Intestinal Anastomosis Complication
Conditions
Keywords
intestinal anastomosis, non-emergent, intestinal revision, magnetic compression, anastomotic leak
Brief summary
Anastomosis of intestine or other viscera currently requires open or laparoscopic surgery and is often the most difficult, time-consuming, and expensive part of many operations. We have developed a device (Magnamosis) that may create compression anastomoses more easily, quickly, and less expensively than sutures or staples. The Magnamosis device consists of two 23-mm diameter, convex-concave, radially symmetric ring magnets encased in polycarbonate. One magnet is placed in the lumen of each viscus to be joined, the magnets self-align, and a compression anastomosis is achieved by tissue remodeling. We have completed extensive pre-clinical studies in animals and have shown that Magnamosis can be used to accomplish gastrojejunostomy, jejunojejunostomy, duodenal-colostomy, and colo-colostomy safely and effectively using available endoscopic and minimally invasive surgery techniques. We are now conducting a small first-in-human study to obtain clinical data in support of the safety and early feasibility of the Magnamosis device.
Detailed description
This is a non-randomized, prospective, single-center pilot study to evaluate the feasibility and safety of creating an intestinal anastomosis using the Magnamosis Magnetic Compression Anastomosis (Magnamosis) device under FDA IDE G130046. Ten, otherwise healthy, subjects, ages 18-60 years, with a disease process necessitating open or laparoscopic surgical anastomosis for re-establishment of intestinal continuity that would otherwise be performed using sutures or stapling devices will be enrolled. Participation in the study requires a time commitment of three months. The total duration of the study is 18 months to ensure three-month follow-up on each subject, with long-term follow-up of each subject at one and two years postoperation. The primary outcome measure will be the incidence of anastomotic leaks related to the use of the Magnamosis device. Anastomotic leakage will be defined as clinical symptoms such as fever, sepsis or peritonitis within 30 days postoperatively leading to a clinical and/or radiological interventional procedure or reoperation that confirms that the leakage is related to the device. Secondary outcome measures will include the rate of occurrence of other device-related complications during hospitalization and post-procedure, including the rate of bleeding, perforation, retention of foreign body, and anastomotic stricture documented by symptoms or imaging within three months of procedure. The duration of hospitalization, number of days to passage of Magnamosis device (found in stool or absent on x-ray), and number of days to the first postoperative bowel movement will also be measured. Post-operative administration of the pain scale will also yield study data. Subject follow-up will be conducted after discharge at 2 weeks (in person, in clinic), and at 1 month, 3 months, 1 year and 2 years either in person, via email, telephone, Skype, or other non in-person method.
Interventions
Single-use, sterile device enables creation of a circular compression intestinal anastomosis. Device is comprised of matched pair of self-centering rare earth magnets encased in medical grade polycarbonate. Geometry of magnets' mating surfaces applies force on inside of mated rings to produce compression necrosis of intervening intestinal walls. Lesser force on outside of mated rings produces inflammatory healing of the two co-apted intestinal walls. Internal lumen creates patency of the anastomosis to prevent intestinal obstruction during compression anastomosis creation. In 5-10 days, mated rings with necrotic tissue falls into intestinal lumen and is naturally expelled with stool, leaving a well-healed anastomosis.
Sponsors
Study design
Eligibility
Inclusion criteria
* Between 18 and ≤ 60 years of age * Requires non-emergent operation to create an intestinal anastomosis for maintenance of intestinal continuity in which the Magnamosis device can be used instead of sutures or staples. * Able to read, speak and understand English * Demonstrates an understanding of the study procedures and risks, and can provide signed informed consent.
Exclusion criteria
* Intestines to be anastomosed are not appropriate in size, thickness or tissue health for the Magnamosis device. For example, * Intestine too small to accommodate 23-mm diameter device; or * Intestine so large that 23-mm diameter anastomotic lumen is not adequate; or * Intestine too thickened to allow two halves of device to come together with sufficient force to produce compression anastomosis (e.g. inflamed or scarred intestinal wall; a foreign body like staples in anastomosis); or * Inadequate blood supply * Requires anastomosis of the stomach * Bowel is not well perfused * Anastomosis will be under tension * Anatomic reconstruction requires crossing a staple line * ASA (American Society of Anesthesiology) score 4 or 5; * Requires more than one anastomosis during operation; * Women possibly or known to be pregnant; * Inability to obtain pre-authorization from insurance company or third party payor * Does not meet all inclusion criteria
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Number of subjects who experience occurrence of anastomotic leak | Up to 2 years postoperation | Subject follow-up will be performed at postoperation 2 weeks, 1 and 3 months, 1 and 2 years to assess the occurrence of an anastomotic leak |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Number of subjects who experience other device-related complications | Up to 2 years postoperation | Other device-related complications include stricture, obstruction, bleeding, delayed expulsion of the device, device failure, bowel perforation will be assessed at 2 weeks, 1 and 3 months, and 1 and 2 years postoperation |
| Number of patients who have an adequate anastomosis | Up to 2 years postoperation | Determine if anastomosis using Magnamosis is adequate. Follow-up will be at 2 weeks, 1 and 3 months, and 1 and 2 years postoperation. |
Countries
United States