HORIZON CE Pivotal Study to Treat Abdominal Aortic Aneurysm

A Prospective Open-label Non-randomized Pivotal Study to Evaluate the Safety and Performance of the Horizon™ Abdominal Aortic Aneurysm (AAA) Stent Graft System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02087501
Enrollment
30
Registered
2014-03-14
Start date
2014-04-30
Completion date
2021-05-31
Last updated
2023-12-05

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aortoiliac Aneurysms, Infrarenal Abdominal Aortic Aneurysms

Keywords

Abdominal Aortic Aneurysm, AAA, Abdominal Aortoiliac Aneurysm, Endovascular Aortic Repair, EVAR

Brief summary

Clinical Investigation Design A prospective, open-label, non-randomized, interventional clinical study, sponsored by Endospan Ltd. Patients will be followed-up for five years. Investigational Device The Horizon™ Abdominal Aortic Aneurysm Stent Graft System and its designated Delivery System. Purpose The purpose of the study is to evaluate the safety and performance of the Horizon™ AAA Stent Graft System for the treatment of infrarenal abdominal aortic and/or aortoiliac aneurysms. The results of this study will be used as supportive data for CE Marking submission in the European Union (EU). Objectives The primary objectives of the study are to evaluate the safety and performance of the Horizon™ AAA Stent graft System. Primary End Points Safety endpoints include proportion of patients free from device related Major Adverse Events (MAEs) within 1 month of the endovascular procedure. Performance endpoints include successful delivery and deployment of the device; and absence of the following at 1 month follow-up: aneurysm growth ≥5mm, type I or III endoleaks, stent graft occlusion, conversion to open surgery, rupture and stent graft migration. Subject population Thirty (30) patients having infrarenal abdominal aortic and/or aortoiliac aneurysms, having Iliac/femoral access vessel morphology that is compatible with vascular access techniques and devices. Treatment All patients will be treated by implantation of the Horizon™ Abdominal Aortic Aneurysm Stent graft System.

Detailed description

Possible benefits: It is believed that the Horizon™ AAA Stent Graft System implantation may offer a number of benefits over conventional and recently developed treatment options for patients suffering from AAA. These other treatment modalities include open surgery and the more recently evolved EVAR. The Horizon™ AAA Stent graft System has a low profile of 14Fr and is implanted through a percutaneous approach. The modular bottom-up construction requires a single-sided femoral access and there is no need for cannulation of the contralateral limb, a requirement present in the majority of commercially available devices. These features are anticipated to result in a simplified procedure with less surgical trauma, shorter procedure time, lower levels of radiation exposure and shorter hospitalization periods. Moreover, the device is flexible enough to reach also tortuous anatomies and its fixation elements reduce the concern of migration and type I endoleaks. As such, it is believed that the Horizon™ AAA Stent graft System may have potential clinical advantages and may exhibit benefits over commercially available stent-grafts. Hypothesis: In this clinical study safety and performance primary endpoints are compared between HORIZON™ and those reported for the Lifeline Registry of Endovascular Aneurysm Repair (Control).

Interventions

DEVICEHorizon™ Abdominal Aortic Aneurysm Stent Graft System

All patient eligible to undergo stent implantation will receive the Horizon™ Abdominal Aortic Aneurysm Stent Graft System and its designated Delivery System.

Sponsors

Endospan Ltd.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Male and female age ≥ 18. 2. Presence of at least one of the following: i. Abdominal aortic aneurysm ≥ 5.0cm in diameter (perpendicular to the line of flow). ii. Abdominal aortic aneurysm 4.5-5.5cm in diameter which increased in size - ≥ 0.5cm over 6 months or 1.0cm over one year. iii. Abdominal aortic aneurysm \>50% larger than the normal aortic diameter. 3. Patient is considered an appropriate candidate for an elective surgery, as evaluated by Physical Status Classification System I, II or III (American Society of Anesthesiologists). 4. Femoral artery diameter of ≥6mm, documented by CTA or MRA that allows endovascular access to the aneurismal site with a 14 Fr delivery catheter. 5. Access vessels morphology suitable for endovascular repair in terms of tortuosity, calcification and angulation, documented by CTA or MRA. 6. To be eligible to receive the Horizon having the following characteristics, as demonstrated on CTA or MRA imaging: I. Infrarenal aortic diameter of 18-28 mm II. Aortic length (lower renal artery to lowest point of aortic bifurcation) of 115-150 mm III. Iliac artery diameter of 10-19 mm IV. Proximal aortic neck length ≥ 15 mm V. Proximal aortic neck angulation ≤ 60° VI. Diameter measured 20mm above aortic bifurcation floor should be ≥ 20mm VII. Aortic bifurcation angulation of ≥ 70°. 7. Patient understands and is voluntarily willing to participate as evidenced by personally signing the Informed Consent document.

