Cardiac Arrhythmia, Cardiomyopathy, Chronic Thromboembolic Pulmonary Hypertension, Heart Failure
Conditions
Keywords
CIED, pacemaker, implantable cardioverter defibrillator, ICD, radiofrequency, cardiac device programming, RF sponge detection, RF assure, cardiac internal electronic devices, temporary pacing, pulmonary thromboendarterectomy, pacing interference, Arrhythmia
Brief summary
If a surgical sponge is mistakenly left inside a patient's body after a surgical procedure, it can cause a serious infection. To prevent this from happening, a new device has been developed that uses radiofrequency (RF) signals to detect the presence of surgical sponges inside the body. The device is now being used routinely to make sure that no sponges are left inside a patient at the end of an operation. However, the RF device has not been implemented in procedures for patients with cardiac implantable electronic devices (CIEDs). While the device is FDA approved for use, there is a theoretical concern that the radiofrequency signals used to detect the sponges will change the settings on the pacemaker or the defibrillator. Changing the settings on a pacemaker might make it pace the heart too quickly or too slowly, while changing the settings on a defibrillator might cause unnecessary shocks or prevent it from shocking the heart if the patient were to have cardiac arrest. The purpose of this study is to test whether the radiofrequency device used to detect sponges can cause a clinically significant change to the settings on pacemakers and defibrillators. To minimize potential risk, the device will be tested only on patients who are having the pacemaker or defibrillator removed or replaced as part of their regular medical care, either because it is infected or because the battery has worn out. Before the pacemaker or defibrillator is removed, the settings will be carefully and completely recorded and the radiofrequency device will be used to scan the body for sponges as it would be done during normal operation. After the pacemaker or defibrillator is taken out, the settings will again be recorded and compared to the settings before the scan. In a standard device removal procedure, no clinically significant change in CIED settings would be expected. If a new pacemaker or defibrillator is implanted in the patient, it will not be exposed to the detection device at all. We will also test whether the RF device has any effect on temporary pacemakers that patients may receive after open heart surgery. We plan to perform testing in a total of 50 patients, 40 with permanent pacemakers or defibrillators and 10 with temporary pacemakers.
Detailed description
Retained surgical items (e.g., sponges, needles, and instruments) are among the most frequently reported medical errors and occur in an estimated 1 in 5000 operations. A retained surgical item frequently causes a serious negative outcome, including re-operation (70-80% of cases), readmission or prolonged hospital stay (30-60% of cases), sepsis (43% of cases), or death (2% of cases). Retained surgical items are classed as never events by the Centers for Medicare and Medicaid Services (CMS), and CMS does not reimburse for additional medical care related to the treatment for such a complication. Sponges are by far the most common retained items and are estimated to account for 50-70% of cases. The standard method to ensure that no sponge is left inside the body is manual counting. Manual counting, however, carries a sensitivity and specificity of only about 77% and 99%, respectively. In one study 62% of retained items were detected after the count was reported as normal. If the manual count suggests that a sponge has been retained, the standard method to detect it is by taking a radiograph. Radiographs, however, are time-consuming and in one large retrospective study failed to detect retained items in 33% of cases. The standard methods of detecting retained surgical sponges are clearly suboptimal. Radiofrequency (RF) technology has recently been employed to improve the detection of retained surgical sponges. A radiofrequency chip is sewn into the fabric of the sponge, and a circular wand that emits a radiofrequency signal is passed over the patient. The wand also serves as an antenna that detects a return signal from the chip in the sponge. If a chip is detected it triggers an audio and a visual alarm on a console attached to the wand. In a study involving 210 patients, RF detection systems had 100% sensitivity and specificity for detecting retained sponges, even in morbidly obese patients. To minimize human error during the scanning process, a new system has been developed where the patient lies on a mat that emits the RF signal and serves as the antenna. In a separate study involving 203 patients, the system using the RF mat had a sensitivity and specificity of 98.5% and 100% respectively. To carry out these studies, sponges were placed underneath patients in a blinded manner. Although it would seem that RF detection systems are superior to manual counting, it should be emphasized that an RF detection system is not a substitute for manual counting, but rather an adjunct to manual counting. The RF Assure ® Detection System manufactured by RF Surgical System Inc. is now routinely used to detect retained sponges in the operating rooms at the University of California, San Diego (UCSD) and at \ 200 other hospitals. There has been some concern that the radio frequency signals emitted by the RF Assure® device might reprogram the settings on CIED such as pacemakers and internal cardiac defibrillators (ICDs), since CIEDs are commonly programmed remotely using RF signals. For this reason, the technology is not commonly used in surgeries with patients with CIEDs. In vitro testing was done in a controlled setting to assess the compatibility of the RF Surgical Detection Technology with permanent pacemakers and ICDs as well as temporary pacemakers, specifically the Boston Scientific Cognis, St. Jude Medical Promote, and Medtronic Virtuoso devices, finding no interference between the RF Assure® system and these devices. Although anecdotal experience and in vitro testing has indicated that the RF Assure® system does not affect CIEDs, it has not been established whether the RF Assure® system is suitable to use in patients with CIEDs in a clinical setting. The purpose of this study is to determine whether the RF Assure® device actually can reprogram CIEDs or cause them to malfunction when used in routine clinical practice. This question is of particular importance because of the rising number of surgical patients, especially cardiac patients, that have an implanted CIED.
