Skip to content

Evaluation of a Novel Electronic Urine Output Monitor (eUOM)

Evaluation of a Novel Electronic Urine Output Monitor (eUOM)

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02195713
Enrollment
20
Registered
2014-07-21
Start date
2014-07-31
Completion date
2015-02-28
Last updated
2016-02-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Oliguria, Polyuria

Brief summary

Urine output and urine drain line pressure were monitored while urine was drained into either: 1. Accuryn Urine Output Monitor (Potrero Medical) OR 2. Criticore Monitor (Bard Medical)

Detailed description

The purpose of this study is to determine the characteristics of urine output and drain line pressure as measured with the Accuryn urinary output monitoring system and urine output as recorded by a commercially available system (Criticore, Bard Medical) when used with a standard Foley catheter and standard urinary drainage tube.

Interventions

DEVICEAccuryn

Accuryn is a novel electronic urine output monitor

Sponsors

Potrero Medical
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patient must ≥ 18 years of age * Patient has a Foley catheter and urine collection system is in place per standard clinical decision * Estimated length of placement of the Foley is 48 hours minimum * Burn injury ≥ 20% and ≤ 80% total body surface area (TBSA) * Subject or subject's legally authorized representative is able to give informed consent before entering the study

Exclusion criteria

* Currently pregnant or breastfeeding * Clinical signs or symptoms of a urinary tract infection (UTI) * Clinical signs or symptoms of a vaginal infection * Currently has bladder or urethral trauma * Use of investigational drug/device therapy within the past 4 weeks

Design outcomes

Primary

MeasureTime frameDescription
Urine Output of SOC Device Versus Accuryn2 - 5 daysUrine output as recorded by the SOC when connected to a standard Foley catheter and standard urinary drainage tube will be compared to urine output recorded by Accuryn and the SOC when connected to a standard Foley catheter and the Accuryn Drainage Tube.

Countries

United States

Participant flow

Participants by arm

ArmCount
Test- Accuryn Urine Output Monitor
Accuryn Anti-airlock Drainage System was attached to the Foley catheter and urine output / drainage line pressure was monitored. Accuryn: Accuryn is a novel electronic urine output monitor
9
Control- Critcore Urine Output Monitor
A commercially available urine output monitor (Criticore, Bard Medical) was attached to the Foley catheter and urine output / drainage line pressure was monitored.
9
Total18

Baseline characteristics

CharacteristicTest- Accuryn Urine Output MonitorControl- Critcore Urine Output MonitorTotal
Age, Continuous44 years
STANDARD_DEVIATION 22
52 years
STANDARD_DEVIATION 17.9
49 years
STANDARD_DEVIATION 20.2
Region of Enrollment
United States
9 participants9 participants18 participants
Sex: Female, Male
Female
2 Participants1 Participants3 Participants
Sex: Female, Male
Male
7 Participants8 Participants15 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
— / —— / —
other
Total, other adverse events
0 / 00 / 0
serious
Total, serious adverse events
0 / 00 / 0

Outcome results

Primary

Urine Output of SOC Device Versus Accuryn

Urine output as recorded by the SOC when connected to a standard Foley catheter and standard urinary drainage tube will be compared to urine output recorded by Accuryn and the SOC when connected to a standard Foley catheter and the Accuryn Drainage Tube.

Time frame: 2 - 5 days

ArmMeasureValue (NUMBER)
Test- Accuryn Urine Output MonitorUrine Output of SOC Device Versus Accuryn0 Patients with airlocks
Control- Critcore Urine Output MonitorUrine Output of SOC Device Versus Accuryn5 Patients with airlocks

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026