Exclusion criteria

1. If female and of childbearing potential , patients who are: 1. pregnant (as determined by a positive pregnancy test performed between 10 to 5 days before implantation date), or 2. intend to become pregnant during the study period, or 3. do not accept to use adequate double barrier contraception methods for the entire study duration to avoid pregnancy 2. Life expectancy of less than 1 year. 3. Any medical condition that, according to the investigator's decision, might expose the patient to increased risk by the investigational device or procedure. 4. Patient is in need for an emergent surgery for a ruptured aneurysm. 5. Patient with an increased risk for aneurysm rupture, such as, saccular aneurysm, aneurysm with isolated wall protrusion and penetrating ulcers of the aorta. 6. A dissecting, acutely ruptured, or leaking aneurysm, or an acute vascular injury due to trauma. 7. Presence of thrombus or atheroma in proximal aortic neck covering \>50% of the endoluminal surface. 8. Presence of diffuse atherosclerotic disease in either common iliac artery that reduces the iliac artery diameter to \<6mm. 9. Congenital vascular disease in which the placement of the stent graft will cause occlusion of major arterial flow. 10. Aneurysm with any of the following characteristics: suprarenal, isolated ilio-femoral, mycotic, inflammatory or para-anastomotic pseudoaneurysm. 11. Patient has an untreated thoracic aneurysm \> 5.5 cm in diameter. 12. Patient has an aneurysm that involves the part of the aorta at the ostia of the renal arteries. 13. Patient has a reversed conical neck defined as a \> 4mm distal increase over a 10 mm length. 14. Patient has ectatic iliac arteries requiring bilateral exclusion of hypogastric blood flow. 15. Patient whose arterial access site is not anticipated to accommodate the 14 Fr diameter of the Horizon™ Delivery System, due to size, tortuosity or hostile groins (scarring, obesity, or previous failed puncture). 16. Patient underwent major surgery or interventional procedure in the last three months. 17. Patient is suffering from unstable angina. 18. Patient has had a myocardial infarction (MI) or cerebral vascular accident (CVA) within 3 months prior to the planned implantation. 19. Patient has a known hypersensitivity or contraindication to anticoagulants, antiplatelets, or contrast media, which is not amenable to pre-treatment. 20. Contraindication to undergoing angiography. 21. Patients with known hypersensitivity or allergy to device materials - Nitinol and Polyester. 22. Morbid obesity or other clinical conditions that severely inhibit x-ray visualization of the aorta. 23. Connective tissue disease (e.g., Marfan's or Ehler's-Danlos syndromes). 24. Patient has a history of bleeding diathesis or coagulopathy. 25. Active systemic infection at the time of the index procedure documented by either one of the following: pain, fever, drainage, positive culture and/or leukocytosis (WBC \> 11,000 mm3). 26. Acute renal failure documented by either one of the following: Creatinine \> 2.00 mg/dl or \> 182 pmol/L; patient on dialysis. 27. Any other medical, social, or psychological issues that in the opinion of the investigator preclude them from receiving this treatment, or the procedures and evaluations pre- and post- treatment. 28. Active participation in another research study involving an investigational device or new drug. \-

Design outcomes

Primary

MeasureTime frameDescription
Safety Endpoint: Major Adverse Events (MAEs)1 month post implantationSafety endpoint includes proportion of patients free from MAEs within 1 month post implantation, as adjudicated by an independent clinical events committee (CEC).
Performance Endpoint1 month post implantPerformance endpoints will assess the rate of successful aneurysm treatment at 1 month, defined as: * Successful delivery and deployment of the device. * Absence of the following, as determined by an independent core-lab: aneurysm growth ≥5mm, type I or III endoleaks, stent graft occlusion, conversion to open surgery, rupture and clinically significant stent graft migration.

