Evaluation of an Karajishi TS Investigational Blood Glucose Monitoring System

User Performance of the Karajishi Contour TS Blood Glucose Monitoring System

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT02041377
Enrollment
134
Registered
2014-01-22
Start date
2014-01-31
Completion date
2014-01-31
Last updated
2016-02-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetes

Brief summary

The purpose of this study was to determine if subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) with no training and obtain valid glucose results.

Interventions

DEVICEKarajishi TS Investigational Blood Glucose Monitoring System

Subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi TS Investigational Blood Glucose Monitoring System with no training. All BG results were compared to reference method results obtained from subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.

Sponsors

Ascensia Diabetes Care
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Males and females, 18 years of age and older * People with type 1 or type 2 diabetes * Able to speak, read, and understand English * Willing to complete all study procedures

Exclusion criteria

* Hemophilia or any other bleeding disorder * Pregnancy * Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM * Previously participated in a BG monitor study using the Karajishi TS BGMS (or used a Bayer Contour TS meter) * Working for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors * Working for a competitive medical device company, or having an immediate family member who works for such a company * A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk

Design outcomes

Primary

MeasureTime frameDescription
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourSubjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Secondary

MeasureTime frameDescription
Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourStudy staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).
Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method1 hourSubjects with diabetes self-tested Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff1 hourStudy staff obtained and tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).
Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements1 hourStaff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree.

Countries

United States

Participant flow

Participants by arm

ArmCount
Intended Users of the Monitoring System
Subjects with diabetes used the Karajishi TS Investigational Blood Glucose Monitoring System with no training. The criteria for the intended use population: 1. At least 60% of subjects will be younger than age 65 2. At least 10% of subjects will have type 1 diabetes Karajishi TS Investigational Blood Glucose Monitoring System: Subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi TS Investigational Blood Glucose Monitoring System with no training. All BG results were compared to reference method results obtained from subject capillary plasma. Also, study staff tested subject venous blood and BG results were compared to reference method results obtained from subject venous plasma.
134
Total134

Baseline characteristics

CharacteristicIntended Users of the Monitoring System
Age, Customized
Age of enrolled subjects
53.8 years
Ethnicity (NIH/OMB)
Hispanic or Latino
12 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
121 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
1 Participants
Race (NIH/OMB)
American Indian or Alaska Native
3 responses
Race (NIH/OMB)
Asian
8 responses
Race (NIH/OMB)
Black or African American
11 responses
Race (NIH/OMB)
More than one race
5 responses
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
3 responses
Race (NIH/OMB)
Unknown or Not Reported
6 responses
Race (NIH/OMB)
White
108 responses
Region of Enrollment
United States
134 participants
Sex: Female, Male
Female
71 Participants
Sex: Female, Male
Male
63 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
— / —
other
Total, other adverse events
2 / 134
serious
Total, serious adverse events
0 / 134

Outcome results

Primary

Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 132 (134-2)Blood glucose results were analyzed. Lab reference replicates for one subject were discrepant and not evaluable per protocol. One subject had no fingerstick result.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method127 Blood glucose results
Secondary

Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Subjects with diabetes self-tested Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 129 (134-5) Blood glucose results were analyzed. Lab reference replicates for one subject were discrepant and not evaluable per protocol. One subject had low blood sugar and AST result was not evaluable per protocol. One subject with low blood sugar did not attempt AST testing per protocol. No AST palm results were obtained for 2 subjects.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method121 Blood glucose results
Secondary

Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff

Study staff obtained and tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI capillary plasma) and +/-15% (\>=100 mg/dL YSI capillary plasma).

Time frame: 1 hour

Population: 133 (134-1) Blood glucose results were analyzed. Lab reference replicates for one subject were discrepant and not evaluable per protocol.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff131 Blood glucose results
Secondary

Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements

Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree.

Time frame: 1 hour

ArmMeasureGroupValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsThe instruction (User Guide) is easy to understand132 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsI find it easy to do a blood test with this meter.132 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsThe meter display is easy to see and read.133 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsIt is easy to understand my test results.133 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsI like the overall meter design.134 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsI find the meter easy to use.132 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsInstructions clearly explain how to run a test.134 participants
Intended Users of the Monitoring SystemNumber of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire StatementsInstructions clearly explain meter error messages.134 participants
Secondary

Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method

Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the number of BGMS results within +/-15 mg/dL (\<100 mg/dL YSI venous plasma) and +/-15% (\>=100 mg/dL YSI venous plasma).

Time frame: 1 hour

Population: 131 (134-3) Blood glucose results were analyzed. Lab reference replicates for one subject were discrepant and not evaluable per protocol. Venipuncture was not successful for 2 subjects.

ArmMeasureValue (NUMBER)
Intended Users of the Monitoring SystemNumber of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method131 Blood glucose results

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026