrequency cannula placed at the site of adjacent pedicle screws for temperature measurement only. | 6 |
| Total | 6 |
Baseline characteristics
| Characteristic | Radiofrequency Neurotomy Subjects |
|---|---|
| Age, Categorical <=18 years | 0 Participants |
| Age, Categorical >=65 years | 1 Participants |
| Age, Categorical Between 18 and 65 years | 5 Participants |
| Age, Continuous | 47.8 years |
| Region of Enrollment United States | 6 participants |
| Sex: Female, Male Female | 1 Participants |
| Sex: Female, Male Male | 5 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | 0 / 6 |
| other Total, other adverse events | 0 / 6 |
| serious Total, serious adverse events | 0 / 6 |
Outcome results
Number of Procedures in Which Temperature of Screws Increased
The study will measure for increased temperature in pedicle screws adjacent to lumbar facet joints during radiofrequency neurotomy (RFN). Temperatures will be recorded by placing a thermistor probe on the surface of the adjacent hardware.
Time frame: 2 hours
Population: There were 6 participants, but some participants had more than 1 procedure. There were a total of 10 procedures analyzed.
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Radiofrequency Neurotomy Subjects | Number of Procedures in Which Temperature of Screws Increased | Non-significant increase | 4 procedures |
| Radiofrequency Neurotomy Subjects | Number of Procedures in Which Temperature of Screws Increased | Significant increase; procedure stopped | 2 procedures |