Interventions
CIED and temporary pacemaker patients will be scanned with RF surgical sponge detection wand and mat as well as with and without the RF sponge. The scanning will only be conducted on the device scheduled for removal. Patients will have device interrogated before and after scanning (for CIEDs) to determine if there have been any clinically significant changes in programming or settings due to scanning with RF technology. Vitals will be monitored and recorded before, during, and after scanning.
Sponsors
Study design
Eligibility
Inclusion criteria
* Patients must be at least 18 years of age * Patients must be willing and able to provide consent * Patients must be eligible for pacemaker/ICD removal or temporary pacemaker
Exclusion criteria
* Pregnancy * Inability or unwillingness to comply with the protocol * Medical condition that would limit study participation
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The Pacing Mode on the CIED was measured prior to RF scanning (with sponge) and after the scanning (after sponge removal) in order to evaluate if any significant changes in the pacing mode setting resulted. The following CIED modes were evaluated: DDD (dual chamber pacing, sensing, triggered and inhibited mode), VVI (ventricular pacing, sensing, and inhibited mode), DDI (dual pacing, sensing, and inhibited mode), AAI (atrial pacing, sensing, and inhibited mode). The number of patients' device mode switched between these settings was tabulated and is shown in the below table. Multiple post-scan assessments were not made. |
| Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The patient's pacing polarity was measured prior to scanning with sponge and after scanning with sponge via device interrogation. The following pacing polarity measurements were evaluated: right atrium/right ventricle (RA/RV) bipolar polarity, left ventricle (LV) bipolar polarity, and left ventricle (LV) unipolar polarity. Note that not all study patients had LV leads implanted. Multiple post-scan assessments were not made. |
| Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure | Baseline and 15 minutes | The base rate on patients' devices was measured before scanning with sponge and after RF scanning with sponge. The median was determined and is presented below with standard deviation for both times. Multiple post-scan assessments were not made. |
| Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The max tracking rate on the CIEDs was measured prior to scanning with the sponge and post scanning with sponge. The max tracking rate is the maximum atrial rate at which a pacemaker will deliver a ventricular pacing stimulus following each sensed atrial beat. Below, the median and standard deviation are presented. Please note that multiple post-scan assessments were not made for any patient. |
| Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The atrio-ventricular delay was measured prior to scanning with sponge and post RF scanning with sponge. Please note that multiple post-scan assessments were not made for any of the patients presented here. |
| Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | Of the patients enrolled, battery capacity changes were measured prior to and post RF scanning with sponge. Multiple post-scan assessments were not made for any of the patients presented here. |
| Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The impedance values of right ventricle (RV), right atrium (RA), and left ventricle (LV) leads was measured prior to RF scanning with sponge and post RF scanning with sponge. Multiple post-scan assessments were not made for any patient represented here. |
| Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The shock impedance changes were measured (right ventricle (RV) coil and superior vena cava (SVC) coil) prior to and post RF scanning with sponge. This was measured by performing a device interrogation in the electrophysiology (EP) lab. Multiple post-scan measurements were not taken for any of the patients presented here. |
| Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | The P and R waves were measured via device interrogation prior to and post RF scanning with sponge. Multiple post-scan measurements were not made for any of the participants represented here. |
| Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Baseline and 15 minutes | Via device interrogation, capture threshold changes (the minimum amount of electricity that the box has to emit to pace the heart) were measured prior to and post RF Scanning with sponge. Multiple post-scan measurements were not taken for any participants presented here. |
Countries
United States
Participant flow
Recruitment details
Patients that required a pacemaker/ICD generator change or device implant were approached for participation in this study in both the Sulpizio Cardiovascular Center Clinic and Pre-procedural Unit. Patients undergoing PTE Surgery were also approached in these settings.
