Evaluation of the Samsung LABGEO IVD-A20 CHF Test in a Point-of-Care Setting

Evaluation of the Samsung LABGEO IVD-A20 CHF Test in a Point-of-Care Setting

Status
Terminated
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT02107495
Enrollment
191
Registered
2014-04-08
Start date
2014-03-31
Completion date
2015-02-28
Last updated
2016-04-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Congestive Heart Failure

Keywords

CHF

Brief summary

To establish the performance characteristics of the Samsung LABGEO IVD-A20 CHF Test in intended use settings, by comparing test results of the A20 CHF Test with results obtained from an FDA-cleared comparator assay.

Interventions

DEVICESamsung LABGEO IVD-A20 CHF Test

Sponsors

Nexus DX
CollaboratorINDUSTRY
Samsung Electronics
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
21 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* CHF Subjects: 21 years of age or greater, with clinically confirmed CHF or have presented to the clinical site with signs, symptoms and/or risk factors suggestive of heart failure. * Subjects with Potentially confounding comorbidities: Non-CHF subjects 21 years of age or greater, with potentially confounding comorbidities such as diabetes, renal insufficiency, hypertension and chronic obstructive pulmonary disease (COPD) * Healthy Subjects: apparently healthy subjects greater than 45 years of age, with no prior history of cardiac-related disease.

Exclusion criteria

* Apparently healthy subjects with a history of MI, CHF or other cardiac-related disease; * Subjects with acute decompensated heart failure currently on nesiritide therapy; * Subjects having participated in another experimental drug, biologic, or invasive device study within 30 days prior to signing informed consent for this study, or enrolled concurrently in any other investigative study; and * Subjects unable to or refusing to provide written informed consent.

Design outcomes

Primary

MeasureTime frameDescription
The proportion of clinically-confirmed CHF subjects for whom the Samsung LABGEO IVD-A20 CHF Test result was positive (clinical sensitivity).baseline
The proportion of non-CHF subjects for whom the LABGEO IVD A20 CHF Test result was negative (clinical specificity).baseline
Using a Passing-Bablok regression analysis, estimates of intercept and slope will be computed. Estimates of bias and corresponding 95% confidence interval for true bias at clinical decision threshold of 125mg/mL and 450 pg/mL will also be computed.baselineMethod comparison will be assessed using a Passing-Bablok regression analysis to determine the relationship between the Samsung LABGEO IVD-A20 CHF Test measurement in plasma and the corresponding measurement in plasma obtained from the comparator assay.

Secondary

MeasureTime frameDescription
Computation of Receiver Operating Characteristic (ROC) curve and area under the curve (AUC) with 95% confidence interval for the true AUC.baselineThis analysis will assess the ability of the Samsung LABGEO IVD-A20 CHF Test to discriminate between confirmed CHF subjects and non-CHF subjects.
Variance components analysis will be conducted on data collected from each whole blood specimen undergoing point-of-care precision testing at study site. Estimates of within-run, between run, between day and total precision will be computed.10 daysThe Samsung LABGEO IVD-A20 CHF Test result will be modeled as a function of testing days and runs, with replicates fully nested within runs, and runs fully nested within days.
Mixed model regression analysis will be conducted on each specimen type (whole blood vs. plasma) and storage condition.48 hoursThe natural logarithm of the Samsung LABGEO IVD-A20 CHF Test measurement will be modeled as a linear function of time (in hours) after baseline testing.A separate linear function will be fitted to data collected from each study specimen and the intercepts and slops from individual specimens will be modeled as random effects.
Variance components analysis will be conducted on data collected from each plasma specimen undergoing point-of-care precision testing at study site. Estimates of within-run, between run, between day and total precision will be computed.20 daysThe Samsung LABGEO IVD-A20 CHF Test result will be modeled as a function of testing days and runs, with replicates fully nested within runs, and runs fully nested within days.
The proportion of subjects with a positive LABGEO IVD-A20 CHF Test result, for whom a clinical diagnosis of CHF was confirmed (positive predictive value).baseline
The proportion of subjects with a negative LABGEO IVD-A20 CHF Test result, for whom a clinical diagnosis of CHF was confirmed (negative predictive value).baseline

Countries

United States

Outcome results

None listed