Measurement of NonInvasive Blood Pressure in Neonates Through Adults and Special Populations

Measurement of NonInvasive Blood Pressure With DINAMAP SuperSTAT and Datex-Ohmeda With Intra-arterial Blood Pressure in Neonates Through Adults and Special Populations

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02125331
Acronym
MISSION
Enrollment
83
Registered
2014-04-29
Start date
2014-04-30
Completion date
2019-03-31
Last updated
2020-06-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure

Keywords

Blood Pressure

Brief summary

This study is required to demonstrate that the non-invasive blood pressure (NIBP) measurement algorithms on two commercially available multifunction hemodynamic acquisition modules, the Patient Data Module (PDM) equipped with the DINAMAP® SUPERSTAT algorithm (PDM-SUPERSTAT) and the Patient Side Module (PSM) equipped with Datex-Ohmeda GE algorithm (PSM-Datex-Ohmeda), provide accurate NIBP measurements in accordance with the guidelines provided by the most recent International Organization for Standardization (ISO) 81060-2:2013. Notably, the new ISO 81060-2:2013 supersedes the previous 2009 version of this standard, and this study is being conducted in accordance with the most recent applicable standards. This study will assess neonate, infant, children, pediatric, adolescent, and adult patients as well as adults with chronic atrial fibrillation in a population that requires non-emergent heart catheterization. This is a study conducted to satisfy both standards for two NIBP devices and, as such, does not require a statistical hypothesis.

Detailed description

This study is required to demonstrate that the non-invasive blood pressure (NIBP) measurement algorithms on two commercially available multifunction hemodynamic acquisition modules, the Patient Data Module (PDM) equipped with the DINAMAP® SUPERSTAT algorithm (PDM-SUPERSTAT) and the Patient Side Module (PSM) equipped with Datex-Ohmeda GE algorithm (PSM-Datex-Ohmeda), provide accurate NIBP measurements in accordance with the guidelines provided by the most recent International Organization for Standardization (ISO) 81060-2:2013. Notably, the new ISO 81060-2:2013 supersedes the previous 2009 version of this standard, and this study is being conducted in accordance with the most recent applicable standards. This study will assess neonate, infant, children, pediatric, adolescent, and adult patients as well as adults with chronic atrial fibrillation in a population that requires non-emergent heart catheterization.

Interventions

PROCEDUREDevice: Blood pressure monitoring with GE Healthcare CARESCAPE B650 Patient Monitor-Patient Side Module (PSM) or Patient Data Module (PDM)

3-15 Non-Invasive Blood Pressure readings on the CARESCAPE B650 Patient Monitor equipped with PSM-Datex-Ohmeda and PDM-SuperSTAT NIBP measurement devices during non-emergent heart catheterization

Sponsors

GE Healthcare
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

Subjects will be included that: 1. Are aged \>29 days requiring clinically indicated non-emergent heart catheterization OR aged ≤29 days with placed or scheduled placement of an indwelling femoral, radial, or umbilical arterial monitoring line; 2. Have an upper limb (right OR left side) that fits a cuff size of the device (circumference ranging 3 cm to 40 cm) OR have a thigh (right OR left side) that fits a cuff size of the device (circumference ranging 38 to 50 cm); 3. Are expected to be able to provide blood pressure measurements using both IBP and NIBP; 4. Are able and willing to provide written informed consent or have a legally authorized representative willing to provide written informed consent with assent from minor patients, as required by IRB policy.

Exclusion criteria

Subjects will be excluded that: 1. Have previously participated in this study (no subject may participate more than once). 2. Exhibit signs or symptoms or have a current diagnosis of peripheral vascular disease in either upper AND/OR lower limbs; 3. Have current, uncontrolled circulatory shock; 4. Exhibit injuries, deformities, intravenous lines, or other abnormalities that, in the opinion of the investigator, may prevent proper cuff application or functioning; 5. For women of child-bearing potential, are currently pregnant, suspected to be pregnant, or are currently lactating; 6. Have any condition that could interfere with the subjects ability to tolerate the procedure, including having a maximum of 4 fast flushes (adult, adolescent, or pediatric patients aged \>29 days) or 1 fast flush (neonates aged \<29 days); 7. If aged greater than 29 days but less than 12 years of age, have previously had any clinical or research procedure requiring general anesthesia in the last 3 month period; 8. If aged greater than 29 days but less than 12 years of age, are expected to require more than three (3) total hours of continuous general anesthesia for the scheduled procedure (including clinically necessary anesthesia and anticipated 25 minute extension for study purposes).