Participants by arm
| Arm | Count |
|---|---|
| RF Assure Scanning Cardiac implantable electronic device (CIED) (implantable cardiac defibrillator (ICD) and permanent pacemakers (PPM)) and temporary pacemaker patients (during pulmonary thromboendartectomy (PTE) surgery) were scanned with the Radiofrequency (RF) surgical sponge detection wand and mat as well as with and without the RF sponge. The scanning was conducted on the device scheduled for removal. Patients had device interrogated before and after scanning (for CIEDs) to determine if there were any clinically significant changes in programming or settings due to scanning with RF technology. | 50 |
| Total | 50 |
Baseline characteristics
| Characteristic | RF Assure Scanning |
|---|---|
| Age, Continuous ICD | 64 years STANDARD_DEVIATION 17.4 |
| Age, Continuous Pacemaker | 74.5 years STANDARD_DEVIATION 17 |
| Age, Continuous PTE | 54.9 years STANDARD_DEVIATION 13.1 |
| Body Mass Index (BMI) ICD | 29.8 kg/m^2 STANDARD_DEVIATION 6.1 |
| Body Mass Index (BMI) Pacemaker | 26.3 kg/m^2 STANDARD_DEVIATION 3.6 |
| Body Mass Index (BMI) PTE | 30.8 kg/m^2 STANDARD_DEVIATION 3.9 |
| History of Atrial Fibrillation (AF) ICD | 5 participants |
| History of Atrial Fibrillation (AF) Pacemaker | 14 participants |
| History of Atrial Fibrillation (AF) PTE | 2 participants |
| History of Coronary Artery Disease (CAD) ICD | 6 participants |
| History of Coronary Artery Disease (CAD) Pacemaker | 6 participants |
| History of Coronary Artery Disease (CAD) PTE | 1 participants |
| History of Diabetes Mellitus (DM) ICD | 6 participants |
| History of Diabetes Mellitus (DM) Pacemaker | 3 participants |
| History of Diabetes Mellitus (DM) PTE | 3 participants |
| History of Hypertension (HTN) ICD | 10 participants |
| History of Hypertension (HTN) Pacemaker | 15 participants |
| History of Hypertension (HTN) PTE | 5 participants |
| Measured Ejection Fraction (EF) (%) ICD | 37 percentage STANDARD_DEVIATION 17 |
| Measured Ejection Fraction (EF) (%) Pacemaker | 56.2 percentage STANDARD_DEVIATION 17 |
| Measured Ejection Fraction (EF) (%) PTE | 67.6 percentage STANDARD_DEVIATION 10 |
| Region of Enrollment United States | 50 participants |
| Sex: Female, Male Female | 16 Participants |
| Sex: Female, Male Male | 34 Participants |
Adverse events
| Event type | EG000 affected / at risk |
|---|---|
| deaths Total, all-cause mortality | — / — |
| other Total, other adverse events | 0 / 50 |
| serious Total, serious adverse events | 0 / 50 |
Outcome results
Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The atrio-ventricular delay was measured prior to scanning with sponge and post RF scanning with sponge. Please note that multiple post-scan assessments were not made for any of the patients presented here.