Secondary

MeasureTime frameDescription
MAEs1 month - 5 years post implantationProportion of patients free from MAEs at 1 month through 5 years post implantation.
Mortality12 monthsAll-cause mortality and aneurysm-related mortality at 12 months, Kaplan-Meier survival analysis.
Secondary Performance Endpoint12 months post implantationAbsence of all of the following at 1 year: aneurysm growth ≥5mm, type I or III endoleaks, stent graft occlusion, conversion to open surgery, rupture and stent graft migration.

Countries

Italy, Netherlands, Serbia, Switzerland

Participant flow

Recruitment details

Study population included 30 patients enrolled in Hospitals, enrollment period lasted 1 year.

Pre-assignment details

There were no events occurring after patient enrollment and before assignment of the single arm foreseen by the study protocol

Participants by arm

ArmCount
HORIZON AAA Stent Graft
All eligible patients received the HORIZON AAA (Abdominal Aortic Aneurysm) Stent Graft System.
30
Total30

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyDeath7
Overall StudyLost to Follow-up2
Overall StudyOpen repair of aneurysm1

Baseline characteristics

CharacteristicHORIZON AAA Stent Graft
Age, Continuous69.3 years
STANDARD_DEVIATION 7.9
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Italy
12 participants
Region of Enrollment
Netherlands
2 participants
Region of Enrollment
Serbia
11 participants
Region of Enrollment
Switzerland
5 participants
Sex: Female, Male
Female
1 Participants
Sex: Female, Male
Male
29 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
7 / 30
other
Total, other adverse events
18 / 30
serious
Total, serious adverse events
19 / 30

Outcome results

Primary

Performance Endpoint

Performance endpoints will assess the rate of successful aneurysm treatment at 1 month, defined as: * Successful delivery and deployment of the device. * Absence of the following, as determined by an independent core-lab: aneurysm growth ≥5mm, type I or III endoleaks, stent graft occlusion, conversion to open surgery, rupture and clinically significant stent graft migration.

Time frame: 1 month post implant

Population: All subjects for whom the implantation procedure was initiated and for whom there were no major entry violations likely to affect outcome

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
HORIZON AAA Stent GraftPerformance Endpoint30 Participants
95% CI: [88.4, 100]
Primary

Safety Endpoint: Major Adverse Events (MAEs)

Safety endpoint includes proportion of patients free from MAEs within 1 month post implantation, as adjudicated by an independent clinical events committee (CEC).

Time frame: 1 month post implantation

Population: The safety analysis population consisted of all subjects for whom the implantation procedure was initiated (i.e. date of implantation was not missing).~Out of the 30 subjects in the safety analysis set, none had Major Adverse Events (MAEs) within 1 month, and therefore the sample proportion of subjects having MAE's within 1 month is 0%. The 95% confidence interval is \[0%, 11.6%\], meaning that we are 95% confident that the proportion of subjects having MAE's within 1 month is lower than 11.6%.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
HORIZON AAA Stent GraftSafety Endpoint: Major Adverse Events (MAEs)30 Participants
95% CI: [0, 11.6]
95% CI: [0, 11.6]
Secondary

MAEs

Proportion of patients free from MAEs at 1 month through 5 years post implantation.

Time frame: 1 month - 5 years post implantation

Population: All patients for whom the implantation procedure was initiated and for whom there was no major entry violation

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
HORIZON AAA Stent GraftMAEs24 Participants
95% CI: [61, 92]
Secondary

Mortality

All-cause mortality and aneurysm-related mortality at 12 months, Kaplan-Meier survival analysis.

Time frame: 12 months

ArmMeasureValue (NUMBER)
HORIZON AAA Stent GraftMortality93.3 percentage free from mortality
Secondary

Secondary Performance Endpoint

Absence of all of the following at 1 year: aneurysm growth ≥5mm, type I or III endoleaks, stent graft occlusion, conversion to open surgery, rupture and stent graft migration.

Time frame: 12 months post implantation

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
HORIZON AAA Stent GraftSecondary Performance Endpoint28 Participants