Evaluation of an Karajishi Contour Investigational Blood Glucose Monitoring System (ISO/FDA Study)

User Performance of the Karajishi Contour Blood Glucose Monitoring System (ISO/FDA Study)

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02150109
Enrollment
372
Registered
2014-05-29
Start date
2014-06-30
Completion date
2014-06-30
Last updated
2016-02-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study was to determine if untrained persons with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid blood glucose results.

Interventions

DEVICESubjects WITH/WITHOUT Diabetes Use Karajishi Contour BGMS

Untrained subjects WITH and WITHOUT diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

DEVICESubjects WITH Diabetes Use Karajishi Contour BGMS

Untrained subjects WITH diabetes use the Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Males and females, 18 years of age and older * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a BG monitor study using the Karajishi Contour BGMS (or used a Bayer Contour meter) * Work for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors * Work for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk General enrollment guidelines: * Most subjects will be persons with type 1 or type 2 diabetes * At least 10% of subjects with diabetes will have type 1 diabetes * Approximately 10% of subjects will be persons without diabetes * At least 60% of subjects will be younger than age 65

Design outcomes

Primary

MeasureTime frameDescription
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourUntrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Secondary

MeasureTime frameDescription
Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff1 hourStudy staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements1 hourStaff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range.
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourStudy staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff1 hourStudy staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range.
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff1 hourStudy staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range.
Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements1 hourStaff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range1 hourUntrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range.

Countries

United States

Participant flow

Participants by arm

ArmCount
Persons With and Without Diabetes
Untrained subjects WITH/WITHOUT Diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects WITH/WITHOUT Diabetes Use Karajishi Contour BGMS: Untrained subjects WITH (329) and WITHOUT (43) diabetes used Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
372
Total372

Baseline characteristics

CharacteristicPersons With and Without Diabetes
Age, Continuous51.1 years
Ethnicity (NIH/OMB)
Hispanic or Latino
47 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
320 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
7 responses
Race (NIH/OMB)
Asian
23 responses
Race (NIH/OMB)
Black or African American
32 responses
Race (NIH/OMB)
More than one race
6 responses
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
4 responses
Race (NIH/OMB)
Unknown or Not Reported
7 responses
Race (NIH/OMB)
White
305 responses
Region of Enrollment
United States
372 participants
Sex: Female, Male
Female
201 Participants
Sex: Female, Male
Male
171 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
16 / 372
serious
Total, serious adverse events
0 / 372

Outcome results

Primary

Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Untrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 324 (329-5) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Three subjects did not obtain meter BG result after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method317 Blood glucose results within 15mg/dL/15%
Secondary

Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff

Study staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 326 (329-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. One subject had no BG result obtained by study staff. One subject's staff-obtained BG result was not evaluable due to protocol deviation.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff325 Blood glucose results within 15mg/dL/15%
Secondary

Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements

Staff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI find the meter easy to use.328 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI find it easy to do a blood test with this meter326 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsThe meter display is easy to see and read.329 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsIt was easy to understand my test results.329 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI like the overall meter design.326 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsInstructions (User Guide) were easy to understand.326 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsInstructions clearly explain how to run a test.327 number who strongly agree,agree,neutral
Persons With DiabetesNumber of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsInstructions clearly explain meter error messages.329 number who strongly agree,agree,neutral
Secondary

Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 365 (372-7) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Five subjects did not obtain meter BG result after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range353 Blood glucose results within +/- 15%
Secondary

Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range

Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range.

Time frame: 1 hour

Population: 365 (372-7) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Five subjects did not obtain meter BG result after three attempts.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range359 Blood glucose results within +/- 20%
Secondary

Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff

Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range.

Time frame: 1 hour

Population: 369 (372-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject staff test was not performed and for one subject staff test was not evaluable due to protocol deviation.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff365 Blood glucose results within +/- 20%
Secondary

Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff

Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range.

Time frame: 1 hour

Population: 369 (372-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject staff test was not performed and for one subject staff test was not evaluable due to protocol deviation.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff367 Blood glucose results within +/- 20%
Secondary

Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements

Staff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI find it easy to do a blood test with this meter.368 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsThe meter display is easy to see and read.371 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsIt is easy to understand my test results.371 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI like the overall meter design.368 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsI find the meter easy to use.369 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsThe instruction (User Guide) is easy to understand368 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsInstructions clearly explain how to run a test.368 Number who strongly agree,agree,neutral
Persons With DiabetesNumber of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire StatementsInstructions clearly explain meter error messages.371 Number who strongly agree,agree,neutral
Secondary

Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Study staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).

Time frame: 1 hour

Population: 318 (329-11) Blood glucose results were analyzed. Eleven (11) subjects had unsuccessful venipuncture attempts, so no venous results were obtained for them.

ArmMeasureValue (NUMBER)
Persons With DiabetesNumber of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method318 Blood glucose results within 15mg/dL/15%

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026