Time frame: Baseline and 15 minutes
Population: 31 patients were evaluated of the 40 in this analysis of the AV Delay 9 paced AV Delays were either not measured or unable to be measured during this study. The PTE participants were not included in this analysis because temporary devices do not have capabilities for interrogation (ie: recording AV Delay Changes).
| Arm | Measure | Group | Value (MEDIAN) | Dispersion |
|---|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Sensed AV Delay | 230 ms | Standard Deviation 69.3 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Paced AV Delay | 250 ms | Standard Deviation 68.3 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Sensed AV Delay | 230 ms | Standard Deviation 69.3 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Paced AV Delay | 250 ms | Standard Deviation 68.3 |
Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure
The base rate on patients' devices was measured before scanning with sponge and after RF scanning with sponge. The median was determined and is presented below with standard deviation for both times. Multiple post-scan assessments were not made.
Time frame: Baseline and 15 minutes
Population: The PTE temporary pacing participants were not included in this analysis because temporary devices do not have capabilities for interrogation (ie: recording base rate measurement changes).
| Arm | Measure | Value (MEDIAN) | Dispersion |
|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure | 60 bpm | Standard Deviation 10.27 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure | 60 bpm | Standard Deviation 10.27 |
Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
Of the patients enrolled, battery capacity changes were measured prior to and post RF scanning with sponge. Multiple post-scan assessments were not made for any of the patients presented here.
Time frame: Baseline and 15 minutes
Population: Of all participants, only 24 were analyzed in this outcome, as their devices reported battery life in millivolts (while other devices reported battery life in % or years). The PTE temporary pacing participants weren't included in this analysis because temporary devices do not have capabilities for interrogation.
| Arm | Measure | Value (MEDIAN) | Dispersion |
|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | 2.745 mV | Standard Deviation 0.23 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | 2.744 mV | Standard Deviation 0.23 |
Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
Via device interrogation, capture threshold changes (the minimum amount of electricity that the box has to emit to pace the heart) were measured prior to and post RF Scanning with sponge. Multiple post-scan measurements were not taken for any participants presented here.
Time frame: Baseline and 15 minutes
Population: Because not all devices are equipped with RA, RV, and LV leads, note the population values in the below table for each group: right atrium (RA) thresholds, right ventricle (RV) thresholds, and left ventricle (LV) thresholds. The PTE participants were not included in this analysis because temporary devices do not have capabilities for interrogation.
| Arm | Measure | Group | Value (MEDIAN) | Dispersion |
|---|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Threshold | 0.955 V | Standard Deviation 0.49 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA Threshold | 0.806 V | Standard Deviation 0.35 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV Threshold | 2.616 V | Standard Deviation 1.16 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA Threshold | 0.819 V | Standard Deviation 0.36 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Threshold | 0.972 V | Standard Deviation 0.5 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV Threshold | 2.320 V | Standard Deviation 1.16 |
Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The P and R waves were measured via device interrogation prior to and post RF scanning with sponge. Multiple post-scan measurements were not made for any of the participants represented here.
Time frame: Baseline and 15 minutes
Population: 32 of the 40 participants were evaluated in this study for P and R measurements. 8 patients' devices either didn't have the P or R measured or they was unable to be measured. The PTE temporary pacing participants were not included in this analysis because temporary devices do not have capabilities for interrogation.
| Arm | Measure | Group | Value (MEDIAN) | Dispersion |
|---|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | P Wave | 2.5 mV | Standard Deviation 1.38 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | R Wave | 10.73 mV | Standard Deviation 5.25 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | P Wave | 2.469 mV | Standard Deviation 1.37 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | R Wave | 10.76 mV | Standard Deviation 5.29 |
Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The max tracking rate on the CIEDs was measured prior to scanning with the sponge and post scanning with sponge. The max tracking rate is the maximum atrial rate at which a pacemaker will deliver a ventricular pacing stimulus following each sensed atrial beat. Below, the median and standard deviation are presented. Please note that multiple post-scan assessments were not made for any patient.
