Clinical Assessment of the External Sound Processor Worn by Patients Implanted With Codacs

Clinical Validation of the Nucleus® CP810 Sound Processor for the Codacs™ System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02156167
Enrollment
15
Registered
2014-06-05
Start date
2014-03-31
Completion date
2015-06-30
Last updated
2016-07-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Severe to Profound Mixed Hearing Loss

Keywords

Otosclerosis, cochlear implant, acoustic stimulation

Brief summary

The purpose of this study is to evaluate the effect of the sound processor upgrade from the C-DACS investigational device Freedom sound processor to the Nucleus® CP810 sound processor for the Codacs™ system on the speech reception threshold in noise, to evaluate the usability of the Codacs™ Fitting Software, to evaluate the quality of life with the Nucleus® CP810 Sound Processor for the Codacs™ system, to collect long term data, to evaluate the acceptance criteria of the postoperative Codacs™ system test and to evaluate the usability of the Codacs™ Test System.

Interventions

DEVICECP810 and Codacs™ system test

Nucleus® CP810 Sound Processor for the Codacs™ system (CE marked) and Codacs™ Test System (CE marked)

Sponsors

Cochlear
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* Subjects implanted with a Cochlear C-DACS investigational device

Exclusion criteria

* Participation in another medical device study * Unwillingness or inability of the subject to comply with the study requirements

Design outcomes

Primary

MeasureTime frameDescription
Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence Test3 months after initial upgraded fittingThe Oldenburger Sentence test is an adaptive speech in noise test with a fixed noise level of typically 65 decibel (dB) sound pressure level (SPL) and a varying speech level, depending on how many words in a sentence were repeated correctly by the subject. The outcome of this test is the signal-to-noise ratio (SNR) at which 50% of words in the sentence list were correctly repeated by the subject. The SRT in noise with the Freedom sound processor at the initial study visit was compared to the SRT in noise with the CP810 at the 3 months follow-up at different measurement conditions. Speech and noise coming from the front (S0N0)- Speech coming from the front and noise coming from the implanted side (S0N90)- Everyday program (E): omnidirectional microphone- Noise program (N): directional microphone- Freedom Sound Processor (Freedom)- CP810 Sound Processor (CP810).

Countries

Germany

Participant flow

Recruitment details

Center contacted their patients, who received a Codacs device in another clinical trial by phone in October 2013

Participants by arm

ArmCount
CP810
Nucleus® CP810 Sound Processor for the Codacs™ system (CE marked)
15
Total15

Baseline characteristics

CharacteristicCP810
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
9 Participants
Age, Categorical
Between 18 and 65 years
6 Participants
Age, Continuous66 years
Region of Enrollment
Germany
15 participants
Sex: Female, Male
Female
12 Participants
Sex: Female, Male
Male
3 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
0 / 15
serious
Total, serious adverse events
0 / 15

Outcome results

Primary

Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence Test

The Oldenburger Sentence test is an adaptive speech in noise test with a fixed noise level of typically 65 decibel (dB) sound pressure level (SPL) and a varying speech level, depending on how many words in a sentence were repeated correctly by the subject. The outcome of this test is the signal-to-noise ratio (SNR) at which 50% of words in the sentence list were correctly repeated by the subject. The SRT in noise with the Freedom sound processor at the initial study visit was compared to the SRT in noise with the CP810 at the 3 months follow-up at different measurement conditions. Speech and noise coming from the front (S0N0)- Speech coming from the front and noise coming from the implanted side (S0N90)- Everyday program (E): omnidirectional microphone- Noise program (N): directional microphone- Freedom Sound Processor (Freedom)- CP810 Sound Processor (CP810).

Time frame: 3 months after initial upgraded fitting

Population: One subject did not visit the clinic for the 3 months follow-up, a second subject did not perform the speech in noise test at the 3 months follow-up due to tiredness

ArmMeasureGroupValue (MEAN)Dispersion
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N0 - E - Freedom-0.83 dB SNRStandard Deviation 2.04
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N0 - E - CP810-1.45 dB SNRStandard Deviation 2.29
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N90 - E - Freedom0.64 dB SNRStandard Deviation 2.92
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N90 - E - CP8100.89 dB SNRStandard Deviation 2.83
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N0 - N - CP810-1.22 dB SNRStandard Deviation 1.69
CP810Aided Speech Reception Threshold (SRT) in Noise Measured by Signal to Noise Ratio for 50% Correct Scores With the Oldenburger Sentence TestS0N90 - N - CP810-3.78 dB SNRStandard Deviation 3.33