Diabetes
Conditions
Brief summary
The purpose of this study was to determine if untrained persons with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid blood glucose results.
Interventions
Untrained subjects WITH and WITHOUT diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
Untrained subjects WITH diabetes use the Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
Sponsors
Study design
Eligibility
Inclusion criteria
* Males and females, 18 years of age and older * Able to speak, read, and understand English * Willing to complete all study procedures
Exclusion criteria
* Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a BG monitor study using the Karajishi Contour BGMS (or used a Bayer Contour meter) * Work for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors * Work for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk General enrollment guidelines: * Most subjects will be persons with type 1 or type 2 diabetes * At least 10% of subjects with diabetes will have type 1 diabetes * Approximately 10% of subjects will be persons without diabetes * At least 60% of subjects will be younger than age 65
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 1 hour | Untrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma). |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff | 1 hour | Study staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma). |
| Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | 1 hour | Staff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.' |
| Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range. |
| Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 1 hour | Study staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma). |
| Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff | 1 hour | Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range. |
| Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff | 1 hour | Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range. |
| Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | 1 hour | Staff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.' |
| Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 1 hour | Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Persons With and Without Diabetes Untrained subjects WITH/WITHOUT Diabetes use Karajishi Contour BGMS (Blood Glucose Monitoring System).
Subjects WITH/WITHOUT Diabetes Use Karajishi Contour BGMS: Untrained subjects WITH (329) and WITHOUT (43) diabetes used Karajishi Contour BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick blood (and study staff tested subject fingerstick blood) using the Karajishi Contour BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma. | 372 |
| Total | 372 |
Baseline characteristics
| Characteristic | Persons With and Without Diabetes |
|---|---|
| Age, Continuous | 51.1 years |
| Ethnicity (NIH/OMB) Hispanic or Latino | 47 Participants |
| Ethnicity (NIH/OMB) Not Hispanic or Latino | 320 Participants |
| Ethnicity (NIH/OMB) Unknown or Not Reported | 5 Participants |
| Race (NIH/OMB) American Indian or Alaska Native | 7 responses |
| Race (NIH/OMB) Asian | 23 responses |
| Race (NIH/OMB) Black or African American | 32 responses |
| Race (NIH/OMB) More than one race | 6 responses |
| Race (NIH/OMB) Native Hawaiian or Other Pacific Islander | 4 responses |
| Race (NIH/OMB) Unknown or Not Reported | 7 responses |
| Race (NIH/OMB) White | 305 responses |
| Region of Enrollment United States | 372 participants |
| Sex: Female, Male Female | 201 Participants |
| Sex: Female, Male Male | 171 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 16 / 372 |
| serious Total, serious adverse events | 0 / 372 |
Outcome results
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Untrained subjects WITH diabetes (329) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 324 (329-5) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Three subjects did not obtain meter BG result after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 317 Blood glucose results within 15mg/dL/15% |
Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (329 WITH diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Time frame: 1 hour
Population: 326 (329-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. One subject had no BG result obtained by study staff. One subject's staff-obtained BG result was not evaluable due to protocol deviation.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff | 325 Blood glucose results within 15mg/dL/15% |
Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained responses from persons WITH diabetes (329) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I find the meter easy to use. | 328 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I find it easy to do a blood test with this meter | 326 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | The meter display is easy to see and read. | 329 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | It was easy to understand my test results. | 329 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I like the overall meter design. | 326 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | Instructions (User Guide) were easy to understand. | 326 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | Instructions clearly explain how to run a test. | 327 number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | Instructions clearly explain meter error messages. | 329 number who strongly agree,agree,neutral |
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 365 (372-7) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Five subjects did not obtain meter BG result after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | 353 Blood glucose results within +/- 15% |
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range
Untrained subjects WITH diabetes (329) and WITHOUT diabetes (43) self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-20% of the laboratory method across the entire tested YSI glucose range.
Time frame: 1 hour
Population: 365 (372-7) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject time between subject meter test and collection of subject reference sample exceeded time allowed in protocol. Five subjects did not obtain meter BG result after three attempts.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | 359 Blood glucose results within +/- 20% |
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15% across the tested YSI glucose range.
Time frame: 1 hour
Population: 369 (372-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject staff test was not performed and for one subject staff test was not evaluable due to protocol deviation.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Obtained and Tested by Study Staff | 365 Blood glucose results within +/- 20% |
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff
Study staff obtained and tested subject (WITH and WITHOUT diabetes) fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-20% across the tested YSI glucose range.
Time frame: 1 hour
Population: 369 (372-3) Blood glucose results were analyzed. One (1) subject had no hematocrit result, required per protocol. For one subject staff test was not performed and for one subject staff test was not evaluable due to protocol deviation.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Obtained and Tested by Study Staff | 367 Blood glucose results within +/- 20% |
Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained responses from subjects WITH and WITHOUT diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or 'Neutral' or 'Disagree' or 'Strongly Disagree.'
Time frame: 1 hour
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I find it easy to do a blood test with this meter. | 368 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | The meter display is easy to see and read. | 371 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | It is easy to understand my test results. | 371 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I like the overall meter design. | 368 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | I find the meter easy to use. | 369 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | The instruction (User Guide) is easy to understand | 368 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | Instructions clearly explain how to run a test. | 368 Number who strongly agree,agree,neutral |
| Persons With Diabetes | Number of Subject Responses That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements | Instructions clearly explain meter error messages. | 371 Number who strongly agree,agree,neutral |
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method
Study staff tested venous blood of subjects WITH diabetes (329) using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).
Time frame: 1 hour
Population: 318 (329-11) Blood glucose results were analyzed. Eleven (11) subjects had unsuccessful venipuncture attempts, so no venous results were obtained for them.
| Arm | Measure | Value (NUMBER) |
|---|---|---|
| Persons With Diabetes | Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | 318 Blood glucose results within 15mg/dL/15% |