Time frame: Baseline and 15 minutes
Population: 34 of 40 participants' Max Tracking Rate was measured. 6 MTRs were not measured or unable to be measured during testing. The PTE temporary pacing participants were not included in this analysis because temporary devices do not have capabilities for interrogation (ie: recording max tracking rate changes).
| Arm | Measure | Value (MEDIAN) | Dispersion |
|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | 130 bpm | Standard Deviation 14.51 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | 130 bpm | Standard Deviation 14.51 |
Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The impedance values of right ventricle (RV), right atrium (RA), and left ventricle (LV) leads was measured prior to RF scanning with sponge and post RF scanning with sponge. Multiple post-scan assessments were not made for any patient represented here.
Time frame: Baseline and 15 minutes
Population: 40 subjects were analyzed. Note that not all had both RA, RV, and LV leads. Therefore, the number of leads (and therefore their impedance values reported below) does not match for each group. The PTE temporary pacing participants were not included in this analysis because temporary devices do not have capabilities for interrogation.
| Arm | Measure | Group | Value (MEDIAN) | Dispersion |
|---|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA Impedance | 500.1 ohms | Standard Deviation 161.18 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Impedance | 531.9 ohms | Standard Deviation 135.96 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV Impedance | 592.8 ohms | Standard Deviation 255.19 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA Impedance | 495.1 ohms | Standard Deviation 161.48 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Impedance | 530.6 ohms | Standard Deviation 135.08 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV Impedance | 610.1 ohms | Standard Deviation 236.89 |
Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The Pacing Mode on the CIED was measured prior to RF scanning (with sponge) and after the scanning (after sponge removal) in order to evaluate if any significant changes in the pacing mode setting resulted. The following CIED modes were evaluated: DDD (dual chamber pacing, sensing, triggered and inhibited mode), VVI (ventricular pacing, sensing, and inhibited mode), DDI (dual pacing, sensing, and inhibited mode), AAI (atrial pacing, sensing, and inhibited mode). The number of patients' device mode switched between these settings was tabulated and is shown in the below table. Multiple post-scan assessments were not made.
Time frame: Baseline and 15 minutes
Population: The PTE participants were not included in this analysis because temporary devices do not have capabilities for interrogation (ie: recording pacing mode changes).
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - DDD | 29 participants |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - VVI | 8 participants |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - DDI | 2 participants |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - AAI | 1 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - AAI | 1 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - DDD | 29 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - DDI | 2 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | Pacing Mode - VVI | 8 participants |
Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The patient's pacing polarity was measured prior to scanning with sponge and after scanning with sponge via device interrogation. The following pacing polarity measurements were evaluated: right atrium/right ventricle (RA/RV) bipolar polarity, left ventricle (LV) bipolar polarity, and left ventricle (LV) unipolar polarity. Note that not all study patients had LV leads implanted. Multiple post-scan assessments were not made.
Time frame: Baseline and 15 minutes
Population: The PTE participants were not included in this analysis because temporary devices do not have capabilities for interrogation (ie: recording pacing polarity changes).
| Arm | Measure | Group | Value (NUMBER) |
|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA/RV (bipolar) | 40 participants |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV (bipolar) | 5 participants |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV (unipolar) | 3 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RA/RV (bipolar) | 40 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV (bipolar) | 5 participants |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | LV (unipolar) | 3 participants |
Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
The shock impedance changes were measured (right ventricle (RV) coil and superior vena cava (SVC) coil) prior to and post RF scanning with sponge. This was measured by performing a device interrogation in the electrophysiology (EP) lab. Multiple post-scan measurements were not taken for any of the patients presented here.
Time frame: Baseline and 15 minutes
Population: SVC and RV Coils are specific to ICDs, and therefore, PPMs and Temporary Pacemakers were not included in this analysis.
| Arm | Measure | Group | Value (MEAN) | Dispersion |
|---|---|---|---|---|
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Coil | 51.30 ohms | Standard Deviation 18.36 |
| Pre-Scan Measures for PPM/ICDs Scanned With RF Assure | Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | SVC Coil | 53.45 ohms | Standard Deviation 17.53 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | RV Coil | 51.5 ohms | Standard Deviation 18.14 |
| Post-Scan Measures for PPM/ICDs Scanned With RF Assure | Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure | SVC Coil | 53.70 ohms | Standard Deviation 17.13 |