Design outcomes

Primary

MeasureTime frameDescription
Number of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements60 minutesComparison of the following measurements: non-invasive blood pressure to invasive blood pressure. This is done with a beat to beat comparison. Each pulse from the analyzed time interval (non-invasive blood pressure determination) provides a systolic, MAP and diastolic value and that is compared to the waveform collected at the same timepoint for the invasive BP (measurement). Criteria is defined in ANSI/AAMI/ISO 81060-2 NIBP Standard. Mean & SD of the IBP derived from the IBP wave recordings during a NIBP determination will be calculated; and the range of reference IBP will be determined as mean ±1 standard deviation (SD) of the IBPs. The same method will be used for both systolic and diastolic blood pressures. Data may be excluded for analysis when the range of invasive systolic BP is wider than 20 mmHg or when the range of invasive diastolic BP is wider than 12 mmHg, as such that analysis is performed in accordance with ISO 81060-2:2013.

Countries

India, United States

Participant flow

Participants by arm

ArmCount
PDM-SuperSTAT
PDM-SuperSTAT Arm (minimum of 55 evaluable patients): GE DINAMAP® SuperSTAT algorithm delivered by the PDM acquisition module connected to the CARESCAPE Monitor B650 Device: Blood pressure monitoring with GE Healthcare CARESCAPE B650 Patient Monitor-Patient Side Module (PSM) or Patient Data Module (PDM): 3-15 Non-Invasive Blood Pressure readings on the CARESCAPE B650 Patient Monitor equipped with PSM-Datex-Ohmeda and PDM-SuperSTAT NIBP measurement devices during non-emergent heart catheterization
58
PSM-Datex-Ohmeda
PSM-Datex-Ohmeda Arm (minimum of 45 evaluable patients): Datex-Ohmeda delivered by the PSM acquisition module connected to the CARESCAPE Monitor B650 Device: Blood pressure monitoring with GE Healthcare CARESCAPE B650 Patient Monitor-Patient Side Module (PSM) or Patient Data Module (PDM): 3-15 Non-Invasive Blood Pressure readings on the CARESCAPE B650 Patient Monitor equipped with PSM-Datex-Ohmeda and PDM-SuperSTAT NIBP measurement devices during non-emergent heart catheterization
25
Total83

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyProtocol Violation10
Overall Studyscreen failure20
Overall Studytechnical problem20

Baseline characteristics

CharacteristicPDM-SuperSTATPSM-Datex-OhmedaTotal
Age, Customized
Subjects < 3 years
24 Participants6 Participants30 Participants
Age, Customized
Subjects >= 3 years
34 Participants19 Participants53 Participants
Race and Ethnicity Not Collected0 Participants
Sex: Female, Male
Female
25 Participants13 Participants38 Participants
Sex: Female, Male
Male
33 Participants12 Participants45 Participants
Subjects with Chronic Atrial Fibrillation2 Participants0 Participants2 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 580 / 25
other
Total, other adverse events
3 / 580 / 25
serious
Total, serious adverse events
0 / 580 / 25

Outcome results

Primary

Number of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements

Comparison of the following measurements: non-invasive blood pressure to invasive blood pressure. This is done with a beat to beat comparison. Each pulse from the analyzed time interval (non-invasive blood pressure determination) provides a systolic, MAP and diastolic value and that is compared to the waveform collected at the same timepoint for the invasive BP (measurement). Criteria is defined in ANSI/AAMI/ISO 81060-2 NIBP Standard. Mean & SD of the IBP derived from the IBP wave recordings during a NIBP determination will be calculated; and the range of reference IBP will be determined as mean ±1 standard deviation (SD) of the IBPs. The same method will be used for both systolic and diastolic blood pressures. Data may be excluded for analysis when the range of invasive systolic BP is wider than 20 mmHg or when the range of invasive diastolic BP is wider than 12 mmHg, as such that analysis is performed in accordance with ISO 81060-2:2013.

Time frame: 60 minutes

Population: The PDM-SuperSTAT arm was analyzed according to the ISO 81060-2:2013(E) standard. As the PSM-Datex-Ohmeda was prematurely terminated, PASS/FAIL criteria were unable to be determined.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
PDM-SuperSTATNumber of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements40 Participants
PSM-Datex-OhmedaNumber of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements19 Participants

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026