Brief summary
The FloRite™ device is a disposable urine flow meter, indicated for use in a home setting. The device provides Urofowmetry measurements, similar to clinic's measurements. Clinic-based uroflowmetry has some flaws; the setting is artificial and often it is difficult to void at the desired moment. Moreover, a single measurement of the voided parameters is a poor representative of the patient condition due to high variability of the measured parameters. Performing multiple Urofowmetry measurements provides representative and more accurate results, in a more patient-friendly, cost effective method.
Detailed description
The FloRite™ product is a disposable urine flow meter. The patient needs to urinate into the device and at the end of urination the patient needs to disconnect a USB drive from the device, dispose the device's container and keep the USB drive that contains the test data. Primary objectives: • Usability evaluation of the FloRite™ system with LUTS patients. Secondary objectives: * Comparison of uroflowmetric parameters obtained using home disposable digital uroflowmetry device with standard clinic-based uroflowmetry. * Evaluation of diurnal variations in uroflowmetric parameters obtained using home uroflowmetry Primary outcome measures • Successful use of FloRite™ system in LUTS patients, defined as: * Successful unpacking and installation * Successful understanding of the user manual * Ability of the patients to operate the device as define by a successful test completion and data transmission * Successful retrieval of the DOK * Absence of: * Device dropping down * Urination outside the container * Urination on the electronic unit Secondary outcome measures: • Comparison of uroflowmetric parameters as measured by FloRite™ system and standard clinical uroflowmetry as defined by \[Time Frame: Days 0, 1, 6 and 7\]: * Qmax - Max Flow Rate • Evaluation of diurnal variations in uroflowmetric parameters obtained using home uroflowmetry as defined by \[Time Frame: Days 1, 6\]: * Urine flow graph * Tdelay - Delay time * T100 - Voiding Time * TQ - Flow Time * Tqmax - Time to max Flow * Qmax - Max Flow Rate * Qave - Average Flow Rate * Vcomp - Voided Volume Up to 30 completed patients Inclusion criteria * LUTS patients * Male 18\<Age \<65 * Ability to speak, read and understand instructions * Patient willing to sign an Informed Consent Exclusion criteria * Mentally disabled patients * Infectious diseases * Catheterized patients * Buried penis due to obesity The patients will be consented if found eligible. Prospective, self- controlled, interventional, clinical study On the baseline visit, after eligibility is confirmed and after consenting the patients, the following steps will be taken: * Standard clinical uroflowmetry * Training how to use the system * Unpacking and setting up the device (supervised by an observer) The following day, the patient will perform home test with the FloRite™ system. Similar test will be conducted on Day 6 of the study. The test results will be transferred either by e- mailing the data stored on the DOK or by bringing the DOK itself to the clinic on the 7th day. On Day 7 of the study, the patients will be asked to return to the clinic for standard clinical uroflowmetry. In addition, the patient will fill a satisfaction questionnaire. All the measurements must be standard and performed at around the same time.
Interventions
Disposable home use urine flow meter.
Sponsors
Study design
Eligibility
Inclusion criteria
* LUTS patients * Male 18\<Age \<65 * Ability to speak, read and understand instructions * Patient willing to sign an Informed Consent
Exclusion criteria
* Mentally disabled patients * Infectious diseases * Catheterized patients * Buried penis due to obesity
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Usability evaluation of the FloRite™ system with LUTS patients. | Dec-2014 | Primary outcome measures • Successful use of FloRite™ system in LUTS patients, defined as: * Successful unpacking and installation * Successful understanding of the user manual * Ability of the patients to operate the device as define by a successful test completion and data transmission * Successful retrieval of the DOK * Absence of: * Device dropping down * Urination outside the container * Urination on the electronic unit |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Comparison of Tdelay | Dec-2014 | Comparison of uroflowmetric parameters of Tdelay as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of uroflowmetric parameter: T100 - Voiding Time | Dec-2014 | Comparison of uroflowmetric parameters of T100 as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of uroflowmetric parameters of TQ - Flow Time | Dec-2014 | Comparison of uroflowmetric parameters of TQ - Flow Time as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of Qmax (Max Flow Rate) | Dec-2014 | Comparison of uroflowmetric parameters of Qmax as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of uroflowmetric parameters of Qave | Dec-2014 | Comparison of uroflowmetric parameters of Qave as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of uroflowmetric parameters of Vcomp - Voided volume | Dec-2014 | Comparison of uroflowmetric parameters of Vcomp - Voided volume as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
| Comparison of uroflowmetric parameters of Tqmax - Time to max Flow | Dec-2014 | Comparison of uroflowmetric parameters of Tqmax - Time to max Flow as measured by FloRite™ system and standard clinical uroflowmetry \[Time Frame: Days 0, 1, 6 and 7\]. |
Countries
Israel