Cardiac Patients, ECG Monitoring
Conditions
Brief summary
The purpose of this research study is to learn if a new 7 day patch ECG recording monitor can serve the same purpose as standard monitoring systems. The new 7day patch is significantly smaller and lighter than previous ECG recorders.
Detailed description
The primary purpose of this research is to evaluate ECG signal quality of the BDx 7 day ECG Patch heart monitor after 7 full days of wear, establishing that the device is safe and effective to wear for up to 7 days.
Interventions
None listed
Sponsors
Study design
Eligibility
Inclusion criteria
* cardiac patient volunteers (5 male and 5 female)
Exclusion criteria
* skin rash or infection over the sternum * Sternal incision within 3 months from the date of enrollment
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| adhesive performance/durability | 7 days | The study will evaluate adhesive performance/durability by evaluating the Signal Quality of ECG data on the subjects (5 men and 5 women) in accordance with the requirements of FDA's ECG Electrode Guidance (section 5.A.3)1 and the ANSI/AAMI EC-12 Disposable ECG electrode technical standard (sections 4.4 and 5.4)2. |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| human factors | 7 days | Skin comfort or discomfort Device comfort Device stability and contact |
Countries